A66 THE MANY FACES OF ASTHMA IN CHILDHOOD: Systematic Bias In Respiratory Impedance Measured By Commercial Equipment Calibrated Using A Low Impedance
[...]we have measured Zrs using a more accurate non-commercial wave-tube assembly and noticed differences between Zrs measured with the wave-tube and i2M devices and respiratory resistance (Rrs) and reactance (Xrs) calculated. The i2M was calibrated according to manufacturer's instructions as o...
Gespeichert in:
Veröffentlicht in: | American journal of respiratory and critical care medicine 2017-01, Vol.195 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | [...]we have measured Zrs using a more accurate non-commercial wave-tube assembly and noticed differences between Zrs measured with the wave-tube and i2M devices and respiratory resistance (Rrs) and reactance (Xrs) calculated. The i2M was calibrated according to manufacturer's instructions as outlined above and the wave-tube calibrated using a technique in which the FMA impedance was considered as an impedance in series and removed accordingly. |
---|---|