A theoretical study of the robustness of the isovolume calibration method for a two-compartment model of breathing, based on an analysis of the connected cylinders model

The use of the isovolume manoeuvre method as a calibration technique for respiratory monitoring instrumentation that detects the movement of the ribcage and the abdominal wall is analysed based on a model of two connected cylinders whose radii and heights may vary, and evidence is presented which su...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 1997-02, Vol.42 (2), p.283-291
1. Verfasser: Augousti, A T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The use of the isovolume manoeuvre method as a calibration technique for respiratory monitoring instrumentation that detects the movement of the ribcage and the abdominal wall is analysed based on a model of two connected cylinders whose radii and heights may vary, and evidence is presented which suggests that this calibration method is robust in most circumstances. Some possible functional forms relating the variations in cylinder radius and height are examined, and methods for obtaining calibration constants based on these functional forms, purely from measurements of variations in the cylinder radius, are presented.
ISSN:0031-9155
1361-6560
DOI:10.1088/0031-9155/42/2/002