It is often necessary to estimate the dose of radiation to a fetus from a series of CT scans. To assist in making this process easier and more accurate, we measured fetal doses in an adult anthropomorphic phantom for four CT scanners: Picker 1200, Siemens DRH, GE 9800, and GE 8800. Measurements were...
Gespeichert in:
Veröffentlicht in: | American journal of roentgenology (1976) 1990-01, Vol.154 (1), p.185-190 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | It is often necessary to estimate the dose of radiation to a fetus from a series of CT scans. To assist in making this process easier and more accurate, we measured fetal doses in an adult anthropomorphic phantom for four CT scanners: Picker 1200, Siemens DRH, GE 9800, and GE 8800. Measurements were made at four kilovoltages (100, 120, 130, and 140 kVp), for 2-, 5-, 8-, and 10-mm thicknesses, for two scanning angles (360 degrees and 398 degrees) and for two patient orientations (prone and supine). The fetal-dose estimates are based on the CT dose index measured by using a pencil ionization chamber at the center position of a 16-cm-diameter cylindrical phantom. Comparison with data from other investigators shows reasonable agreement when phantom differences, X-ray tube output, and equipment calibration are considered. |
---|---|