Charge-Injection-Device Performance in the High-Energy-Neutron Environment of Laser-Fusion Experiments

Charge-injection devices (CIDs) are being used to image x rays in laser-fusion experiments on the University of Rochester’s OMEGA Laser System. The CID cameras are routinely used up to the maximum neutron yields generated (~10^14 DT). The detectors are deployed in x-ray pinhole cameras and Kirkpatri...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2010-10, Vol.81 (10)
Hauptverfasser: Marshall, F.J., DeHaas, T., Glebov, V.Yu
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Charge-injection devices (CIDs) are being used to image x rays in laser-fusion experiments on the University of Rochester’s OMEGA Laser System. The CID cameras are routinely used up to the maximum neutron yields generated (~10^14 DT). The detectors are deployed in x-ray pinhole cameras and Kirkpatrick–Baez microscopes. The neutron fluences ranged from ~10^7 to ~10^9 neutrons/cm^2 and useful x-ray images were obtained even at the highest fluences. It is intended to use CID cameras at the National Ignition Facility (NIF) as a supporting means of recording x-ray images. The results of this work predict that x-ray images should be obtainable on the NIF at yields up to ~10^15, depending on distance and shielding.