Realization of the triple point of carbon dioxide in a transportable cell using long-stem SPRTs
The action of the Minamata Convention requests in the near term to develop alternative fixed points to replace the mercury triple point (Hg TP) from the set of the defining points of the International Temperature Scale (ITS) of 1990. Carbon dioxide (CO 2 ) TP is a candidate that exhibits the advanta...
Gespeichert in:
Veröffentlicht in: | Metrologia 2023-02, Vol.60 (1), p.15006 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | |
---|---|
container_issue | 1 |
container_start_page | 15006 |
container_title | Metrologia |
container_volume | 60 |
creator | Liang, Yu Zhang, J T Feng, X J Qiu, P |
description | The action of the Minamata Convention requests in the near term to develop alternative fixed points to replace the mercury triple point (Hg TP) from the set of the defining points of the International Temperature Scale (ITS) of 1990. Carbon dioxide (CO
2
) TP is a candidate that exhibits the advantages of high chemical and physical stability and low health risk, and the disadvantage of relatively high pressure at room temperature. We report our study on the realization of the CO
2
TP using long-stem standard platinum resistance thermometers (SPRTs). We present a novel design for a CO
2
TP cell with thermometer immersion depth extending from 110 –200 mm with no increase in sample gas pressure. The outer dimension of the new cell is suitable for a drop-in compatible replacement of a normal cell for the Hg TP. We realized 14 plateaus of the CO
2
TP using two long-stem SPRTs over a span of 12 months. The mean temperature at the CO
2
TP is 216.591 30 (36) K at the melt fraction
F
= 0.5 and 216.591 36 (37) K at
F =
1.0. Within the given uncertainties, the uncertainty arising from the transformation to ITS-90 contributes a major part of 0.29 –0.30 mK. The uncertainty formed by the pressure-head effect and the axial heat leak through the thermometer well of the new cell, the impurities of the sample gas and the inaccuracy of the electronic instruments contribute a smaller part of 0.21 –0.22 mK. The plateaus consistently exhibit small slopes of heights of less than 0.21 mK and time spans from 147 –347 h between
F
= 0.3 and
F
= 0.75. The small slopes of the plateaus demonstrate the repeatability of 0.056 and 0.059 mK, respectively, for the melt fraction
F
= 0.5 and 1.0, and the long-term stability of 0.031 mK. |
doi_str_mv | 10.1088/1681-7575/aca417 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_journals_2753718625</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2753718625</sourcerecordid><originalsourceid>FETCH-LOGICAL-c308t-2f853c63ce1461c3ff8815ebea76652f057a275809476c7d13824c1542fa23223</originalsourceid><addsrcrecordid>eNp1kEtLAzEURoMoWKt7lwG3js1j8uhSilahoNS6Dmma1JTpZExSUH-9GUZ05erCd893LxwALjG6wUjKCeYSV4IJNtFG11gcgdFvdAxGCBFeYTqtT8FZSjuEsCBMjIBaWt34L519aGFwML9ZmKPvGgu74NvcZ0bHddlufPjwGwt9C3VhdJu6ELNeF9TYpoGH5NstbEK7rVK2e_jyvFylc3DidJPsxc8cg9f7u9XsoVo8zR9nt4vKUCRzRZxk1HBqLK45NtQ5KTGza6sF54w4xIQmgkk0rQU3YoOpJLXBrCZOE0oIHYOr4W4Xw_vBpqx24RDb8lKVHhVYcsIKhQbKxJBStE510e91_FQYqV6j6p2p3pkaNJbK9VDxofu7-S_-Dekscnw</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2753718625</pqid></control><display><type>article</type><title>Realization of the triple point of carbon dioxide in a transportable cell using long-stem SPRTs</title><source>IOP Publishing Journals</source><source>Institute of Physics (IOP) Journals - HEAL-Link</source><creator>Liang, Yu ; Zhang, J T ; Feng, X J ; Qiu, P</creator><creatorcontrib>Liang, Yu ; Zhang, J T ; Feng, X J ; Qiu, P</creatorcontrib><description>The action of the Minamata Convention requests in the near term to develop alternative fixed points to replace the mercury triple point (Hg TP) from the set of the defining points of the International Temperature Scale (ITS) of 1990. Carbon dioxide (CO
2
) TP is a candidate that exhibits the advantages of high chemical and physical stability and low health risk, and the disadvantage of relatively high pressure at room temperature. We report our study on the realization of the CO
2
TP using long-stem standard platinum resistance thermometers (SPRTs). We present a novel design for a CO
2
TP cell with thermometer immersion depth extending from 110 –200 mm with no increase in sample gas pressure. The outer dimension of the new cell is suitable for a drop-in compatible replacement of a normal cell for the Hg TP. We realized 14 plateaus of the CO
2
TP using two long-stem SPRTs over a span of 12 months. The mean temperature at the CO
2
TP is 216.591 30 (36) K at the melt fraction
F
= 0.5 and 216.591 36 (37) K at
F =
1.0. Within the given uncertainties, the uncertainty arising from the transformation to ITS-90 contributes a major part of 0.29 –0.30 mK. The uncertainty formed by the pressure-head effect and the axial heat leak through the thermometer well of the new cell, the impurities of the sample gas and the inaccuracy of the electronic instruments contribute a smaller part of 0.21 –0.22 mK. The plateaus consistently exhibit small slopes of heights of less than 0.21 mK and time spans from 147 –347 h between
F
= 0.3 and
F
= 0.75. The small slopes of the plateaus demonstrate the repeatability of 0.056 and 0.059 mK, respectively, for the melt fraction
F
= 0.5 and 1.0, and the long-term stability of 0.031 mK.</description><identifier>ISSN: 0026-1394</identifier><identifier>EISSN: 1681-7575</identifier><identifier>DOI: 10.1088/1681-7575/aca417</identifier><identifier>CODEN: MTRGAU</identifier><language>eng</language><publisher>Bristol: IOP Publishing</publisher><subject>Carbon dioxide ; Gas pressure ; International Temperature Scale ; Pressure effects ; Pressure head ; Resistance thermometers ; Room temperature ; Stability ; Stems ; Temperature scales ; Thermometers ; Thermometry ; triple point of carbon dioxide ; triple point of mercury ; Uncertainty</subject><ispartof>Metrologia, 2023-02, Vol.60 (1), p.15006</ispartof><rights>2022 BIPM & IOP Publishing Ltd</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c308t-2f853c63ce1461c3ff8815ebea76652f057a275809476c7d13824c1542fa23223</citedby><cites>FETCH-LOGICAL-c308t-2f853c63ce1461c3ff8815ebea76652f057a275809476c7d13824c1542fa23223</cites><orcidid>0000-0002-5204-3403 ; 0000-0002-0612-187X</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://iopscience.iop.org/article/10.1088/1681-7575/aca417/pdf$$EPDF$$P50$$Giop$$H</linktopdf><link.rule.ids>314,776,780,27901,27902,53821,53868</link.rule.ids></links><search><creatorcontrib>Liang, Yu</creatorcontrib><creatorcontrib>Zhang, J T</creatorcontrib><creatorcontrib>Feng, X J</creatorcontrib><creatorcontrib>Qiu, P</creatorcontrib><title>Realization of the triple point of carbon dioxide in a transportable cell using long-stem SPRTs</title><title>Metrologia</title><addtitle>MET</addtitle><addtitle>Metrologia</addtitle><description>The action of the Minamata Convention requests in the near term to develop alternative fixed points to replace the mercury triple point (Hg TP) from the set of the defining points of the International Temperature Scale (ITS) of 1990. Carbon dioxide (CO
2
) TP is a candidate that exhibits the advantages of high chemical and physical stability and low health risk, and the disadvantage of relatively high pressure at room temperature. We report our study on the realization of the CO
2
TP using long-stem standard platinum resistance thermometers (SPRTs). We present a novel design for a CO
2
TP cell with thermometer immersion depth extending from 110 –200 mm with no increase in sample gas pressure. The outer dimension of the new cell is suitable for a drop-in compatible replacement of a normal cell for the Hg TP. We realized 14 plateaus of the CO
2
TP using two long-stem SPRTs over a span of 12 months. The mean temperature at the CO
2
TP is 216.591 30 (36) K at the melt fraction
F
= 0.5 and 216.591 36 (37) K at
F =
1.0. Within the given uncertainties, the uncertainty arising from the transformation to ITS-90 contributes a major part of 0.29 –0.30 mK. The uncertainty formed by the pressure-head effect and the axial heat leak through the thermometer well of the new cell, the impurities of the sample gas and the inaccuracy of the electronic instruments contribute a smaller part of 0.21 –0.22 mK. The plateaus consistently exhibit small slopes of heights of less than 0.21 mK and time spans from 147 –347 h between
F
= 0.3 and
F
= 0.75. The small slopes of the plateaus demonstrate the repeatability of 0.056 and 0.059 mK, respectively, for the melt fraction
F
= 0.5 and 1.0, and the long-term stability of 0.031 mK.</description><subject>Carbon dioxide</subject><subject>Gas pressure</subject><subject>International Temperature Scale</subject><subject>Pressure effects</subject><subject>Pressure head</subject><subject>Resistance thermometers</subject><subject>Room temperature</subject><subject>Stability</subject><subject>Stems</subject><subject>Temperature scales</subject><subject>Thermometers</subject><subject>Thermometry</subject><subject>triple point of carbon dioxide</subject><subject>triple point of mercury</subject><subject>Uncertainty</subject><issn>0026-1394</issn><issn>1681-7575</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><recordid>eNp1kEtLAzEURoMoWKt7lwG3js1j8uhSilahoNS6Dmma1JTpZExSUH-9GUZ05erCd893LxwALjG6wUjKCeYSV4IJNtFG11gcgdFvdAxGCBFeYTqtT8FZSjuEsCBMjIBaWt34L519aGFwML9ZmKPvGgu74NvcZ0bHddlufPjwGwt9C3VhdJu6ELNeF9TYpoGH5NstbEK7rVK2e_jyvFylc3DidJPsxc8cg9f7u9XsoVo8zR9nt4vKUCRzRZxk1HBqLK45NtQ5KTGza6sF54w4xIQmgkk0rQU3YoOpJLXBrCZOE0oIHYOr4W4Xw_vBpqx24RDb8lKVHhVYcsIKhQbKxJBStE510e91_FQYqV6j6p2p3pkaNJbK9VDxofu7-S_-Dekscnw</recordid><startdate>20230201</startdate><enddate>20230201</enddate><creator>Liang, Yu</creator><creator>Zhang, J T</creator><creator>Feng, X J</creator><creator>Qiu, P</creator><general>IOP Publishing</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7U5</scope><scope>8FD</scope><scope>L7M</scope><orcidid>https://orcid.org/0000-0002-5204-3403</orcidid><orcidid>https://orcid.org/0000-0002-0612-187X</orcidid></search><sort><creationdate>20230201</creationdate><title>Realization of the triple point of carbon dioxide in a transportable cell using long-stem SPRTs</title><author>Liang, Yu ; Zhang, J T ; Feng, X J ; Qiu, P</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c308t-2f853c63ce1461c3ff8815ebea76652f057a275809476c7d13824c1542fa23223</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Carbon dioxide</topic><topic>Gas pressure</topic><topic>International Temperature Scale</topic><topic>Pressure effects</topic><topic>Pressure head</topic><topic>Resistance thermometers</topic><topic>Room temperature</topic><topic>Stability</topic><topic>Stems</topic><topic>Temperature scales</topic><topic>Thermometers</topic><topic>Thermometry</topic><topic>triple point of carbon dioxide</topic><topic>triple point of mercury</topic><topic>Uncertainty</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Liang, Yu</creatorcontrib><creatorcontrib>Zhang, J T</creatorcontrib><creatorcontrib>Feng, X J</creatorcontrib><creatorcontrib>Qiu, P</creatorcontrib><collection>CrossRef</collection><collection>Solid State and Superconductivity Abstracts</collection><collection>Technology Research Database</collection><collection>Advanced Technologies Database with Aerospace</collection><jtitle>Metrologia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Liang, Yu</au><au>Zhang, J T</au><au>Feng, X J</au><au>Qiu, P</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Realization of the triple point of carbon dioxide in a transportable cell using long-stem SPRTs</atitle><jtitle>Metrologia</jtitle><stitle>MET</stitle><addtitle>Metrologia</addtitle><date>2023-02-01</date><risdate>2023</risdate><volume>60</volume><issue>1</issue><spage>15006</spage><pages>15006-</pages><issn>0026-1394</issn><eissn>1681-7575</eissn><coden>MTRGAU</coden><abstract>The action of the Minamata Convention requests in the near term to develop alternative fixed points to replace the mercury triple point (Hg TP) from the set of the defining points of the International Temperature Scale (ITS) of 1990. Carbon dioxide (CO
2
) TP is a candidate that exhibits the advantages of high chemical and physical stability and low health risk, and the disadvantage of relatively high pressure at room temperature. We report our study on the realization of the CO
2
TP using long-stem standard platinum resistance thermometers (SPRTs). We present a novel design for a CO
2
TP cell with thermometer immersion depth extending from 110 –200 mm with no increase in sample gas pressure. The outer dimension of the new cell is suitable for a drop-in compatible replacement of a normal cell for the Hg TP. We realized 14 plateaus of the CO
2
TP using two long-stem SPRTs over a span of 12 months. The mean temperature at the CO
2
TP is 216.591 30 (36) K at the melt fraction
F
= 0.5 and 216.591 36 (37) K at
F =
1.0. Within the given uncertainties, the uncertainty arising from the transformation to ITS-90 contributes a major part of 0.29 –0.30 mK. The uncertainty formed by the pressure-head effect and the axial heat leak through the thermometer well of the new cell, the impurities of the sample gas and the inaccuracy of the electronic instruments contribute a smaller part of 0.21 –0.22 mK. The plateaus consistently exhibit small slopes of heights of less than 0.21 mK and time spans from 147 –347 h between
F
= 0.3 and
F
= 0.75. The small slopes of the plateaus demonstrate the repeatability of 0.056 and 0.059 mK, respectively, for the melt fraction
F
= 0.5 and 1.0, and the long-term stability of 0.031 mK.</abstract><cop>Bristol</cop><pub>IOP Publishing</pub><doi>10.1088/1681-7575/aca417</doi><tpages>21</tpages><orcidid>https://orcid.org/0000-0002-5204-3403</orcidid><orcidid>https://orcid.org/0000-0002-0612-187X</orcidid></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0026-1394 |
ispartof | Metrologia, 2023-02, Vol.60 (1), p.15006 |
issn | 0026-1394 1681-7575 |
language | eng |
recordid | cdi_proquest_journals_2753718625 |
source | IOP Publishing Journals; Institute of Physics (IOP) Journals - HEAL-Link |
subjects | Carbon dioxide Gas pressure International Temperature Scale Pressure effects Pressure head Resistance thermometers Room temperature Stability Stems Temperature scales Thermometers Thermometry triple point of carbon dioxide triple point of mercury Uncertainty |
title | Realization of the triple point of carbon dioxide in a transportable cell using long-stem SPRTs |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-02T23%3A35%3A03IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Realization%20of%20the%20triple%20point%20of%20carbon%20dioxide%20in%20a%20transportable%20cell%20using%20long-stem%20SPRTs&rft.jtitle=Metrologia&rft.au=Liang,%20Yu&rft.date=2023-02-01&rft.volume=60&rft.issue=1&rft.spage=15006&rft.pages=15006-&rft.issn=0026-1394&rft.eissn=1681-7575&rft.coden=MTRGAU&rft_id=info:doi/10.1088/1681-7575/aca417&rft_dat=%3Cproquest_cross%3E2753718625%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2753718625&rft_id=info:pmid/&rfr_iscdi=true |