Measurements of electrode position inside the cochlea for different cochlear implant systems

Conclusions This study demonstrates that the exact location of an electrode inside the cochlea needs to be assessed using two complementary measures, namely the length and angle of insertion, both of which are mandatory if one wants to prevent erroneous outcomes. Knowledge of the contact position ma...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta oto-laryngologica 2005-05, Vol.125 (5), p.474-480
Hauptverfasser: Kós, Maria-Izabel, Boëx, Colette, Sigrist, Alain, Guyot, Jean-Philippe, Pelizzone, Marco
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Conclusions This study demonstrates that the exact location of an electrode inside the cochlea needs to be assessed using two complementary measures, namely the length and angle of insertion, both of which are mandatory if one wants to prevent erroneous outcomes. Knowledge of the contact position may become very useful when tuning a cochlear implant processor in a patient with contralateral residual hearing, or in cases of binaural implants. Objective Multichannel cochlear implants restore useful hearing to deaf patients. However, several types of intracochlear electrodes are presently available, each featuring a specific technology or design. The aim of this study was to determine precisely the intracochlear position of the contacts for different electrode arrays. Material and methods Electrode array insertions were estimated using special radiographs. A total of 26 cochlear implantations were included in the study: 6 Ineraid™; 5 Clarion™ HiFocus I; 11 Clarion HiFocus II; and 4 Med-El™ Combi40+. In each case, a measurable reference or marker ring placed close to the round window (within 2 mm) could be identified. Insertion lengths and angles were measured and then plotted on a graph based on 3D reconstructions. Results Both Clarion HiFocus I and II electrode arrays were found to be placed close to the inner wall of the cochlea. Ineraid and Med-El Combi40+ electrode arrays were both placed close to the organ of Corti, the Med-El Combi40+ arrays demonstrating the deepest insertions overall. In spite of marked differences in the positions of the contacts, we did not find any correlation with speech perception performance for the different types of implants studied.
ISSN:0963-7486
0001-6489
1465-3478
1651-2251
DOI:10.1080/00016480510039995