72-LB: Clinical Evaluation of a Long-Term Optical Continuous Glucose Monitoring (CGM) System: Lumee Glucose

Profusa is developing a CGM System which tracks glucose concentrations in interstitial fluid for several months. It consists of a small hydrogel sensor (2.5 x 0.4 x 0.3 mm) that is permanently inserted by subcutaneous injection, and a wireless optical reader that is placed over the skin to measure t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Diabetes (New York, N.Y.) N.Y.), 2021-06, Vol.70 (Supplement_1)
Hauptverfasser: TRAN, KHANH Q., NGUYEN, CHLOE, NGUYEN, HIEP T., GAMSEY, SOYA J., LEPAK, CLAYTON G., REBRIN, KERSTIN
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue Supplement_1
container_start_page
container_title Diabetes (New York, N.Y.)
container_volume 70
creator TRAN, KHANH Q.
NGUYEN, CHLOE
NGUYEN, HIEP T.
GAMSEY, SOYA J.
LEPAK, CLAYTON G.
REBRIN, KERSTIN
description Profusa is developing a CGM System which tracks glucose concentrations in interstitial fluid for several months. It consists of a small hydrogel sensor (2.5 x 0.4 x 0.3 mm) that is permanently inserted by subcutaneous injection, and a wireless optical reader that is placed over the skin to measure the sensor's fluorescent signal, which changes proportionally with glucose. We evaluated the sensor signal response compared to reference capillary blood glucose (BG) taken from a commercial glucose meter in seven subjects with type 1 and type 2 diabetes. Each subject received two sensor injections in the upper arm during the first visit, followed by multiple 6-hour ambulatory monitoring visits over a three-month period. Of the 14 sensors injected, 13 tracked glucose excursions during every monitoring visit up to 88 days. Applying retrospective calibration analysis to 106 sensor traces showed that 84% of the sensor traces had strong correlation with BG reference values, defined by r2 > 0.5. Parkes Error Grid Analysis revealed that of the paired sensor-reference values, 88.9% fell in zone A, 10.9% in zone B, and 0.2% in zone C (Figure 1). The Lumee Glucose System demonstrates glucose tracking functionality for a minimum of 3 months in subjects with type 1 and type 2 diabetes. Additional studies are ongoing to establish prospective calibration algorithms and to assess measurement timepoints beyond three months.
doi_str_mv 10.2337/db21-72-LB
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_journals_2562268245</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2562268245</sourcerecordid><originalsourceid>FETCH-LOGICAL-c635-3c9fc4c24c1852dd655d6c4e3720c9339a723884b606c8ede572813b331b084e3</originalsourceid><addsrcrecordid>eNo10EFLwzAUB_AgCs7pxU8Q8KJCNHlpm3Y3V-YUOnawB2-hTdPR2SYzaYV9e7tNeYd3-fF_vD9Ct4w-AefiuSqBEQEkm5-hCUt4QjiIz3M0oZQBYSIRl-jK-y2lNBpngr6OeIbTtjGNKlq8-Cnaoegba7CtcYEzazYk167D611_FKk1fWMGO3i8bAdlvcYra5reusZs8H26XD3gj73vdTfD2dBp_c-u0UVdtF7f_O0pyl8XefpGsvXyPX3JiIp4SLhKahUoCBSLQ6iqKAyrSAWaC6Aq4TwpBPA4DsrxARXrSocCYsZLzllJ49FN0d0pdufs96B9L7d2cGa8KCGMAKIYgnBUjyelnPXe6VruXNMVbi8ZlYcu5aFLKUBmc_4LKrhktw</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2562268245</pqid></control><display><type>article</type><title>72-LB: Clinical Evaluation of a Long-Term Optical Continuous Glucose Monitoring (CGM) System: Lumee Glucose</title><source>PubMed Central Free</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>TRAN, KHANH Q. ; NGUYEN, CHLOE ; NGUYEN, HIEP T. ; GAMSEY, SOYA J. ; LEPAK, CLAYTON G. ; REBRIN, KERSTIN</creator><creatorcontrib>TRAN, KHANH Q. ; NGUYEN, CHLOE ; NGUYEN, HIEP T. ; GAMSEY, SOYA J. ; LEPAK, CLAYTON G. ; REBRIN, KERSTIN</creatorcontrib><description>Profusa is developing a CGM System which tracks glucose concentrations in interstitial fluid for several months. It consists of a small hydrogel sensor (2.5 x 0.4 x 0.3 mm) that is permanently inserted by subcutaneous injection, and a wireless optical reader that is placed over the skin to measure the sensor's fluorescent signal, which changes proportionally with glucose. We evaluated the sensor signal response compared to reference capillary blood glucose (BG) taken from a commercial glucose meter in seven subjects with type 1 and type 2 diabetes. Each subject received two sensor injections in the upper arm during the first visit, followed by multiple 6-hour ambulatory monitoring visits over a three-month period. Of the 14 sensors injected, 13 tracked glucose excursions during every monitoring visit up to 88 days. Applying retrospective calibration analysis to 106 sensor traces showed that 84% of the sensor traces had strong correlation with BG reference values, defined by r2 &gt; 0.5. Parkes Error Grid Analysis revealed that of the paired sensor-reference values, 88.9% fell in zone A, 10.9% in zone B, and 0.2% in zone C (Figure 1). The Lumee Glucose System demonstrates glucose tracking functionality for a minimum of 3 months in subjects with type 1 and type 2 diabetes. Additional studies are ongoing to establish prospective calibration algorithms and to assess measurement timepoints beyond three months.</description><identifier>ISSN: 0012-1797</identifier><identifier>EISSN: 1939-327X</identifier><identifier>DOI: 10.2337/db21-72-LB</identifier><language>eng</language><publisher>New York: American Diabetes Association</publisher><subject>Diabetes ; Diabetes mellitus (insulin dependent) ; Diabetes mellitus (non-insulin dependent) ; Glucose ; Glucose monitoring ; Hydrogels ; Sensors</subject><ispartof>Diabetes (New York, N.Y.), 2021-06, Vol.70 (Supplement_1)</ispartof><rights>Copyright American Diabetes Association Jun 1, 2021</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids></links><search><creatorcontrib>TRAN, KHANH Q.</creatorcontrib><creatorcontrib>NGUYEN, CHLOE</creatorcontrib><creatorcontrib>NGUYEN, HIEP T.</creatorcontrib><creatorcontrib>GAMSEY, SOYA J.</creatorcontrib><creatorcontrib>LEPAK, CLAYTON G.</creatorcontrib><creatorcontrib>REBRIN, KERSTIN</creatorcontrib><title>72-LB: Clinical Evaluation of a Long-Term Optical Continuous Glucose Monitoring (CGM) System: Lumee Glucose</title><title>Diabetes (New York, N.Y.)</title><description>Profusa is developing a CGM System which tracks glucose concentrations in interstitial fluid for several months. It consists of a small hydrogel sensor (2.5 x 0.4 x 0.3 mm) that is permanently inserted by subcutaneous injection, and a wireless optical reader that is placed over the skin to measure the sensor's fluorescent signal, which changes proportionally with glucose. We evaluated the sensor signal response compared to reference capillary blood glucose (BG) taken from a commercial glucose meter in seven subjects with type 1 and type 2 diabetes. Each subject received two sensor injections in the upper arm during the first visit, followed by multiple 6-hour ambulatory monitoring visits over a three-month period. Of the 14 sensors injected, 13 tracked glucose excursions during every monitoring visit up to 88 days. Applying retrospective calibration analysis to 106 sensor traces showed that 84% of the sensor traces had strong correlation with BG reference values, defined by r2 &gt; 0.5. Parkes Error Grid Analysis revealed that of the paired sensor-reference values, 88.9% fell in zone A, 10.9% in zone B, and 0.2% in zone C (Figure 1). The Lumee Glucose System demonstrates glucose tracking functionality for a minimum of 3 months in subjects with type 1 and type 2 diabetes. Additional studies are ongoing to establish prospective calibration algorithms and to assess measurement timepoints beyond three months.</description><subject>Diabetes</subject><subject>Diabetes mellitus (insulin dependent)</subject><subject>Diabetes mellitus (non-insulin dependent)</subject><subject>Glucose</subject><subject>Glucose monitoring</subject><subject>Hydrogels</subject><subject>Sensors</subject><issn>0012-1797</issn><issn>1939-327X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNo10EFLwzAUB_AgCs7pxU8Q8KJCNHlpm3Y3V-YUOnawB2-hTdPR2SYzaYV9e7tNeYd3-fF_vD9Ct4w-AefiuSqBEQEkm5-hCUt4QjiIz3M0oZQBYSIRl-jK-y2lNBpngr6OeIbTtjGNKlq8-Cnaoegba7CtcYEzazYk167D611_FKk1fWMGO3i8bAdlvcYra5reusZs8H26XD3gj73vdTfD2dBp_c-u0UVdtF7f_O0pyl8XefpGsvXyPX3JiIp4SLhKahUoCBSLQ6iqKAyrSAWaC6Aq4TwpBPA4DsrxARXrSocCYsZLzllJ49FN0d0pdufs96B9L7d2cGa8KCGMAKIYgnBUjyelnPXe6VruXNMVbi8ZlYcu5aFLKUBmc_4LKrhktw</recordid><startdate>20210601</startdate><enddate>20210601</enddate><creator>TRAN, KHANH Q.</creator><creator>NGUYEN, CHLOE</creator><creator>NGUYEN, HIEP T.</creator><creator>GAMSEY, SOYA J.</creator><creator>LEPAK, CLAYTON G.</creator><creator>REBRIN, KERSTIN</creator><general>American Diabetes Association</general><scope>AAYXX</scope><scope>CITATION</scope><scope>K9.</scope><scope>NAPCQ</scope></search><sort><creationdate>20210601</creationdate><title>72-LB: Clinical Evaluation of a Long-Term Optical Continuous Glucose Monitoring (CGM) System: Lumee Glucose</title><author>TRAN, KHANH Q. ; NGUYEN, CHLOE ; NGUYEN, HIEP T. ; GAMSEY, SOYA J. ; LEPAK, CLAYTON G. ; REBRIN, KERSTIN</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c635-3c9fc4c24c1852dd655d6c4e3720c9339a723884b606c8ede572813b331b084e3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Diabetes</topic><topic>Diabetes mellitus (insulin dependent)</topic><topic>Diabetes mellitus (non-insulin dependent)</topic><topic>Glucose</topic><topic>Glucose monitoring</topic><topic>Hydrogels</topic><topic>Sensors</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>TRAN, KHANH Q.</creatorcontrib><creatorcontrib>NGUYEN, CHLOE</creatorcontrib><creatorcontrib>NGUYEN, HIEP T.</creatorcontrib><creatorcontrib>GAMSEY, SOYA J.</creatorcontrib><creatorcontrib>LEPAK, CLAYTON G.</creatorcontrib><creatorcontrib>REBRIN, KERSTIN</creatorcontrib><collection>CrossRef</collection><collection>ProQuest Health &amp; Medical Complete (Alumni)</collection><collection>Nursing &amp; Allied Health Premium</collection><jtitle>Diabetes (New York, N.Y.)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>TRAN, KHANH Q.</au><au>NGUYEN, CHLOE</au><au>NGUYEN, HIEP T.</au><au>GAMSEY, SOYA J.</au><au>LEPAK, CLAYTON G.</au><au>REBRIN, KERSTIN</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>72-LB: Clinical Evaluation of a Long-Term Optical Continuous Glucose Monitoring (CGM) System: Lumee Glucose</atitle><jtitle>Diabetes (New York, N.Y.)</jtitle><date>2021-06-01</date><risdate>2021</risdate><volume>70</volume><issue>Supplement_1</issue><issn>0012-1797</issn><eissn>1939-327X</eissn><abstract>Profusa is developing a CGM System which tracks glucose concentrations in interstitial fluid for several months. It consists of a small hydrogel sensor (2.5 x 0.4 x 0.3 mm) that is permanently inserted by subcutaneous injection, and a wireless optical reader that is placed over the skin to measure the sensor's fluorescent signal, which changes proportionally with glucose. We evaluated the sensor signal response compared to reference capillary blood glucose (BG) taken from a commercial glucose meter in seven subjects with type 1 and type 2 diabetes. Each subject received two sensor injections in the upper arm during the first visit, followed by multiple 6-hour ambulatory monitoring visits over a three-month period. Of the 14 sensors injected, 13 tracked glucose excursions during every monitoring visit up to 88 days. Applying retrospective calibration analysis to 106 sensor traces showed that 84% of the sensor traces had strong correlation with BG reference values, defined by r2 &gt; 0.5. Parkes Error Grid Analysis revealed that of the paired sensor-reference values, 88.9% fell in zone A, 10.9% in zone B, and 0.2% in zone C (Figure 1). The Lumee Glucose System demonstrates glucose tracking functionality for a minimum of 3 months in subjects with type 1 and type 2 diabetes. Additional studies are ongoing to establish prospective calibration algorithms and to assess measurement timepoints beyond three months.</abstract><cop>New York</cop><pub>American Diabetes Association</pub><doi>10.2337/db21-72-LB</doi></addata></record>
fulltext fulltext
identifier ISSN: 0012-1797
ispartof Diabetes (New York, N.Y.), 2021-06, Vol.70 (Supplement_1)
issn 0012-1797
1939-327X
language eng
recordid cdi_proquest_journals_2562268245
source PubMed Central Free; EZB-FREE-00999 freely available EZB journals
subjects Diabetes
Diabetes mellitus (insulin dependent)
Diabetes mellitus (non-insulin dependent)
Glucose
Glucose monitoring
Hydrogels
Sensors
title 72-LB: Clinical Evaluation of a Long-Term Optical Continuous Glucose Monitoring (CGM) System: Lumee Glucose
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-18T18%3A05%3A40IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=72-LB:%20Clinical%20Evaluation%20of%20a%20Long-Term%20Optical%20Continuous%20Glucose%20Monitoring%20(CGM)%20System:%20Lumee%20Glucose&rft.jtitle=Diabetes%20(New%20York,%20N.Y.)&rft.au=TRAN,%20KHANH%20Q.&rft.date=2021-06-01&rft.volume=70&rft.issue=Supplement_1&rft.issn=0012-1797&rft.eissn=1939-327X&rft_id=info:doi/10.2337/db21-72-LB&rft_dat=%3Cproquest_cross%3E2562268245%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2562268245&rft_id=info:pmid/&rfr_iscdi=true