Fast, Miniaturized, Real-Time Unit for Sampling, Modulation, and Separation in Detection of Hazardous Chemicals
A sampling, modulation, and separation (SMS) unit was tested for detection of hazardous chemicals. The SMS unit, designed and developed for on-site sampling and analysis, consists of a dynamic inlet system coupled with a fast, miniaturized gas chromatograph (GC). Feasibility of the SMS unit was eval...
Gespeichert in:
Veröffentlicht in: | Analytical chemistry (Washington) 2020-11, Vol.92 (21), p.14589-14593 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A sampling, modulation, and separation (SMS) unit was tested for detection of hazardous chemicals. The SMS unit, designed and developed for on-site sampling and analysis, consists of a dynamic inlet system coupled with a fast, miniaturized gas chromatograph (GC). Feasibility of the SMS unit was evaluated together with a hazardous chemical vapor generator. The performance of the SMS unit was tested with automated thermal desorption after SMS to collect samples for GC-mass spectrometry (GC-MS) measurements. Detection of sarin nerve agent was verified. Additionally, the vapor generator was connected to the SMS unit, which was hyphenated with a photoionization detector (PID), thus creating a fast GC-PID system. This system gave a positive response for degradation products of sulfur mustard, thereby indicating suitability of the SMS-PID unit for field drone applications. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/acs.analchem.0c02898 |