Comparative study of Single Crystal (SC)-Diamond and 4H-SiC bulk radiation detectors for room temperature alpha spectroscopy
In this work, the fabrication, electrical characterization and alpha spectral characterization of single-crystal (SC) diamond bulk radiation detector have been carried out. The performance metrics of the SC-diamond bulk detector such as the leakage current density, charge collection efficiency, spec...
Gespeichert in:
Veröffentlicht in: | Journal of instrumentation 2021-06, Vol.16 (6), p.P06020 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this work, the fabrication, electrical characterization
and alpha spectral characterization of single-crystal (SC) diamond
bulk radiation detector have been carried out. The performance
metrics of the SC-diamond bulk detector such as the leakage current
density, charge collection efficiency, spectral resolution and
polarization effects are studied and are compared with a bulk
detector fabricated using High-Pure Semi-Insulating (HPSI)
4H-SiC. It is found that the SC-diamond detector exhibits lower
leakage current density, higher μτ product, better spectral
resolution and no polarization effects in comparison to the HPSI
4H-SiC detector. The time response analysis of the SC-diamond
detector is carried out to study the effect of alpha particles on
the operation of the SC-diamond detector. Accordingly, the stability
of the SC-diamond detector is tested by evaluating the total counts
versus real time and the performance of the diamond detector is
found to be stable during the alpha-time spectroscopy. Hence,
diamond bulk detectors using commercially available electronic grade
diamond may be better suited for alpha particle spectroscopy
compared to HPSI 4H-SiC of nearly equivalent thickness. |
---|---|
ISSN: | 1748-0221 1748-0221 |
DOI: | 10.1088/1748-0221/16/06/P06020 |