İki farklı endodontik cetvelin (endoblok) cetvel ve kalibrasyon deliklerinin ölçüm doğruluklarının incelenmesi

The depth of the file-length-calibration holes located on the upper side of the endoblock (between 12.0-27.5) was measured using the same caliper but by means of transferring the depth to a #40 file with silicone stopper. Statistical control limits and specification limits were compared to examine t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta odontologica turcica 2020-09, Vol.37 (3), p.84-90
Hauptverfasser: Yücel, Olga, Sarıçam, Esma, Altunkaynak, Bülent, Kayaoğlu, Güven
Format: Artikel
Sprache:tur
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The depth of the file-length-calibration holes located on the upper side of the endoblock (between 12.0-27.5) was measured using the same caliper but by means of transferring the depth to a #40 file with silicone stopper. Statistical control limits and specification limits were compared to examine the measurement quality of each endoblock brand. The specification limits were calculated in 3 standard deviation (sigma) widths taking the reference values into account indicated on the product. Control limits were calculated utilizing the measurements obtained during product testing.
ISSN:2147-690X
DOI:10.17214/aaziaot.662943