A six-port reflectometer calibration using Schottky diodes operating in AC detection mode
This works presents a six-port reflectometer in W frequency band (75-110 GHz) using Schottky diodes as power detectors. A linearization procedure and an AC detection technique are used to improve the measurement accuracy. In addition, a calibration procedure requiring two standards, a matched wavegu...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 1993-04, Vol.42 (2), p.505-510 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This works presents a six-port reflectometer in W frequency band (75-110 GHz) using Schottky diodes as power detectors. A linearization procedure and an AC detection technique are used to improve the measurement accuracy. In addition, a calibration procedure requiring two standards, a matched waveguide and a short circuit, is described. A comparison between experimental results obtained by the six-port reflectometer and other automatic network analyzers is presented.< > |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/19.278612 |