Determination of Aniline Blue dye by Flow Injection Analysis With Home Made Valve

Flow injection unit with merging zone technique was designed for aniline blue determination spectrophotometrically. This unit consists of a valve was designed for dye injection. The scope of the study is to find the optimum conditions :volume of sample, flow rate of carrier, dead volume, dispersion...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Oriental journal of chemistry 2017-01, Vol.33 (2), p.944-950
Hauptverfasser: Saleh Farhood, Ahmed, Ahmed Mohammed Ali, Luma, Fadhel Ali, Firas
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Flow injection unit with merging zone technique was designed for aniline blue determination spectrophotometrically. This unit consists of a valve was designed for dye injection. The scope of the study is to find the optimum conditions :volume of sample, flow rate of carrier, dead volume, dispersion coefficient and reproducibility. They found 117µl, 4.3 ml/min, zero dead volume, 2.62 and high reproducibility for six times respectively. The range of aniline blue concentration by flow injection analysis was 0. 1 – 8 ppm and detection limit 0.1 with R2= 0.9952 and 0.01 – 16 ppm for spectrophotometric method with R2=0.9947 with detection limit 0.01.
ISSN:0970-020X
2231-5039
DOI:10.13005/ojc/330246