A Prototype of a Ring Neutron Detector

— When the half-ring of a prototype neutron detector with an inner radius of 50 mm is calibrated using a 60 Co ionizing-radiation source, its relative detection efficiency gradually decreases from ~15% at the center of the half-ring to ~5% at its ends. When calibrated using cosmic muons, the relativ...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Instruments and experimental techniques (New York) 2019-07, Vol.62 (3), p.305-308
Hauptverfasser: Baskov, V. A., Govorkov, B. B., Poliansky, V. V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:— When the half-ring of a prototype neutron detector with an inner radius of 50 mm is calibrated using a 60 Co ionizing-radiation source, its relative detection efficiency gradually decreases from ~15% at the center of the half-ring to ~5% at its ends. When calibrated using cosmic muons, the relative detection efficiency is approximately 100% over the entire detector volume. The presence of zones with reduced and increased efficiency values are associated with the intricate detector geometry.