A Scanner Jitter Test Method of CR System Based on Improved Hough Transform Algorithm

Computed radiology system (CR) test technology has been applied in the testing of industrial welding defect after the promulgation and implementation of digital radiographic imaging test standards. The performance of CR system directly affects the testing rate of welding defects. Therefore, testing...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of physics. Conference series 2021-03, Vol.1827 (1), p.12161
Hauptverfasser: Wenming, Guo, Huilin, Zhao
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Computed radiology system (CR) test technology has been applied in the testing of industrial welding defect after the promulgation and implementation of digital radiographic imaging test standards. The performance of CR system directly affects the testing rate of welding defects. Therefore, testing and evaluating the performance of CR system becomes the premise and guarantee for the application of CR system. In this paper, based on the testing image of CR phantom and the simulation of scanner dither results, an algorithm combining fault-tolerant voting and improved point-based hoof transform is proposed to test the performance index of scanner dither in CR system. Experimental results show that the proposed algorithm can test not only the row and column dithering, but also the subtle arc dithering. Compared with standard Hough transform algorithm, this algorithm has higher accuracy and can avoid the influence of other image indicators around the T-shaped image quality indicator used in CR phantom.
ISSN:1742-6588
1742-6596
DOI:10.1088/1742-6596/1827/1/012161