Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace
To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster. The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 device...
Gespeichert in:
Veröffentlicht in: | PloS one 2016-01, Vol.11 (1), p.e0147086-e0147086 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | e0147086 |
---|---|
container_issue | 1 |
container_start_page | e0147086 |
container_title | PloS one |
container_volume | 11 |
creator | Hua, Yanjun Xu, Zequan Qiu, Wei Wu, Qiang |
description | To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster.
The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 devices. Another observer performed additional 3 consecutive scans using the Topcon KR-1W and iTrace in the same session. About one week later, the first observer repeated the measurements using the Topcon KR-1W and iTrace. The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), J0 and J45 were analyzed. Repeatability and reproducibility of measurements were evaluated by the within-subject standard deviation (Sw), coefficient of variation (CoV), test-retest repeatability (2.77Sw), and intraclass correlation coefficient (ICC). Agreements between devices were assessed using Bland-Altman analysis and 95% limits of agreement (LoA).
Intraobserver repeatability and interobserver and intersession reproducibility of the Ks, Kf and Km showed a CoV of no more than 0.5%, a 2.77Sw of 0.70 D or less, and an ICC of no less than 0.99. However, J0 and J45 showed poor intraobserver repeatability and interobserver and intersession reproducibility (all ICCs not greater than 0.446). Statistically significant differences existed between Topcon KR-1W and IOLMaster, Topcon KR-1W and iTrace, Topcon KR-1W and Topolyzer, iTrace and Topolyzer, iTrace and IOLMaster for Ks, Kf and Km measurements (all P < 0.05). The mean differences between Topcon KR-1W, iTrace, and the other 2 devices were small. The 95% LoA were approximately 1.0 D to 1.5 D for all measurements.
The Ks, Kf and Km obtained by Topcon KR-1W and iTrace showed excellent intraobserver repeatability and interobserver and intersession reproducibility in normal eyes. The agreement between Topcon KR-1W and Topolyzer, Topcon KR-1W and IOLMaster, iTrace and Topolyzer, iTrace and IOLMaster, Topcon KR-1W and iTrace were not so good, they should not be interchangeable in clinical application. Given that the intraobserver repeatability and interobserver and intersession reproducibility of corneal astigmatism measurements obtained by Topcon KR-1W and iTrace were poor, it should be cautious that Topcon KR-1W and iTrace were applied for the preparation of toric lens implantation. |
doi_str_mv | 10.1371/journal.pone.0147086 |
format | Article |
fullrecord | <record><control><sourceid>gale_plos_</sourceid><recordid>TN_cdi_plos_journals_1755799217</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A439600441</galeid><doaj_id>oai_doaj_org_article_ebd36aff7d1a457f830b84893717c669</doaj_id><sourcerecordid>A439600441</sourcerecordid><originalsourceid>FETCH-LOGICAL-c692t-a55f3e189e205039c83afe947e0eb1c7ed214bc334b38b3e4ce729089a03e3b63</originalsourceid><addsrcrecordid>eNqNk02P0zAQhiMEYpeFf4AgEhLaPbTYceKPC1JV8VGxqKtS4Gg5zqR1lcTFToAe-O-4bXbVoD2gHBy9fuYde8YTRc8xGmPC8JuN7VyjqvHWNjBGOGWI0wfRORYkGdEEkYcn_2fRE-83CGWEU_o4OksoyxKUifPoz40DbbyxTXy5gC2oVuWmMu0uVk0RB8XZotPmqF0dxMnKAdTQtLEt46l1DagqvrG_wMWfQfnOHTZ9PM9bZRoo4nwXL-1WhxSfFiP8_WBilk5peBo9KlXl4Vm_XkRf379bTj-OrucfZtPJ9UhTkbQjlWUlAcwFhEMjIjQnqgSRMkCQY82gSHCaa0LSnPCcQKqBJQJxoRABklNyEb08-m4r62VfOS8xyzImRIJZIGZHorBqI7fO1MrtpFVGHgTrVlK51ugKJOQFoaosWYFVmrGSE5TzlIvQFKYpFcHrbZ-ty2sodKiGU9XAdLjTmLVc2Z8y9FBgjoPBZW_g7I8OfCtr4zVUlWrAdvtzU8QzTg65Xv2D3n-7nlqpcAHTlDbk1XtTOUmJoAil6T7t-B4qfAXUJrQPShP0QcDVICAwLfxuV6rzXs6-LP6fnX8bsq9P2HV4X-3a26prwzP1QzA9gtpZ7x2Ud0XGSO6n5LYacj8lsp-SEPbitEF3QbdjQf4CPAUMYQ</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1755799217</pqid></control><display><type>article</type><title>Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace</title><source>MEDLINE</source><source>DOAJ Directory of Open Access Journals</source><source>Public Library of Science (PLoS)</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><source>Free Full-Text Journals in Chemistry</source><creator>Hua, Yanjun ; Xu, Zequan ; Qiu, Wei ; Wu, Qiang</creator><contributor>Bhattacharya, Sanjoy</contributor><creatorcontrib>Hua, Yanjun ; Xu, Zequan ; Qiu, Wei ; Wu, Qiang ; Bhattacharya, Sanjoy</creatorcontrib><description>To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster.
The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 devices. Another observer performed additional 3 consecutive scans using the Topcon KR-1W and iTrace in the same session. About one week later, the first observer repeated the measurements using the Topcon KR-1W and iTrace. The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), J0 and J45 were analyzed. Repeatability and reproducibility of measurements were evaluated by the within-subject standard deviation (Sw), coefficient of variation (CoV), test-retest repeatability (2.77Sw), and intraclass correlation coefficient (ICC). Agreements between devices were assessed using Bland-Altman analysis and 95% limits of agreement (LoA).
Intraobserver repeatability and interobserver and intersession reproducibility of the Ks, Kf and Km showed a CoV of no more than 0.5%, a 2.77Sw of 0.70 D or less, and an ICC of no less than 0.99. However, J0 and J45 showed poor intraobserver repeatability and interobserver and intersession reproducibility (all ICCs not greater than 0.446). Statistically significant differences existed between Topcon KR-1W and IOLMaster, Topcon KR-1W and iTrace, Topcon KR-1W and Topolyzer, iTrace and Topolyzer, iTrace and IOLMaster for Ks, Kf and Km measurements (all P < 0.05). The mean differences between Topcon KR-1W, iTrace, and the other 2 devices were small. The 95% LoA were approximately 1.0 D to 1.5 D for all measurements.
The Ks, Kf and Km obtained by Topcon KR-1W and iTrace showed excellent intraobserver repeatability and interobserver and intersession reproducibility in normal eyes. The agreement between Topcon KR-1W and Topolyzer, Topcon KR-1W and IOLMaster, iTrace and Topolyzer, iTrace and IOLMaster, Topcon KR-1W and iTrace were not so good, they should not be interchangeable in clinical application. Given that the intraobserver repeatability and interobserver and intersession reproducibility of corneal astigmatism measurements obtained by Topcon KR-1W and iTrace were poor, it should be cautious that Topcon KR-1W and iTrace were applied for the preparation of toric lens implantation.</description><identifier>ISSN: 1932-6203</identifier><identifier>EISSN: 1932-6203</identifier><identifier>DOI: 10.1371/journal.pone.0147086</identifier><identifier>PMID: 26752059</identifier><language>eng</language><publisher>United States: Public Library of Science</publisher><subject>Adult ; Aged ; Analysis ; Astigmatism ; Automation ; Cataracts ; Coefficient of variation ; Contact lenses ; Cornea ; Cornea - physiology ; Corneal Diseases - diagnosis ; Corneal Topography - instrumentation ; Corneal Topography - methods ; Correlation coefficient ; Correlation coefficients ; Eye lens ; Female ; Healthy Volunteers ; Humans ; Implantation ; Interferometry ; Lenses, Intraocular ; Male ; Medical care ; Middle Aged ; Observer Variation ; Power measurement ; Prospective Studies ; Refractometry ; Reproducibility ; Reproducibility of Results ; Statistical analysis ; Studies ; Surgery ; Topography ; United States ; Young Adult</subject><ispartof>PloS one, 2016-01, Vol.11 (1), p.e0147086-e0147086</ispartof><rights>COPYRIGHT 2016 Public Library of Science</rights><rights>2016 Hua et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><rights>2016 Hua et al 2016 Hua et al</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c692t-a55f3e189e205039c83afe947e0eb1c7ed214bc334b38b3e4ce729089a03e3b63</citedby><cites>FETCH-LOGICAL-c692t-a55f3e189e205039c83afe947e0eb1c7ed214bc334b38b3e4ce729089a03e3b63</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC4709181/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC4709181/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,864,885,2102,2928,23866,27924,27925,53791,53793,79600,79601</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/26752059$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><contributor>Bhattacharya, Sanjoy</contributor><creatorcontrib>Hua, Yanjun</creatorcontrib><creatorcontrib>Xu, Zequan</creatorcontrib><creatorcontrib>Qiu, Wei</creatorcontrib><creatorcontrib>Wu, Qiang</creatorcontrib><title>Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace</title><title>PloS one</title><addtitle>PLoS One</addtitle><description>To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster.
The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 devices. Another observer performed additional 3 consecutive scans using the Topcon KR-1W and iTrace in the same session. About one week later, the first observer repeated the measurements using the Topcon KR-1W and iTrace. The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), J0 and J45 were analyzed. Repeatability and reproducibility of measurements were evaluated by the within-subject standard deviation (Sw), coefficient of variation (CoV), test-retest repeatability (2.77Sw), and intraclass correlation coefficient (ICC). Agreements between devices were assessed using Bland-Altman analysis and 95% limits of agreement (LoA).
Intraobserver repeatability and interobserver and intersession reproducibility of the Ks, Kf and Km showed a CoV of no more than 0.5%, a 2.77Sw of 0.70 D or less, and an ICC of no less than 0.99. However, J0 and J45 showed poor intraobserver repeatability and interobserver and intersession reproducibility (all ICCs not greater than 0.446). Statistically significant differences existed between Topcon KR-1W and IOLMaster, Topcon KR-1W and iTrace, Topcon KR-1W and Topolyzer, iTrace and Topolyzer, iTrace and IOLMaster for Ks, Kf and Km measurements (all P < 0.05). The mean differences between Topcon KR-1W, iTrace, and the other 2 devices were small. The 95% LoA were approximately 1.0 D to 1.5 D for all measurements.
The Ks, Kf and Km obtained by Topcon KR-1W and iTrace showed excellent intraobserver repeatability and interobserver and intersession reproducibility in normal eyes. The agreement between Topcon KR-1W and Topolyzer, Topcon KR-1W and IOLMaster, iTrace and Topolyzer, iTrace and IOLMaster, Topcon KR-1W and iTrace were not so good, they should not be interchangeable in clinical application. Given that the intraobserver repeatability and interobserver and intersession reproducibility of corneal astigmatism measurements obtained by Topcon KR-1W and iTrace were poor, it should be cautious that Topcon KR-1W and iTrace were applied for the preparation of toric lens implantation.</description><subject>Adult</subject><subject>Aged</subject><subject>Analysis</subject><subject>Astigmatism</subject><subject>Automation</subject><subject>Cataracts</subject><subject>Coefficient of variation</subject><subject>Contact lenses</subject><subject>Cornea</subject><subject>Cornea - physiology</subject><subject>Corneal Diseases - diagnosis</subject><subject>Corneal Topography - instrumentation</subject><subject>Corneal Topography - methods</subject><subject>Correlation coefficient</subject><subject>Correlation coefficients</subject><subject>Eye lens</subject><subject>Female</subject><subject>Healthy Volunteers</subject><subject>Humans</subject><subject>Implantation</subject><subject>Interferometry</subject><subject>Lenses, Intraocular</subject><subject>Male</subject><subject>Medical care</subject><subject>Middle Aged</subject><subject>Observer Variation</subject><subject>Power measurement</subject><subject>Prospective Studies</subject><subject>Refractometry</subject><subject>Reproducibility</subject><subject>Reproducibility of Results</subject><subject>Statistical analysis</subject><subject>Studies</subject><subject>Surgery</subject><subject>Topography</subject><subject>United States</subject><subject>Young Adult</subject><issn>1932-6203</issn><issn>1932-6203</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2016</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><sourceid>ABUWG</sourceid><sourceid>AFKRA</sourceid><sourceid>AZQEC</sourceid><sourceid>BENPR</sourceid><sourceid>CCPQU</sourceid><sourceid>DWQXO</sourceid><sourceid>GNUQQ</sourceid><sourceid>DOA</sourceid><recordid>eNqNk02P0zAQhiMEYpeFf4AgEhLaPbTYceKPC1JV8VGxqKtS4Gg5zqR1lcTFToAe-O-4bXbVoD2gHBy9fuYde8YTRc8xGmPC8JuN7VyjqvHWNjBGOGWI0wfRORYkGdEEkYcn_2fRE-83CGWEU_o4OksoyxKUifPoz40DbbyxTXy5gC2oVuWmMu0uVk0RB8XZotPmqF0dxMnKAdTQtLEt46l1DagqvrG_wMWfQfnOHTZ9PM9bZRoo4nwXL-1WhxSfFiP8_WBilk5peBo9KlXl4Vm_XkRf379bTj-OrucfZtPJ9UhTkbQjlWUlAcwFhEMjIjQnqgSRMkCQY82gSHCaa0LSnPCcQKqBJQJxoRABklNyEb08-m4r62VfOS8xyzImRIJZIGZHorBqI7fO1MrtpFVGHgTrVlK51ugKJOQFoaosWYFVmrGSE5TzlIvQFKYpFcHrbZ-ty2sodKiGU9XAdLjTmLVc2Z8y9FBgjoPBZW_g7I8OfCtr4zVUlWrAdvtzU8QzTg65Xv2D3n-7nlqpcAHTlDbk1XtTOUmJoAil6T7t-B4qfAXUJrQPShP0QcDVICAwLfxuV6rzXs6-LP6fnX8bsq9P2HV4X-3a26prwzP1QzA9gtpZ7x2Ud0XGSO6n5LYacj8lsp-SEPbitEF3QbdjQf4CPAUMYQ</recordid><startdate>20160111</startdate><enddate>20160111</enddate><creator>Hua, Yanjun</creator><creator>Xu, Zequan</creator><creator>Qiu, Wei</creator><creator>Wu, Qiang</creator><general>Public Library of Science</general><general>Public Library of Science (PLoS)</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>IOV</scope><scope>ISR</scope><scope>3V.</scope><scope>7QG</scope><scope>7QL</scope><scope>7QO</scope><scope>7RV</scope><scope>7SN</scope><scope>7SS</scope><scope>7T5</scope><scope>7TG</scope><scope>7TM</scope><scope>7U9</scope><scope>7X2</scope><scope>7X7</scope><scope>7XB</scope><scope>88E</scope><scope>8AO</scope><scope>8C1</scope><scope>8FD</scope><scope>8FE</scope><scope>8FG</scope><scope>8FH</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABJCF</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>ARAPS</scope><scope>ATCPS</scope><scope>AZQEC</scope><scope>BBNVY</scope><scope>BENPR</scope><scope>BGLVJ</scope><scope>BHPHI</scope><scope>C1K</scope><scope>CCPQU</scope><scope>D1I</scope><scope>DWQXO</scope><scope>FR3</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>GNUQQ</scope><scope>H94</scope><scope>HCIFZ</scope><scope>K9.</scope><scope>KB.</scope><scope>KB0</scope><scope>KL.</scope><scope>L6V</scope><scope>LK8</scope><scope>M0K</scope><scope>M0S</scope><scope>M1P</scope><scope>M7N</scope><scope>M7P</scope><scope>M7S</scope><scope>NAPCQ</scope><scope>P5Z</scope><scope>P62</scope><scope>P64</scope><scope>PATMY</scope><scope>PDBOC</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>PTHSS</scope><scope>PYCSY</scope><scope>RC3</scope><scope>7X8</scope><scope>5PM</scope><scope>DOA</scope></search><sort><creationdate>20160111</creationdate><title>Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace</title><author>Hua, Yanjun ; Xu, Zequan ; Qiu, Wei ; Wu, Qiang</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c692t-a55f3e189e205039c83afe947e0eb1c7ed214bc334b38b3e4ce729089a03e3b63</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2016</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Analysis</topic><topic>Astigmatism</topic><topic>Automation</topic><topic>Cataracts</topic><topic>Coefficient of variation</topic><topic>Contact lenses</topic><topic>Cornea</topic><topic>Cornea - physiology</topic><topic>Corneal Diseases - diagnosis</topic><topic>Corneal Topography - instrumentation</topic><topic>Corneal Topography - methods</topic><topic>Correlation coefficient</topic><topic>Correlation coefficients</topic><topic>Eye lens</topic><topic>Female</topic><topic>Healthy Volunteers</topic><topic>Humans</topic><topic>Implantation</topic><topic>Interferometry</topic><topic>Lenses, Intraocular</topic><topic>Male</topic><topic>Medical care</topic><topic>Middle Aged</topic><topic>Observer Variation</topic><topic>Power measurement</topic><topic>Prospective Studies</topic><topic>Refractometry</topic><topic>Reproducibility</topic><topic>Reproducibility of Results</topic><topic>Statistical analysis</topic><topic>Studies</topic><topic>Surgery</topic><topic>Topography</topic><topic>United States</topic><topic>Young Adult</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Hua, Yanjun</creatorcontrib><creatorcontrib>Xu, Zequan</creatorcontrib><creatorcontrib>Qiu, Wei</creatorcontrib><creatorcontrib>Wu, Qiang</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Gale In Context: Opposing Viewpoints</collection><collection>Gale In Context: Science</collection><collection>ProQuest Central (Corporate)</collection><collection>Animal Behavior Abstracts</collection><collection>Bacteriology Abstracts (Microbiology B)</collection><collection>Biotechnology Research Abstracts</collection><collection>Nursing & Allied Health Database</collection><collection>Ecology Abstracts</collection><collection>Entomology Abstracts (Full archive)</collection><collection>Immunology Abstracts</collection><collection>Meteorological & Geoastrophysical Abstracts</collection><collection>Nucleic Acids Abstracts</collection><collection>Virology and AIDS Abstracts</collection><collection>Agricultural Science Collection</collection><collection>Health & Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Medical Database (Alumni Edition)</collection><collection>ProQuest Pharma Collection</collection><collection>Public Health Database</collection><collection>Technology Research Database</collection><collection>ProQuest SciTech Collection</collection><collection>ProQuest Technology Collection</collection><collection>ProQuest Natural Science Collection</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>Materials Science & Engineering Collection</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>Advanced Technologies & Aerospace Collection</collection><collection>Agricultural & Environmental Science Collection</collection><collection>ProQuest Central Essentials</collection><collection>Biological Science Collection</collection><collection>ProQuest Central</collection><collection>Technology Collection (ProQuest)</collection><collection>Natural Science Collection (ProQuest)</collection><collection>Environmental Sciences and Pollution Management</collection><collection>ProQuest One Community College</collection><collection>ProQuest Materials Science Collection</collection><collection>ProQuest Central Korea</collection><collection>Engineering Research Database</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Central Student</collection><collection>AIDS and Cancer Research Abstracts</collection><collection>SciTech Premium Collection</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Materials Science Database</collection><collection>Nursing & Allied Health Database (Alumni Edition)</collection><collection>Meteorological & Geoastrophysical Abstracts - Academic</collection><collection>ProQuest Engineering Collection</collection><collection>ProQuest Biological Science Collection</collection><collection>Agricultural Science Database</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>Medical Database</collection><collection>Algology Mycology and Protozoology Abstracts (Microbiology C)</collection><collection>Biological Science Database</collection><collection>Engineering Database</collection><collection>Nursing & Allied Health Premium</collection><collection>Advanced Technologies & Aerospace Database</collection><collection>ProQuest Advanced Technologies & Aerospace Collection</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>Environmental Science Database</collection><collection>Materials Science Collection</collection><collection>Publicly Available Content Database</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>Engineering Collection</collection><collection>Environmental Science Collection</collection><collection>Genetics Abstracts</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>DOAJ Directory of Open Access Journals</collection><jtitle>PloS one</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Hua, Yanjun</au><au>Xu, Zequan</au><au>Qiu, Wei</au><au>Wu, Qiang</au><au>Bhattacharya, Sanjoy</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace</atitle><jtitle>PloS one</jtitle><addtitle>PLoS One</addtitle><date>2016-01-11</date><risdate>2016</risdate><volume>11</volume><issue>1</issue><spage>e0147086</spage><epage>e0147086</epage><pages>e0147086-e0147086</pages><issn>1932-6203</issn><eissn>1932-6203</eissn><abstract>To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster.
The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 devices. Another observer performed additional 3 consecutive scans using the Topcon KR-1W and iTrace in the same session. About one week later, the first observer repeated the measurements using the Topcon KR-1W and iTrace. The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), J0 and J45 were analyzed. Repeatability and reproducibility of measurements were evaluated by the within-subject standard deviation (Sw), coefficient of variation (CoV), test-retest repeatability (2.77Sw), and intraclass correlation coefficient (ICC). Agreements between devices were assessed using Bland-Altman analysis and 95% limits of agreement (LoA).
Intraobserver repeatability and interobserver and intersession reproducibility of the Ks, Kf and Km showed a CoV of no more than 0.5%, a 2.77Sw of 0.70 D or less, and an ICC of no less than 0.99. However, J0 and J45 showed poor intraobserver repeatability and interobserver and intersession reproducibility (all ICCs not greater than 0.446). Statistically significant differences existed between Topcon KR-1W and IOLMaster, Topcon KR-1W and iTrace, Topcon KR-1W and Topolyzer, iTrace and Topolyzer, iTrace and IOLMaster for Ks, Kf and Km measurements (all P < 0.05). The mean differences between Topcon KR-1W, iTrace, and the other 2 devices were small. The 95% LoA were approximately 1.0 D to 1.5 D for all measurements.
The Ks, Kf and Km obtained by Topcon KR-1W and iTrace showed excellent intraobserver repeatability and interobserver and intersession reproducibility in normal eyes. The agreement between Topcon KR-1W and Topolyzer, Topcon KR-1W and IOLMaster, iTrace and Topolyzer, iTrace and IOLMaster, Topcon KR-1W and iTrace were not so good, they should not be interchangeable in clinical application. Given that the intraobserver repeatability and interobserver and intersession reproducibility of corneal astigmatism measurements obtained by Topcon KR-1W and iTrace were poor, it should be cautious that Topcon KR-1W and iTrace were applied for the preparation of toric lens implantation.</abstract><cop>United States</cop><pub>Public Library of Science</pub><pmid>26752059</pmid><doi>10.1371/journal.pone.0147086</doi><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1932-6203 |
ispartof | PloS one, 2016-01, Vol.11 (1), p.e0147086-e0147086 |
issn | 1932-6203 1932-6203 |
language | eng |
recordid | cdi_plos_journals_1755799217 |
source | MEDLINE; DOAJ Directory of Open Access Journals; Public Library of Science (PLoS); EZB-FREE-00999 freely available EZB journals; PubMed Central; Free Full-Text Journals in Chemistry |
subjects | Adult Aged Analysis Astigmatism Automation Cataracts Coefficient of variation Contact lenses Cornea Cornea - physiology Corneal Diseases - diagnosis Corneal Topography - instrumentation Corneal Topography - methods Correlation coefficient Correlation coefficients Eye lens Female Healthy Volunteers Humans Implantation Interferometry Lenses, Intraocular Male Medical care Middle Aged Observer Variation Power measurement Prospective Studies Refractometry Reproducibility Reproducibility of Results Statistical analysis Studies Surgery Topography United States Young Adult |
title | Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T08%3A03%3A51IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_plos_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Precision%20(Repeatability%20and%20Reproducibility)%20and%20Agreement%20of%20Corneal%20Power%20Measurements%20Obtained%20by%20Topcon%20KR-1W%20and%20iTrace&rft.jtitle=PloS%20one&rft.au=Hua,%20Yanjun&rft.date=2016-01-11&rft.volume=11&rft.issue=1&rft.spage=e0147086&rft.epage=e0147086&rft.pages=e0147086-e0147086&rft.issn=1932-6203&rft.eissn=1932-6203&rft_id=info:doi/10.1371/journal.pone.0147086&rft_dat=%3Cgale_plos_%3EA439600441%3C/gale_plos_%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1755799217&rft_id=info:pmid/26752059&rft_galeid=A439600441&rft_doaj_id=oai_doaj_org_article_ebd36aff7d1a457f830b84893717c669&rfr_iscdi=true |