Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound

Objective: During application of a distal coronary bypass connector, we employed 13 MHz epicardial ultrasound to evaluate quantitative caliper measurements for vessel size matching and to assess anastomosis quality after connector deployment. Methods: Two S2AS connector anastomoses were constructed...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of cardio-thoracic surgery 2005-12, Vol.28 (6), p.833-837
Hauptverfasser: Budde, Ricardo P.J, Suyker, Willem J.L, Suyker, Paul T.W, Verlaan, Cees W.J, Meijer, Rudy, Borst, Cornelius, Gründeman, Paul F
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 837
container_issue 6
container_start_page 833
container_title European journal of cardio-thoracic surgery
container_volume 28
creator Budde, Ricardo P.J
Suyker, Willem J.L
Suyker, Paul T.W
Verlaan, Cees W.J
Meijer, Rudy
Borst, Cornelius
Gründeman, Paul F
description Objective: During application of a distal coronary bypass connector, we employed 13 MHz epicardial ultrasound to evaluate quantitative caliper measurements for vessel size matching and to assess anastomosis quality after connector deployment. Methods: Two S2AS connector anastomoses were constructed on ex vivo pressure-perfused porcine hearts. Epicardial ultrasound measurements of the connector ring and anastomosis were compared to intravascular ultrasound measurement and cast dimensions. In 21 pigs, anastomotic sites with internal diameter of 2.25–3.0 mm (internal mammary artery, IMA) and 1.8–2.2 mm (left anterior descending coronary artery, LAD) were selected using external caliper and epicardial ultrasound measurements. Anastomoses were visualized and assessed intraoperatively (beating heart, n = 21) and at 3 and 6 months’ follow-up (explanted heart, n = 10 each). Results: Epicardial ultrasound underestimated connector dimension by ≤5% versus intravascular ultrasound and deviated ≤13% from cast dimensions for other anastomotic measurements. Caliper estimates of internal IMA and LAD diameter differed from ultrasound by −3 ± 6% and −2 ± 7% (mean ± SD), respectively. Intraoperatively, the anastomotic orifice was flawless in all animals. It remained fully patent at 3 and 6 months by ultrasound, which was confirmed by histology. The connector to LAD percentage diameter stenosis changed from −12 ± 5% intraoperatively to −1 ± 7% at 3 months and from −5 ± 6% intraoperatively to −16 ± 13% at 6 months, in the growing pig model. Conclusions: In the pig, external caliper measurements provided a reliable quantitative estimate of inner graft and coronary diameter for connector size matching. Epicardial 13 MHz ultrasound is a promising method to assess coronary anastomosis quality even when connector metal is present.
doi_str_mv 10.1016/j.ejcts.2005.08.020
format Article
fullrecord <record><control><sourceid>oup_pasca</sourceid><recordid>TN_cdi_pascalfrancis_primary_17334598</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><oup_id>10.1016/j.ejcts.2005.08.020</oup_id><sourcerecordid>10.1016/j.ejcts.2005.08.020</sourcerecordid><originalsourceid>FETCH-LOGICAL-i1078-a3e2ca76d8c6c77568a491771b490529b947985166300c8e2f5fbe4f7dd5c9163</originalsourceid><addsrcrecordid>eNpFkE1Lw0AQQIMoWKu_wMtePCbO7ma_jrVYK1RU6kfpZdlsNrAxTUo2AeuvN7WipxmY9-bwougSQ4IB8-sycaXtQkIAWAIyAQJH0QhLQWNB09XxsAOGWKgUTqOzEEoA4JSIUfT-3JvKdztkQnAhbFzdoaZAuQ-dqdCSTJbINnXtbNe0yNQmdM2mCT6gbIcwRQ_zL-S23po29wPfV11rQtPX-Xl0UpgquIvfOY5eZ7cv03m8eLy7n04WsccgZGyoI9YInkvLrRCMS5MqLATOUgWMqEylQkmGOacAVjpSsCJzaSHynFmFOR1HV4e_WxOsqYrW1NYHvW39xrQ7jQWlKVNy4JID1_Tb_yvofT5d6p98ep9Pg9RDvkGID8JQwn3-Kab90FxQwfR8tdbq7elGrKczTeg3bdVzMA</addsrcrecordid><sourcetype>Index Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound</title><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>Oxford University Press Journals All Titles (1996-Current)</source><creator>Budde, Ricardo P.J ; Suyker, Willem J.L ; Suyker, Paul T.W ; Verlaan, Cees W.J ; Meijer, Rudy ; Borst, Cornelius ; Gründeman, Paul F</creator><creatorcontrib>Budde, Ricardo P.J ; Suyker, Willem J.L ; Suyker, Paul T.W ; Verlaan, Cees W.J ; Meijer, Rudy ; Borst, Cornelius ; Gründeman, Paul F</creatorcontrib><description>Objective: During application of a distal coronary bypass connector, we employed 13 MHz epicardial ultrasound to evaluate quantitative caliper measurements for vessel size matching and to assess anastomosis quality after connector deployment. Methods: Two S2AS connector anastomoses were constructed on ex vivo pressure-perfused porcine hearts. Epicardial ultrasound measurements of the connector ring and anastomosis were compared to intravascular ultrasound measurement and cast dimensions. In 21 pigs, anastomotic sites with internal diameter of 2.25–3.0 mm (internal mammary artery, IMA) and 1.8–2.2 mm (left anterior descending coronary artery, LAD) were selected using external caliper and epicardial ultrasound measurements. Anastomoses were visualized and assessed intraoperatively (beating heart, n = 21) and at 3 and 6 months’ follow-up (explanted heart, n = 10 each). Results: Epicardial ultrasound underestimated connector dimension by ≤5% versus intravascular ultrasound and deviated ≤13% from cast dimensions for other anastomotic measurements. Caliper estimates of internal IMA and LAD diameter differed from ultrasound by −3 ± 6% and −2 ± 7% (mean ± SD), respectively. Intraoperatively, the anastomotic orifice was flawless in all animals. It remained fully patent at 3 and 6 months by ultrasound, which was confirmed by histology. The connector to LAD percentage diameter stenosis changed from −12 ± 5% intraoperatively to −1 ± 7% at 3 months and from −5 ± 6% intraoperatively to −16 ± 13% at 6 months, in the growing pig model. Conclusions: In the pig, external caliper measurements provided a reliable quantitative estimate of inner graft and coronary diameter for connector size matching. Epicardial 13 MHz ultrasound is a promising method to assess coronary anastomosis quality even when connector metal is present.</description><identifier>ISSN: 1010-7940</identifier><identifier>EISSN: 1873-734X</identifier><identifier>DOI: 10.1016/j.ejcts.2005.08.020</identifier><identifier>CODEN: EJCSE7</identifier><language>eng</language><publisher>Amsterdam: Elsevier Science B.V</publisher><subject>Anastomosis ; Biological and medical sciences ; Cardiology. Vascular system ; Coronary artery bypass grafting ; Coronary heart disease ; Echocardiography ; Heart ; Medical sciences ; Minimally invasive ; Off-pump ; Pneumology ; Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases ; Surgery of the heart ; Ultrasound</subject><ispartof>European journal of cardio-thoracic surgery, 2005-12, Vol.28 (6), p.833-837</ispartof><rights>Elsevier B.V. © 2005 Elsevier B.V. All rights reserved. 2005</rights><rights>2006 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27922,27923</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=17334598$$DView record in Pascal Francis$$Hfree_for_read</backlink></links><search><creatorcontrib>Budde, Ricardo P.J</creatorcontrib><creatorcontrib>Suyker, Willem J.L</creatorcontrib><creatorcontrib>Suyker, Paul T.W</creatorcontrib><creatorcontrib>Verlaan, Cees W.J</creatorcontrib><creatorcontrib>Meijer, Rudy</creatorcontrib><creatorcontrib>Borst, Cornelius</creatorcontrib><creatorcontrib>Gründeman, Paul F</creatorcontrib><title>Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound</title><title>European journal of cardio-thoracic surgery</title><addtitle>Eur J Cardiothorac Surg</addtitle><addtitle>Eur J Cardiothorac Surg</addtitle><description>Objective: During application of a distal coronary bypass connector, we employed 13 MHz epicardial ultrasound to evaluate quantitative caliper measurements for vessel size matching and to assess anastomosis quality after connector deployment. Methods: Two S2AS connector anastomoses were constructed on ex vivo pressure-perfused porcine hearts. Epicardial ultrasound measurements of the connector ring and anastomosis were compared to intravascular ultrasound measurement and cast dimensions. In 21 pigs, anastomotic sites with internal diameter of 2.25–3.0 mm (internal mammary artery, IMA) and 1.8–2.2 mm (left anterior descending coronary artery, LAD) were selected using external caliper and epicardial ultrasound measurements. Anastomoses were visualized and assessed intraoperatively (beating heart, n = 21) and at 3 and 6 months’ follow-up (explanted heart, n = 10 each). Results: Epicardial ultrasound underestimated connector dimension by ≤5% versus intravascular ultrasound and deviated ≤13% from cast dimensions for other anastomotic measurements. Caliper estimates of internal IMA and LAD diameter differed from ultrasound by −3 ± 6% and −2 ± 7% (mean ± SD), respectively. Intraoperatively, the anastomotic orifice was flawless in all animals. It remained fully patent at 3 and 6 months by ultrasound, which was confirmed by histology. The connector to LAD percentage diameter stenosis changed from −12 ± 5% intraoperatively to −1 ± 7% at 3 months and from −5 ± 6% intraoperatively to −16 ± 13% at 6 months, in the growing pig model. Conclusions: In the pig, external caliper measurements provided a reliable quantitative estimate of inner graft and coronary diameter for connector size matching. Epicardial 13 MHz ultrasound is a promising method to assess coronary anastomosis quality even when connector metal is present.</description><subject>Anastomosis</subject><subject>Biological and medical sciences</subject><subject>Cardiology. Vascular system</subject><subject>Coronary artery bypass grafting</subject><subject>Coronary heart disease</subject><subject>Echocardiography</subject><subject>Heart</subject><subject>Medical sciences</subject><subject>Minimally invasive</subject><subject>Off-pump</subject><subject>Pneumology</subject><subject>Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</subject><subject>Surgery of the heart</subject><subject>Ultrasound</subject><issn>1010-7940</issn><issn>1873-734X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><recordid>eNpFkE1Lw0AQQIMoWKu_wMtePCbO7ma_jrVYK1RU6kfpZdlsNrAxTUo2AeuvN7WipxmY9-bwougSQ4IB8-sycaXtQkIAWAIyAQJH0QhLQWNB09XxsAOGWKgUTqOzEEoA4JSIUfT-3JvKdztkQnAhbFzdoaZAuQ-dqdCSTJbINnXtbNe0yNQmdM2mCT6gbIcwRQ_zL-S23po29wPfV11rQtPX-Xl0UpgquIvfOY5eZ7cv03m8eLy7n04WsccgZGyoI9YInkvLrRCMS5MqLATOUgWMqEylQkmGOacAVjpSsCJzaSHynFmFOR1HV4e_WxOsqYrW1NYHvW39xrQ7jQWlKVNy4JID1_Tb_yvofT5d6p98ep9Pg9RDvkGID8JQwn3-Kab90FxQwfR8tdbq7elGrKczTeg3bdVzMA</recordid><startdate>200512</startdate><enddate>200512</enddate><creator>Budde, Ricardo P.J</creator><creator>Suyker, Willem J.L</creator><creator>Suyker, Paul T.W</creator><creator>Verlaan, Cees W.J</creator><creator>Meijer, Rudy</creator><creator>Borst, Cornelius</creator><creator>Gründeman, Paul F</creator><general>Elsevier Science B.V</general><general>Elsevier Science</general><scope>BSCLL</scope><scope>IQODW</scope></search><sort><creationdate>200512</creationdate><title>Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound</title><author>Budde, Ricardo P.J ; Suyker, Willem J.L ; Suyker, Paul T.W ; Verlaan, Cees W.J ; Meijer, Rudy ; Borst, Cornelius ; Gründeman, Paul F</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-i1078-a3e2ca76d8c6c77568a491771b490529b947985166300c8e2f5fbe4f7dd5c9163</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>Anastomosis</topic><topic>Biological and medical sciences</topic><topic>Cardiology. Vascular system</topic><topic>Coronary artery bypass grafting</topic><topic>Coronary heart disease</topic><topic>Echocardiography</topic><topic>Heart</topic><topic>Medical sciences</topic><topic>Minimally invasive</topic><topic>Off-pump</topic><topic>Pneumology</topic><topic>Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</topic><topic>Surgery of the heart</topic><topic>Ultrasound</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Budde, Ricardo P.J</creatorcontrib><creatorcontrib>Suyker, Willem J.L</creatorcontrib><creatorcontrib>Suyker, Paul T.W</creatorcontrib><creatorcontrib>Verlaan, Cees W.J</creatorcontrib><creatorcontrib>Meijer, Rudy</creatorcontrib><creatorcontrib>Borst, Cornelius</creatorcontrib><creatorcontrib>Gründeman, Paul F</creatorcontrib><collection>Istex</collection><collection>Pascal-Francis</collection><jtitle>European journal of cardio-thoracic surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Budde, Ricardo P.J</au><au>Suyker, Willem J.L</au><au>Suyker, Paul T.W</au><au>Verlaan, Cees W.J</au><au>Meijer, Rudy</au><au>Borst, Cornelius</au><au>Gründeman, Paul F</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound</atitle><jtitle>European journal of cardio-thoracic surgery</jtitle><stitle>Eur J Cardiothorac Surg</stitle><addtitle>Eur J Cardiothorac Surg</addtitle><date>2005-12</date><risdate>2005</risdate><volume>28</volume><issue>6</issue><spage>833</spage><epage>837</epage><pages>833-837</pages><issn>1010-7940</issn><eissn>1873-734X</eissn><coden>EJCSE7</coden><abstract>Objective: During application of a distal coronary bypass connector, we employed 13 MHz epicardial ultrasound to evaluate quantitative caliper measurements for vessel size matching and to assess anastomosis quality after connector deployment. Methods: Two S2AS connector anastomoses were constructed on ex vivo pressure-perfused porcine hearts. Epicardial ultrasound measurements of the connector ring and anastomosis were compared to intravascular ultrasound measurement and cast dimensions. In 21 pigs, anastomotic sites with internal diameter of 2.25–3.0 mm (internal mammary artery, IMA) and 1.8–2.2 mm (left anterior descending coronary artery, LAD) were selected using external caliper and epicardial ultrasound measurements. Anastomoses were visualized and assessed intraoperatively (beating heart, n = 21) and at 3 and 6 months’ follow-up (explanted heart, n = 10 each). Results: Epicardial ultrasound underestimated connector dimension by ≤5% versus intravascular ultrasound and deviated ≤13% from cast dimensions for other anastomotic measurements. Caliper estimates of internal IMA and LAD diameter differed from ultrasound by −3 ± 6% and −2 ± 7% (mean ± SD), respectively. Intraoperatively, the anastomotic orifice was flawless in all animals. It remained fully patent at 3 and 6 months by ultrasound, which was confirmed by histology. The connector to LAD percentage diameter stenosis changed from −12 ± 5% intraoperatively to −1 ± 7% at 3 months and from −5 ± 6% intraoperatively to −16 ± 13% at 6 months, in the growing pig model. Conclusions: In the pig, external caliper measurements provided a reliable quantitative estimate of inner graft and coronary diameter for connector size matching. Epicardial 13 MHz ultrasound is a promising method to assess coronary anastomosis quality even when connector metal is present.</abstract><cop>Amsterdam</cop><pub>Elsevier Science B.V</pub><doi>10.1016/j.ejcts.2005.08.020</doi><tpages>5</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1010-7940
ispartof European journal of cardio-thoracic surgery, 2005-12, Vol.28 (6), p.833-837
issn 1010-7940
1873-734X
language eng
recordid cdi_pascalfrancis_primary_17334598
source Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; Oxford University Press Journals All Titles (1996-Current)
subjects Anastomosis
Biological and medical sciences
Cardiology. Vascular system
Coronary artery bypass grafting
Coronary heart disease
Echocardiography
Heart
Medical sciences
Minimally invasive
Off-pump
Pneumology
Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases
Surgery of the heart
Ultrasound
title Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-09T20%3A48%3A25IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-oup_pasca&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Quality%20assessment%20of%20distal%20S2AS%20connector%20anastomosis%20by%2013%20MHz%20epicardial%20ultrasound&rft.jtitle=European%20journal%20of%20cardio-thoracic%20surgery&rft.au=Budde,%20Ricardo%20P.J&rft.date=2005-12&rft.volume=28&rft.issue=6&rft.spage=833&rft.epage=837&rft.pages=833-837&rft.issn=1010-7940&rft.eissn=1873-734X&rft.coden=EJCSE7&rft_id=info:doi/10.1016/j.ejcts.2005.08.020&rft_dat=%3Coup_pasca%3E10.1016/j.ejcts.2005.08.020%3C/oup_pasca%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rft_oup_id=10.1016/j.ejcts.2005.08.020&rfr_iscdi=true