Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator

A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Fitch, F. Moran, Martinez-Davalos, A., Garcia-Garduno, O. A.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue 1
container_start_page
container_title
container_volume 1310
creator Fitch, F. Moran
Martinez-Davalos, A.
Garcia-Garduno, O. A.
description A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributions using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.
doi_str_mv 10.1063/1.3531582
format Conference Proceeding
fullrecord <record><control><sourceid>osti</sourceid><recordid>TN_cdi_osti_scitechconnect_21506964</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>21506964</sourcerecordid><originalsourceid>FETCH-LOGICAL-c223t-6c3578cfbaa47a590a7a5ab1484889f39a5903a656bb7d3a5fea526595daec5e3</originalsourceid><addsrcrecordid>eNotjEtLxDAURoMoWEcX_oOA64559OaxLONroMWFD9wNt2mqkU4CTVz47x3RzTlw-PgIueRszZmS13wtQXIw4ohUHIDXWnF1TCrGbFOLRr6dkrOcPxkTVmtTkf4m5bD3ZfmmaaJIn_Y4z3Ubw8F0uyw4BiwhRdqn0c_0K4f4fpgp2r_SLkSPC22d87NfsKTlnJxMOGd_8e8Vebm7fd481N3j_XbTdrUTQpZaOQnauGlAbDSCZXggDrwxjTF2kva3SVSghkGPEmHyCEKBhRG9Ay9X5OrvN-USdtmF4t2HSzF6V3aCA1NWNfIHCYVNrg</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>conference_proceeding</recordtype></control><display><type>conference_proceeding</type><title>Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator</title><source>American Institute of Physics (AIP) Journals</source><creator>Fitch, F. Moran ; Martinez-Davalos, A. ; Garcia-Garduno, O. A.</creator><creatorcontrib>Fitch, F. Moran ; Martinez-Davalos, A. ; Garcia-Garduno, O. A.</creatorcontrib><description>A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributions using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.</description><identifier>ISSN: 0094-243X</identifier><identifier>EISSN: 1551-7616</identifier><identifier>DOI: 10.1063/1.3531582</identifier><language>eng</language><publisher>United States</publisher><subject>ACCELERATORS ; ANIMALS ; BODY ; COLLIMATORS ; COMPARATIVE EVALUATIONS ; CONFIGURATION ; CYLINDRICAL CONFIGURATION ; DIFFERENTIAL THERMAL ANALYSIS ; DOSIMETRY ; EVALUATION ; IRRADIATION ; ISODOSE CURVES ; LINEAR ACCELERATORS ; LUNGS ; MAMMALS ; MEDICINE ; MOCKUP ; NUCLEAR MEDICINE ; ORGANS ; PHANTOMS ; RADIATION DOSE DISTRIBUTIONS ; RADIOLOGY ; RADIOLOGY AND NUCLEAR MEDICINE ; RADIOTHERAPY ; RATS ; RESPIRATORY SYSTEM ; RODENTS ; SIMULATION ; STRUCTURAL MODELS ; THERAPY ; THERMAL ANALYSIS ; TRANSMISSION ; VERTEBRATES</subject><ispartof>AIP conference proceedings, 2010, Vol.1310 (1)</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c223t-6c3578cfbaa47a590a7a5ab1484889f39a5903a656bb7d3a5fea526595daec5e3</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,27924,27925</link.rule.ids><backlink>$$Uhttps://www.osti.gov/biblio/21506964$$D View this record in Osti.gov$$Hfree_for_read</backlink></links><search><creatorcontrib>Fitch, F. Moran</creatorcontrib><creatorcontrib>Martinez-Davalos, A.</creatorcontrib><creatorcontrib>Garcia-Garduno, O. A.</creatorcontrib><title>Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator</title><title>AIP conference proceedings</title><description>A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributions using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.</description><subject>ACCELERATORS</subject><subject>ANIMALS</subject><subject>BODY</subject><subject>COLLIMATORS</subject><subject>COMPARATIVE EVALUATIONS</subject><subject>CONFIGURATION</subject><subject>CYLINDRICAL CONFIGURATION</subject><subject>DIFFERENTIAL THERMAL ANALYSIS</subject><subject>DOSIMETRY</subject><subject>EVALUATION</subject><subject>IRRADIATION</subject><subject>ISODOSE CURVES</subject><subject>LINEAR ACCELERATORS</subject><subject>LUNGS</subject><subject>MAMMALS</subject><subject>MEDICINE</subject><subject>MOCKUP</subject><subject>NUCLEAR MEDICINE</subject><subject>ORGANS</subject><subject>PHANTOMS</subject><subject>RADIATION DOSE DISTRIBUTIONS</subject><subject>RADIOLOGY</subject><subject>RADIOLOGY AND NUCLEAR MEDICINE</subject><subject>RADIOTHERAPY</subject><subject>RATS</subject><subject>RESPIRATORY SYSTEM</subject><subject>RODENTS</subject><subject>SIMULATION</subject><subject>STRUCTURAL MODELS</subject><subject>THERAPY</subject><subject>THERMAL ANALYSIS</subject><subject>TRANSMISSION</subject><subject>VERTEBRATES</subject><issn>0094-243X</issn><issn>1551-7616</issn><fulltext>true</fulltext><rsrctype>conference_proceeding</rsrctype><creationdate>2010</creationdate><recordtype>conference_proceeding</recordtype><recordid>eNotjEtLxDAURoMoWEcX_oOA64559OaxLONroMWFD9wNt2mqkU4CTVz47x3RzTlw-PgIueRszZmS13wtQXIw4ohUHIDXWnF1TCrGbFOLRr6dkrOcPxkTVmtTkf4m5bD3ZfmmaaJIn_Y4z3Ubw8F0uyw4BiwhRdqn0c_0K4f4fpgp2r_SLkSPC22d87NfsKTlnJxMOGd_8e8Vebm7fd481N3j_XbTdrUTQpZaOQnauGlAbDSCZXggDrwxjTF2kva3SVSghkGPEmHyCEKBhRG9Ay9X5OrvN-USdtmF4t2HSzF6V3aCA1NWNfIHCYVNrg</recordid><startdate>20100101</startdate><enddate>20100101</enddate><creator>Fitch, F. Moran</creator><creator>Martinez-Davalos, A.</creator><creator>Garcia-Garduno, O. A.</creator><scope>OTOTI</scope></search><sort><creationdate>20100101</creationdate><title>Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator</title><author>Fitch, F. Moran ; Martinez-Davalos, A. ; Garcia-Garduno, O. A.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c223t-6c3578cfbaa47a590a7a5ab1484889f39a5903a656bb7d3a5fea526595daec5e3</frbrgroupid><rsrctype>conference_proceedings</rsrctype><prefilter>conference_proceedings</prefilter><language>eng</language><creationdate>2010</creationdate><topic>ACCELERATORS</topic><topic>ANIMALS</topic><topic>BODY</topic><topic>COLLIMATORS</topic><topic>COMPARATIVE EVALUATIONS</topic><topic>CONFIGURATION</topic><topic>CYLINDRICAL CONFIGURATION</topic><topic>DIFFERENTIAL THERMAL ANALYSIS</topic><topic>DOSIMETRY</topic><topic>EVALUATION</topic><topic>IRRADIATION</topic><topic>ISODOSE CURVES</topic><topic>LINEAR ACCELERATORS</topic><topic>LUNGS</topic><topic>MAMMALS</topic><topic>MEDICINE</topic><topic>MOCKUP</topic><topic>NUCLEAR MEDICINE</topic><topic>ORGANS</topic><topic>PHANTOMS</topic><topic>RADIATION DOSE DISTRIBUTIONS</topic><topic>RADIOLOGY</topic><topic>RADIOLOGY AND NUCLEAR MEDICINE</topic><topic>RADIOTHERAPY</topic><topic>RATS</topic><topic>RESPIRATORY SYSTEM</topic><topic>RODENTS</topic><topic>SIMULATION</topic><topic>STRUCTURAL MODELS</topic><topic>THERAPY</topic><topic>THERMAL ANALYSIS</topic><topic>TRANSMISSION</topic><topic>VERTEBRATES</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fitch, F. Moran</creatorcontrib><creatorcontrib>Martinez-Davalos, A.</creatorcontrib><creatorcontrib>Garcia-Garduno, O. A.</creatorcontrib><collection>OSTI.GOV</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fitch, F. Moran</au><au>Martinez-Davalos, A.</au><au>Garcia-Garduno, O. A.</au><format>book</format><genre>proceeding</genre><ristype>CONF</ristype><atitle>Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator</atitle><btitle>AIP conference proceedings</btitle><date>2010-01-01</date><risdate>2010</risdate><volume>1310</volume><issue>1</issue><issn>0094-243X</issn><eissn>1551-7616</eissn><abstract>A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributions using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.</abstract><cop>United States</cop><doi>10.1063/1.3531582</doi></addata></record>
fulltext fulltext
identifier ISSN: 0094-243X
ispartof AIP conference proceedings, 2010, Vol.1310 (1)
issn 0094-243X
1551-7616
language eng
recordid cdi_osti_scitechconnect_21506964
source American Institute of Physics (AIP) Journals
subjects ACCELERATORS
ANIMALS
BODY
COLLIMATORS
COMPARATIVE EVALUATIONS
CONFIGURATION
CYLINDRICAL CONFIGURATION
DIFFERENTIAL THERMAL ANALYSIS
DOSIMETRY
EVALUATION
IRRADIATION
ISODOSE CURVES
LINEAR ACCELERATORS
LUNGS
MAMMALS
MEDICINE
MOCKUP
NUCLEAR MEDICINE
ORGANS
PHANTOMS
RADIATION DOSE DISTRIBUTIONS
RADIOLOGY
RADIOLOGY AND NUCLEAR MEDICINE
RADIOTHERAPY
RATS
RESPIRATORY SYSTEM
RODENTS
SIMULATION
STRUCTURAL MODELS
THERAPY
THERMAL ANALYSIS
TRANSMISSION
VERTEBRATES
title Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T09%3A54%3A39IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-osti&rft_val_fmt=info:ofi/fmt:kev:mtx:book&rft.genre=proceeding&rft.atitle=Dosimetry%20of%20a%20Small-Animal%20Irradiation%20Model%20using%20a%206%20MV%20Linear%20Accelerator&rft.btitle=AIP%20conference%20proceedings&rft.au=Fitch,%20F.%20Moran&rft.date=2010-01-01&rft.volume=1310&rft.issue=1&rft.issn=0094-243X&rft.eissn=1551-7616&rft_id=info:doi/10.1063/1.3531582&rft_dat=%3Costi%3E21506964%3C/osti%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true