Minimizing the error in the hemoglobin estimation of blood donors as an articulate step toward donors’ blood management
INTRODUCTION: Hemoglobin (Hb) estimation in blood donors is conducted using capillary samples on portable hemoglobinometers, representing measurement methods in practice. The reference standard is conducted using a venous sample on a hematology analyzer, representing the mentor measurement method or...
Gespeichert in:
Veröffentlicht in: | Asian journal of transfusion science 2024-07, Vol.18 (2), p.270-276 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 276 |
---|---|
container_issue | 2 |
container_start_page | 270 |
container_title | Asian journal of transfusion science |
container_volume | 18 |
creator | Roy, Manisha Kumari, Sangeeta Sachdev, Suchet Sharma, Ratti Ram |
description | INTRODUCTION: Hemoglobin (Hb) estimation in blood donors is conducted using capillary samples on portable hemoglobinometers, representing measurement methods in practice. The reference standard is conducted using a venous sample on a hematology analyzer, representing the mentor measurement method or the true value. The correction involves the calculation of the secondary adjustment factor (SAF) to mitigate the difference between the two values. MATERIAL AND METHODS: A cross-sectional study enrolled 187 blood donors after approval from the institute's ethics committee. On each donor, capillary Hb was performed on the first drop and the second drop of blood using the hemoglobinometer (HCC-1 2) and venous Hb using the hemoglobinometer (HC-V) and hematology analyzer (HA-V) consecutively. The secondary adjustment factor was derived from the slope of the regression equation by calculating the ratio of change in HA-V to the corresponding change in HCC -1 2. RESULTS: The Hb on HCC-1 2 was 15.02 ± 1.30 g/dL 15.03 ± 1.34 g/dL, whereas the Hb on HC-V HA-V was 15.00 ± 1.24 g/dL 14.41 ± 1.19 g/dL respectively. No difference in means of Hb between HCC-1 HCC-2 was observed. The equation to calculate SAF was HA-V = 3.25 + 0.74 x HCC-1 and HA-V= 3.58 + 0.72 x HCC-2 respectively. CONCLUSION: The study highlights the need for Hb cut-off for blood donors specific for the type of sample, the drop of blood in case of capillary sample and use correction with secondary adjustment to strengthen quality assurance. Keywords: Bias, donors' blood management, error, measurement method in practice, mentor measurement method, secondary adjustment factor |
doi_str_mv | 10.4103/ajts.ajts_37_24 |
format | Article |
fullrecord | <record><control><sourceid>gale_doaj_</sourceid><recordid>TN_cdi_proquest_journals_3148390132</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A822924678</galeid><doaj_id>oai_doaj_org_article_0fa8590ba5ed4e26907d626bbd7ab113</doaj_id><sourcerecordid>A822924678</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2424-773f387d5a2720cc0350e64cdfe2d3ca69e652ff6a18a5317330486093d001f03</originalsourceid><addsrcrecordid>eNptUsluFDEQbSGQGAJnrpY498S7u49RxBIpUS5wtqq9dDzqtgfbIxRO_Aa_x5fgzAQiJOSSS1V676m2rntL8JYTzM5hV8v24dNMacqfdRsyjkPPhBTPuw0eFesl5epl96qUHcZCyUFsuvubEMMavoc4o3rnkMs5ZRTiMbhza5qXNLXQlRpWqCFFlDyalpQssimmXBA0iwhyDeawQHWoVLdHNX2D_Afz68fPR84KEWa3ulhfdy88LMW9efRn3ZcP7z9ffuqvbz9eXV5c94ZyynulmGeDsgKootgYzAR2khvrHbXMgBydFNR7CWQAwYhiDPNB4pFZjInH7Ky7OunaBDu9z62NfK8TBH1MpDzrY-2L09jDIEY8gXCWOypHrKykcpqsgokQ1rTenbT2OX09tJnoXTrk2MrXjPCBjZgw-oSaoYmG6FPNYNZQjL4YKB0pl2poqO1_UO1ZtwaTovOh5f8hnJ8IJqdSsvN_myFYP5yAPm7_6QTYbxQ1pyg</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>3148390132</pqid></control><display><type>article</type><title>Minimizing the error in the hemoglobin estimation of blood donors as an articulate step toward donors’ blood management</title><source>Medknow Open Access Medical Journals</source><source>DOAJ Directory of Open Access Journals</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>PubMed Central Open Access</source><source>PubMed Central</source><creator>Roy, Manisha ; Kumari, Sangeeta ; Sachdev, Suchet ; Sharma, Ratti Ram</creator><creatorcontrib>Roy, Manisha ; Kumari, Sangeeta ; Sachdev, Suchet ; Sharma, Ratti Ram</creatorcontrib><description>INTRODUCTION: Hemoglobin (Hb) estimation in blood donors is conducted using capillary samples on portable hemoglobinometers, representing measurement methods in practice. The reference standard is conducted using a venous sample on a hematology analyzer, representing the mentor measurement method or the true value. The correction involves the calculation of the secondary adjustment factor (SAF) to mitigate the difference between the two values. MATERIAL AND METHODS: A cross-sectional study enrolled 187 blood donors after approval from the institute's ethics committee. On each donor, capillary Hb was performed on the first drop and the second drop of blood using the hemoglobinometer (HCC-1 2) and venous Hb using the hemoglobinometer (HC-V) and hematology analyzer (HA-V) consecutively. The secondary adjustment factor was derived from the slope of the regression equation by calculating the ratio of change in HA-V to the corresponding change in HCC -1 2. RESULTS: The Hb on HCC-1 2 was 15.02 ± 1.30 g/dL 15.03 ± 1.34 g/dL, whereas the Hb on HC-V HA-V was 15.00 ± 1.24 g/dL 14.41 ± 1.19 g/dL respectively. No difference in means of Hb between HCC-1 HCC-2 was observed. The equation to calculate SAF was HA-V = 3.25 + 0.74 x HCC-1 and HA-V= 3.58 + 0.72 x HCC-2 respectively. CONCLUSION: The study highlights the need for Hb cut-off for blood donors specific for the type of sample, the drop of blood in case of capillary sample and use correction with secondary adjustment to strengthen quality assurance. Keywords: Bias, donors' blood management, error, measurement method in practice, mentor measurement method, secondary adjustment factor</description><identifier>ISSN: 0973-6247</identifier><identifier>EISSN: 1998-3565</identifier><identifier>DOI: 10.4103/ajts.ajts_37_24</identifier><language>eng</language><publisher>Ahmedabad: Medknow Publications and Media Pvt. Ltd</publisher><subject>bias ; Blood & organ donations ; Blood donors ; donors’ blood management ; error ; Glycosylated hemoglobin ; Hematology ; Hemoglobin ; measurement method in practice ; mentor measurement method ; secondary adjustment factor</subject><ispartof>Asian journal of transfusion science, 2024-07, Vol.18 (2), p.270-276</ispartof><rights>COPYRIGHT 2024 Medknow Publications and Media Pvt. Ltd.</rights><rights>2024. This article is published under (http://creativecommons.org/licenses/by-nc-sa/3.0/) (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c2424-773f387d5a2720cc0350e64cdfe2d3ca69e652ff6a18a5317330486093d001f03</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,860,2095,27903,27904</link.rule.ids></links><search><creatorcontrib>Roy, Manisha</creatorcontrib><creatorcontrib>Kumari, Sangeeta</creatorcontrib><creatorcontrib>Sachdev, Suchet</creatorcontrib><creatorcontrib>Sharma, Ratti Ram</creatorcontrib><title>Minimizing the error in the hemoglobin estimation of blood donors as an articulate step toward donors’ blood management</title><title>Asian journal of transfusion science</title><description>INTRODUCTION: Hemoglobin (Hb) estimation in blood donors is conducted using capillary samples on portable hemoglobinometers, representing measurement methods in practice. The reference standard is conducted using a venous sample on a hematology analyzer, representing the mentor measurement method or the true value. The correction involves the calculation of the secondary adjustment factor (SAF) to mitigate the difference between the two values. MATERIAL AND METHODS: A cross-sectional study enrolled 187 blood donors after approval from the institute's ethics committee. On each donor, capillary Hb was performed on the first drop and the second drop of blood using the hemoglobinometer (HCC-1 2) and venous Hb using the hemoglobinometer (HC-V) and hematology analyzer (HA-V) consecutively. The secondary adjustment factor was derived from the slope of the regression equation by calculating the ratio of change in HA-V to the corresponding change in HCC -1 2. RESULTS: The Hb on HCC-1 2 was 15.02 ± 1.30 g/dL 15.03 ± 1.34 g/dL, whereas the Hb on HC-V HA-V was 15.00 ± 1.24 g/dL 14.41 ± 1.19 g/dL respectively. No difference in means of Hb between HCC-1 HCC-2 was observed. The equation to calculate SAF was HA-V = 3.25 + 0.74 x HCC-1 and HA-V= 3.58 + 0.72 x HCC-2 respectively. CONCLUSION: The study highlights the need for Hb cut-off for blood donors specific for the type of sample, the drop of blood in case of capillary sample and use correction with secondary adjustment to strengthen quality assurance. Keywords: Bias, donors' blood management, error, measurement method in practice, mentor measurement method, secondary adjustment factor</description><subject>bias</subject><subject>Blood & organ donations</subject><subject>Blood donors</subject><subject>donors’ blood management</subject><subject>error</subject><subject>Glycosylated hemoglobin</subject><subject>Hematology</subject><subject>Hemoglobin</subject><subject>measurement method in practice</subject><subject>mentor measurement method</subject><subject>secondary adjustment factor</subject><issn>0973-6247</issn><issn>1998-3565</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNptUsluFDEQbSGQGAJnrpY498S7u49RxBIpUS5wtqq9dDzqtgfbIxRO_Aa_x5fgzAQiJOSSS1V676m2rntL8JYTzM5hV8v24dNMacqfdRsyjkPPhBTPuw0eFesl5epl96qUHcZCyUFsuvubEMMavoc4o3rnkMs5ZRTiMbhza5qXNLXQlRpWqCFFlDyalpQssimmXBA0iwhyDeawQHWoVLdHNX2D_Afz68fPR84KEWa3ulhfdy88LMW9efRn3ZcP7z9ffuqvbz9eXV5c94ZyynulmGeDsgKootgYzAR2khvrHbXMgBydFNR7CWQAwYhiDPNB4pFZjInH7Ky7OunaBDu9z62NfK8TBH1MpDzrY-2L09jDIEY8gXCWOypHrKykcpqsgokQ1rTenbT2OX09tJnoXTrk2MrXjPCBjZgw-oSaoYmG6FPNYNZQjL4YKB0pl2poqO1_UO1ZtwaTovOh5f8hnJ8IJqdSsvN_myFYP5yAPm7_6QTYbxQ1pyg</recordid><startdate>20240701</startdate><enddate>20240701</enddate><creator>Roy, Manisha</creator><creator>Kumari, Sangeeta</creator><creator>Sachdev, Suchet</creator><creator>Sharma, Ratti Ram</creator><general>Medknow Publications and Media Pvt. Ltd</general><general>Medknow Publications & Media Pvt. Ltd</general><general>Wolters Kluwer Medknow Publications</general><scope>AAYXX</scope><scope>CITATION</scope><scope>K9.</scope><scope>DOA</scope></search><sort><creationdate>20240701</creationdate><title>Minimizing the error in the hemoglobin estimation of blood donors as an articulate step toward donors’ blood management</title><author>Roy, Manisha ; Kumari, Sangeeta ; Sachdev, Suchet ; Sharma, Ratti Ram</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2424-773f387d5a2720cc0350e64cdfe2d3ca69e652ff6a18a5317330486093d001f03</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>bias</topic><topic>Blood & organ donations</topic><topic>Blood donors</topic><topic>donors’ blood management</topic><topic>error</topic><topic>Glycosylated hemoglobin</topic><topic>Hematology</topic><topic>Hemoglobin</topic><topic>measurement method in practice</topic><topic>mentor measurement method</topic><topic>secondary adjustment factor</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Roy, Manisha</creatorcontrib><creatorcontrib>Kumari, Sangeeta</creatorcontrib><creatorcontrib>Sachdev, Suchet</creatorcontrib><creatorcontrib>Sharma, Ratti Ram</creatorcontrib><collection>CrossRef</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>DOAJ Directory of Open Access Journals</collection><jtitle>Asian journal of transfusion science</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Roy, Manisha</au><au>Kumari, Sangeeta</au><au>Sachdev, Suchet</au><au>Sharma, Ratti Ram</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Minimizing the error in the hemoglobin estimation of blood donors as an articulate step toward donors’ blood management</atitle><jtitle>Asian journal of transfusion science</jtitle><date>2024-07-01</date><risdate>2024</risdate><volume>18</volume><issue>2</issue><spage>270</spage><epage>276</epage><pages>270-276</pages><issn>0973-6247</issn><eissn>1998-3565</eissn><abstract>INTRODUCTION: Hemoglobin (Hb) estimation in blood donors is conducted using capillary samples on portable hemoglobinometers, representing measurement methods in practice. The reference standard is conducted using a venous sample on a hematology analyzer, representing the mentor measurement method or the true value. The correction involves the calculation of the secondary adjustment factor (SAF) to mitigate the difference between the two values. MATERIAL AND METHODS: A cross-sectional study enrolled 187 blood donors after approval from the institute's ethics committee. On each donor, capillary Hb was performed on the first drop and the second drop of blood using the hemoglobinometer (HCC-1 2) and venous Hb using the hemoglobinometer (HC-V) and hematology analyzer (HA-V) consecutively. The secondary adjustment factor was derived from the slope of the regression equation by calculating the ratio of change in HA-V to the corresponding change in HCC -1 2. RESULTS: The Hb on HCC-1 2 was 15.02 ± 1.30 g/dL 15.03 ± 1.34 g/dL, whereas the Hb on HC-V HA-V was 15.00 ± 1.24 g/dL 14.41 ± 1.19 g/dL respectively. No difference in means of Hb between HCC-1 HCC-2 was observed. The equation to calculate SAF was HA-V = 3.25 + 0.74 x HCC-1 and HA-V= 3.58 + 0.72 x HCC-2 respectively. CONCLUSION: The study highlights the need for Hb cut-off for blood donors specific for the type of sample, the drop of blood in case of capillary sample and use correction with secondary adjustment to strengthen quality assurance. Keywords: Bias, donors' blood management, error, measurement method in practice, mentor measurement method, secondary adjustment factor</abstract><cop>Ahmedabad</cop><pub>Medknow Publications and Media Pvt. Ltd</pub><doi>10.4103/ajts.ajts_37_24</doi><tpages>7</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0973-6247 |
ispartof | Asian journal of transfusion science, 2024-07, Vol.18 (2), p.270-276 |
issn | 0973-6247 1998-3565 |
language | eng |
recordid | cdi_proquest_journals_3148390132 |
source | Medknow Open Access Medical Journals; DOAJ Directory of Open Access Journals; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; PubMed Central Open Access; PubMed Central |
subjects | bias Blood & organ donations Blood donors donors’ blood management error Glycosylated hemoglobin Hematology Hemoglobin measurement method in practice mentor measurement method secondary adjustment factor |
title | Minimizing the error in the hemoglobin estimation of blood donors as an articulate step toward donors’ blood management |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-21T14%3A52%3A49IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Minimizing%20the%20error%20in%20the%20hemoglobin%20estimation%20of%20blood%20donors%20as%20an%20articulate%20step%20toward%20donors%E2%80%99%20blood%20management&rft.jtitle=Asian%20journal%20of%20transfusion%20science&rft.au=Roy,%20Manisha&rft.date=2024-07-01&rft.volume=18&rft.issue=2&rft.spage=270&rft.epage=276&rft.pages=270-276&rft.issn=0973-6247&rft.eissn=1998-3565&rft_id=info:doi/10.4103/ajts.ajts_37_24&rft_dat=%3Cgale_doaj_%3EA822924678%3C/gale_doaj_%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=3148390132&rft_id=info:pmid/&rft_galeid=A822924678&rft_doaj_id=oai_doaj_org_article_0fa8590ba5ed4e26907d626bbd7ab113&rfr_iscdi=true |