910-P: Improved Accuracy of 14-Day Factory-Calibrated FreeStyle Libre System with New Glucose Algorithm

We evaluated the accuracy of FreeStyle Libre 14-day system with new glucose algorithm in adults ages 18 years or older. This was a non-randomized, multi-center, single-arm study that enrolled 144 subjects with types 1 and 2 diabetes on intensive insulin therapy at 5 U.S. sites. Subjects participated...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Diabetes (New York, N.Y.) N.Y.), 2019-06, Vol.68 (Supplement_1)
Hauptverfasser: KARINKA, SHRIDHARA ALVA, BAILEY, TIMOTHY S., BRAZG, RONALD L., BUDIMAN, ERWIN S., CASTORINO, KRISTIN, CHRISTIANSEN, MARK P., LILJENQUIST, DAVID R., LIU, HANQING
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue Supplement_1
container_start_page
container_title Diabetes (New York, N.Y.)
container_volume 68
creator KARINKA, SHRIDHARA ALVA
BAILEY, TIMOTHY S.
BRAZG, RONALD L.
BUDIMAN, ERWIN S.
CASTORINO, KRISTIN
CHRISTIANSEN, MARK P.
LILJENQUIST, DAVID R.
LIU, HANQING
description We evaluated the accuracy of FreeStyle Libre 14-day system with new glucose algorithm in adults ages 18 years or older. This was a non-randomized, multi-center, single-arm study that enrolled 144 subjects with types 1 and 2 diabetes on intensive insulin therapy at 5 U.S. sites. Subjects participated in 3 clinic sessions of up to 10 hours each over the 14- day wear period to obtain comparator data. Glucose concentrations were manipulated to obtain glucose concentrations across the system dynamic range of 40-500mg/dL. System performance was evaluated in terms of percent accuracy with respect to venous plasma reference using YSI2300. Accuracy was characterized as the proportion of sensor results that were within ±15%, ± 20% of their paired YSI values for glucose levels ≥ 70 mg/dL or within ± 15 mg/dL, ± 20 mg/dL of YSI for glucose levels < 70 mg/dL. Other accuracy metrics such as mean bias and mean absolute relative difference (MARD) of the FreeStyle Libre System with respect to YSI were also evaluated and are presented in Table 1. FreeStyle Libre System with new algorithm showed 92.8% of the results within ±20mg/dL/±20% of the venous plasma reference with an overall MARD of 9.2% based on the data from 144 subjects with evaluable sensors.
doi_str_mv 10.2337/db19-910-P
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_journals_2248403094</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2248403094</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1044-26814962a090ad4e0d567271ec52ea87939eae69b3d7c101c6b364960a7415373</originalsourceid><addsrcrecordid>eNotkE9LAzEQxYMoWKsXP0HAmxCd_Omm8VaqrYWihSp4C9nstLbsujXZtey3N7Uyh4HH783wHiHXHO6ElPq-yLlhhgNbnJAeN9IwKfTHKekBcMG4NvqcXMS4BYAsTY-s_-AHOqt2of7Bgo68b4PzHa1XlCv26Do6cb6pQ8fGrtzkwTWJmgTEZdOVSOdJQrrsYoMV3W-aT_qCezotW19HpKNyXYckVpfkbOXKiFf_u0_eJ09v42c2f53OxqM58xyUYiIbcmUy4cCAKxRCMci00Bz9QKAb6pQIHWYml4VODu6zXGbJAE4rPpBa9snN8W6K891ibOy2bsNXemmFUEMFEoxK1O2R8qGOMeDK7sKmcqGzHOyhSHso0qZu7EL-AmYkYxs</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2248403094</pqid></control><display><type>article</type><title>910-P: Improved Accuracy of 14-Day Factory-Calibrated FreeStyle Libre System with New Glucose Algorithm</title><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>PubMed Central</source><creator>KARINKA, SHRIDHARA ALVA ; BAILEY, TIMOTHY S. ; BRAZG, RONALD L. ; BUDIMAN, ERWIN S. ; CASTORINO, KRISTIN ; CHRISTIANSEN, MARK P. ; LILJENQUIST, DAVID R. ; LIU, HANQING</creator><creatorcontrib>KARINKA, SHRIDHARA ALVA ; BAILEY, TIMOTHY S. ; BRAZG, RONALD L. ; BUDIMAN, ERWIN S. ; CASTORINO, KRISTIN ; CHRISTIANSEN, MARK P. ; LILJENQUIST, DAVID R. ; LIU, HANQING</creatorcontrib><description>We evaluated the accuracy of FreeStyle Libre 14-day system with new glucose algorithm in adults ages 18 years or older. This was a non-randomized, multi-center, single-arm study that enrolled 144 subjects with types 1 and 2 diabetes on intensive insulin therapy at 5 U.S. sites. Subjects participated in 3 clinic sessions of up to 10 hours each over the 14- day wear period to obtain comparator data. Glucose concentrations were manipulated to obtain glucose concentrations across the system dynamic range of 40-500mg/dL. System performance was evaluated in terms of percent accuracy with respect to venous plasma reference using YSI2300. Accuracy was characterized as the proportion of sensor results that were within ±15%, ± 20% of their paired YSI values for glucose levels ≥ 70 mg/dL or within ± 15 mg/dL, ± 20 mg/dL of YSI for glucose levels &lt; 70 mg/dL. Other accuracy metrics such as mean bias and mean absolute relative difference (MARD) of the FreeStyle Libre System with respect to YSI were also evaluated and are presented in Table 1. FreeStyle Libre System with new algorithm showed 92.8% of the results within ±20mg/dL/±20% of the venous plasma reference with an overall MARD of 9.2% based on the data from 144 subjects with evaluable sensors.</description><identifier>ISSN: 0012-1797</identifier><identifier>EISSN: 1939-327X</identifier><identifier>DOI: 10.2337/db19-910-P</identifier><language>eng</language><publisher>New York: American Diabetes Association</publisher><subject>Accuracy ; Algorithms ; Diabetes ; Diabetes mellitus ; Glucose ; Insulin</subject><ispartof>Diabetes (New York, N.Y.), 2019-06, Vol.68 (Supplement_1)</ispartof><rights>Copyright American Diabetes Association Jun 1, 2019</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c1044-26814962a090ad4e0d567271ec52ea87939eae69b3d7c101c6b364960a7415373</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids></links><search><creatorcontrib>KARINKA, SHRIDHARA ALVA</creatorcontrib><creatorcontrib>BAILEY, TIMOTHY S.</creatorcontrib><creatorcontrib>BRAZG, RONALD L.</creatorcontrib><creatorcontrib>BUDIMAN, ERWIN S.</creatorcontrib><creatorcontrib>CASTORINO, KRISTIN</creatorcontrib><creatorcontrib>CHRISTIANSEN, MARK P.</creatorcontrib><creatorcontrib>LILJENQUIST, DAVID R.</creatorcontrib><creatorcontrib>LIU, HANQING</creatorcontrib><title>910-P: Improved Accuracy of 14-Day Factory-Calibrated FreeStyle Libre System with New Glucose Algorithm</title><title>Diabetes (New York, N.Y.)</title><description>We evaluated the accuracy of FreeStyle Libre 14-day system with new glucose algorithm in adults ages 18 years or older. This was a non-randomized, multi-center, single-arm study that enrolled 144 subjects with types 1 and 2 diabetes on intensive insulin therapy at 5 U.S. sites. Subjects participated in 3 clinic sessions of up to 10 hours each over the 14- day wear period to obtain comparator data. Glucose concentrations were manipulated to obtain glucose concentrations across the system dynamic range of 40-500mg/dL. System performance was evaluated in terms of percent accuracy with respect to venous plasma reference using YSI2300. Accuracy was characterized as the proportion of sensor results that were within ±15%, ± 20% of their paired YSI values for glucose levels ≥ 70 mg/dL or within ± 15 mg/dL, ± 20 mg/dL of YSI for glucose levels &lt; 70 mg/dL. Other accuracy metrics such as mean bias and mean absolute relative difference (MARD) of the FreeStyle Libre System with respect to YSI were also evaluated and are presented in Table 1. FreeStyle Libre System with new algorithm showed 92.8% of the results within ±20mg/dL/±20% of the venous plasma reference with an overall MARD of 9.2% based on the data from 144 subjects with evaluable sensors.</description><subject>Accuracy</subject><subject>Algorithms</subject><subject>Diabetes</subject><subject>Diabetes mellitus</subject><subject>Glucose</subject><subject>Insulin</subject><issn>0012-1797</issn><issn>1939-327X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><recordid>eNotkE9LAzEQxYMoWKsXP0HAmxCd_Omm8VaqrYWihSp4C9nstLbsujXZtey3N7Uyh4HH783wHiHXHO6ElPq-yLlhhgNbnJAeN9IwKfTHKekBcMG4NvqcXMS4BYAsTY-s_-AHOqt2of7Bgo68b4PzHa1XlCv26Do6cb6pQ8fGrtzkwTWJmgTEZdOVSOdJQrrsYoMV3W-aT_qCezotW19HpKNyXYckVpfkbOXKiFf_u0_eJ09v42c2f53OxqM58xyUYiIbcmUy4cCAKxRCMci00Bz9QKAb6pQIHWYml4VODu6zXGbJAE4rPpBa9snN8W6K891ibOy2bsNXemmFUEMFEoxK1O2R8qGOMeDK7sKmcqGzHOyhSHso0qZu7EL-AmYkYxs</recordid><startdate>20190601</startdate><enddate>20190601</enddate><creator>KARINKA, SHRIDHARA ALVA</creator><creator>BAILEY, TIMOTHY S.</creator><creator>BRAZG, RONALD L.</creator><creator>BUDIMAN, ERWIN S.</creator><creator>CASTORINO, KRISTIN</creator><creator>CHRISTIANSEN, MARK P.</creator><creator>LILJENQUIST, DAVID R.</creator><creator>LIU, HANQING</creator><general>American Diabetes Association</general><scope>AAYXX</scope><scope>CITATION</scope><scope>K9.</scope><scope>NAPCQ</scope></search><sort><creationdate>20190601</creationdate><title>910-P: Improved Accuracy of 14-Day Factory-Calibrated FreeStyle Libre System with New Glucose Algorithm</title><author>KARINKA, SHRIDHARA ALVA ; BAILEY, TIMOTHY S. ; BRAZG, RONALD L. ; BUDIMAN, ERWIN S. ; CASTORINO, KRISTIN ; CHRISTIANSEN, MARK P. ; LILJENQUIST, DAVID R. ; LIU, HANQING</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1044-26814962a090ad4e0d567271ec52ea87939eae69b3d7c101c6b364960a7415373</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><topic>Accuracy</topic><topic>Algorithms</topic><topic>Diabetes</topic><topic>Diabetes mellitus</topic><topic>Glucose</topic><topic>Insulin</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>KARINKA, SHRIDHARA ALVA</creatorcontrib><creatorcontrib>BAILEY, TIMOTHY S.</creatorcontrib><creatorcontrib>BRAZG, RONALD L.</creatorcontrib><creatorcontrib>BUDIMAN, ERWIN S.</creatorcontrib><creatorcontrib>CASTORINO, KRISTIN</creatorcontrib><creatorcontrib>CHRISTIANSEN, MARK P.</creatorcontrib><creatorcontrib>LILJENQUIST, DAVID R.</creatorcontrib><creatorcontrib>LIU, HANQING</creatorcontrib><collection>CrossRef</collection><collection>ProQuest Health &amp; Medical Complete (Alumni)</collection><collection>Nursing &amp; Allied Health Premium</collection><jtitle>Diabetes (New York, N.Y.)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>KARINKA, SHRIDHARA ALVA</au><au>BAILEY, TIMOTHY S.</au><au>BRAZG, RONALD L.</au><au>BUDIMAN, ERWIN S.</au><au>CASTORINO, KRISTIN</au><au>CHRISTIANSEN, MARK P.</au><au>LILJENQUIST, DAVID R.</au><au>LIU, HANQING</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>910-P: Improved Accuracy of 14-Day Factory-Calibrated FreeStyle Libre System with New Glucose Algorithm</atitle><jtitle>Diabetes (New York, N.Y.)</jtitle><date>2019-06-01</date><risdate>2019</risdate><volume>68</volume><issue>Supplement_1</issue><issn>0012-1797</issn><eissn>1939-327X</eissn><abstract>We evaluated the accuracy of FreeStyle Libre 14-day system with new glucose algorithm in adults ages 18 years or older. This was a non-randomized, multi-center, single-arm study that enrolled 144 subjects with types 1 and 2 diabetes on intensive insulin therapy at 5 U.S. sites. Subjects participated in 3 clinic sessions of up to 10 hours each over the 14- day wear period to obtain comparator data. Glucose concentrations were manipulated to obtain glucose concentrations across the system dynamic range of 40-500mg/dL. System performance was evaluated in terms of percent accuracy with respect to venous plasma reference using YSI2300. Accuracy was characterized as the proportion of sensor results that were within ±15%, ± 20% of their paired YSI values for glucose levels ≥ 70 mg/dL or within ± 15 mg/dL, ± 20 mg/dL of YSI for glucose levels &lt; 70 mg/dL. Other accuracy metrics such as mean bias and mean absolute relative difference (MARD) of the FreeStyle Libre System with respect to YSI were also evaluated and are presented in Table 1. FreeStyle Libre System with new algorithm showed 92.8% of the results within ±20mg/dL/±20% of the venous plasma reference with an overall MARD of 9.2% based on the data from 144 subjects with evaluable sensors.</abstract><cop>New York</cop><pub>American Diabetes Association</pub><doi>10.2337/db19-910-P</doi></addata></record>
fulltext fulltext
identifier ISSN: 0012-1797
ispartof Diabetes (New York, N.Y.), 2019-06, Vol.68 (Supplement_1)
issn 0012-1797
1939-327X
language eng
recordid cdi_proquest_journals_2248403094
source Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; PubMed Central
subjects Accuracy
Algorithms
Diabetes
Diabetes mellitus
Glucose
Insulin
title 910-P: Improved Accuracy of 14-Day Factory-Calibrated FreeStyle Libre System with New Glucose Algorithm
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-10T07%3A18%3A12IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=910-P:%20Improved%20Accuracy%20of%2014-Day%20Factory-Calibrated%20FreeStyle%20Libre%20System%20with%20New%20Glucose%20Algorithm&rft.jtitle=Diabetes%20(New%20York,%20N.Y.)&rft.au=KARINKA,%20SHRIDHARA%20ALVA&rft.date=2019-06-01&rft.volume=68&rft.issue=Supplement_1&rft.issn=0012-1797&rft.eissn=1939-327X&rft_id=info:doi/10.2337/db19-910-P&rft_dat=%3Cproquest_cross%3E2248403094%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2248403094&rft_id=info:pmid/&rfr_iscdi=true