Effective dose estimation for pediatric upper gastrointestinal examinations using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology
Background There is a need for updated radiation dose estimates in pediatric fluoroscopy given the routine use of new dose-saving technologies and increased radiation safety awareness in pediatric imaging. Objective To estimate effective doses for standardized pediatric upper gastrointestinal (UGI)...
Gespeichert in:
Veröffentlicht in: | Pediatric radiology 2013-09, Vol.43 (9), p.1108-1116 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
There is a need for updated radiation dose estimates in pediatric fluoroscopy given the routine use of new dose-saving technologies and increased radiation safety awareness in pediatric imaging.
Objective
To estimate effective doses for standardized pediatric upper gastrointestinal (UGI) examinations at our institute using direct dose measurement, as well as provide dose-area product (DAP) to effective dose conversion factors to be used for the estimation of UGI effective doses for boys and girls up to 10 years of age at other centers.
Materials and methods
Metal oxide semiconductor field-effect transistor (MOSFET) dosimeters were placed within four anthropomorphic phantoms representing children ≤10 years of age and exposed to mock UGI examinations using exposures much greater than used clinically to minimize measurement error. Measured effective dose was calculated using ICRP 103 weights and scaled to our institution’s standardized clinical UGI (3.6-min fluoroscopy, four spot exposures and four examination beam projections) as determined from patient logs. Results were compared to Monte Carlo simulations and related to fluoroscope-displayed DAP.
Results
Measured effective doses for standardized pediatric UGI examinations in our institute ranged from 0.35 to 0.79 mSv in girls and were 3–8% lower for boys. Simulation-derived and measured effective doses were in agreement (percentage differences 0.18). DAP-to-effective dose conversion factors ranged from 6.5
×
10
−4
mSv per Gy-cm
2
to 4.3
×
10
−3
mSv per Gy-cm
2
for girls and were similarly lower for boys.
Conclusion
Using modern fluoroscopy equipment, the effective dose associated with the UGI examination in children ≤10 years at our institute is |
---|---|