Role of surface finishing on the in vitro biological properties of a silicon nitride–titanium nitride (Si3N4–TiN) composite
Silicon nitride (Si 3 N 4 ) has been introduced clinically as an orthopedic biomaterial for interbody fusion devices and in joint replacements. However, the production of complex shapes through conventional mechanical machining is difficult and expensive and limits interesting applications. Thus, se...
Gespeichert in:
Veröffentlicht in: | Journal of materials science 2017, Vol.52 (1), p.467-477 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Silicon nitride (Si
3
N
4
) has been introduced clinically as an orthopedic biomaterial for interbody fusion devices and in joint replacements. However, the production of complex shapes through conventional mechanical machining is difficult and expensive and limits interesting applications. Thus, several electrically conductive reinforcements to the Si
3
N
4
matrix, like TiN, have been proposed, generating composites suitable to be wrought by electrical discharge machining (EDM). In this study, Si
3
N
4
–TiN with high strength, low density, and good electric conductivity wrought by EDM was studied. The role of surface finishing was investigated comparing the interface generated during the EDM process to that resulting from further polishing. The different topographical features were assessed by electron microscopy, energy dispersive X-ray spectrometry, and profilometry. Surface wettability was also determined based on the measurement of the OCA of water and diiodomethane. The biological responses induced in MC3T3 cells, a widely diffused osteoblast model, were correlated with the surface pattern. The unpolished samples could promote better cell viability, with a more relevant effect on the cytoskeleton arrangement as highlighted by numerous cytoplasmic extensions and filopodia-like structures and the high number of focal adhesions, while MC3T3 cells grown on polished Si
3
N
4
–TiN specimens displayed a flat morphology. In addition, the unpolished Si
3
N
4
–TiN increased osteocalcin production and calcium deposition. Taken together, these data support the biocompatibility and in vitro osteogenic properties of the electroconductive Si
3
N
4
–TiN investigated. Further in vivo studies are required to explore the possible use of bone implants directly obtained by EDM. |
---|---|
ISSN: | 0022-2461 1573-4803 |
DOI: | 10.1007/s10853-016-0346-1 |