A relative humidity calibration from 5°C to 45°C in a mixed-flow humidity generator
The paper presents a method used in the Vinca Institute of Nuclear Sciences for a reliable and traceable relative humidity calibration in the temperature range from 5?C to 45?C. Inside a controllable temperature and humidity environment, supplied by a mixed-flow humidity generator, measurements of h...
Gespeichert in:
Veröffentlicht in: | Thermal science 2012, Vol.16 (1), p.193-205 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The paper presents a method used in the Vinca Institute of Nuclear Sciences
for a reliable and traceable relative humidity calibration in the temperature
range from 5?C to 45?C. Inside a controllable temperature and humidity
environment, supplied by a mixed-flow humidity generator, measurements of
hygrometers under calibration are compared with those of calibrated reference
instruments. A traceability chain from temperature to reference relative
humidity and next to the hygrometers under calibrations is provided by using
a chilled-mirror dew-point temperature system and precise relative humidity
probes. Corresponding calibration uncertainties are analyzed, particularly
those associated to the temperature uniformity of controlled calibration
environment. Two examples of relative humidity calibration with dew-point and
relative humidity reference measurements in the range from 15 to 75% of RH
and 5?C to 45?C are presented and discussed.