A66 THE MANY FACES OF ASTHMA IN CHILDHOOD: Systematic Bias In Respiratory Impedance Measured By Commercial Equipment Calibrated Using A Low Impedance

[...]we have measured Zrs using a more accurate non-commercial wave-tube assembly and noticed differences between Zrs measured with the wave-tube and i2M devices and respiratory resistance (Rrs) and reactance (Xrs) calculated. The i2M was calibrated according to manufacturer's instructions as o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of respiratory and critical care medicine 2017-01, Vol.195
Hauptverfasser: Sly, P D, Shackleton, C, Czovek, D, Hantos, Z
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:[...]we have measured Zrs using a more accurate non-commercial wave-tube assembly and noticed differences between Zrs measured with the wave-tube and i2M devices and respiratory resistance (Rrs) and reactance (Xrs) calculated. The i2M was calibrated according to manufacturer's instructions as outlined above and the wave-tube calibrated using a technique in which the FMA impedance was considered as an impedance in series and removed accordingly.