Tannic Acid: a Novel Calibrator for Facile and Accurate Mass Measurement of Electrospray Ionization Mass Spectrometry
Accurate mass calibration is beneficial to the identification of the unknown compounds quickly and accurately. The ESI mass spectrum of tannic acid (TA) tends to a normal distribution of the cluster ion peaks in m/z range from 371.0368 to 1739.1169. Based on the interesting result, we reported the u...
Gespeichert in:
Veröffentlicht in: | Journal of the American Society for Mass Spectrometry 2019-08, Vol.30 (8), p.1545-1549 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Accurate mass calibration is beneficial to the identification of the unknown compounds quickly and accurately. The ESI mass spectrum of tannic acid (TA) tends to a normal distribution of the cluster ion peaks in
m/z
range from 371.0368 to 1739.1169. Based on the interesting result, we reported the use of TA, a natural plant polyphenol, as a novel calibrator for electrospray ionization mass spectrometry (ESI MS), which has the following three advantages, including (1) easy preparation, (2) the calibration range of
m/z
200~2000, and (3) the calibration error is around 3.00 ppm in positive ion mode, which is less than the use of sodium formate (SF) and Prod #88323 calibrators. This TA calibrator has great potential for the wide applications in biological, chemical, and pharmacal analysis. |
---|---|
ISSN: | 1044-0305 1879-1123 |
DOI: | 10.1007/s13361-019-02211-9 |