Small prototype of Anger camera with submillimeter position resolution

In this work we present a small prototype of Anger camera based on a monolithic array of silicon drift detectors (SDDs) used as photodetectors for a single CsI(Tl) scintillator crystal. This prototype, having a total sensitive area of about 1 cm 2 , has been realized in order to evaluate the perform...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2005-04, Vol.76 (4), p.044303-044303-8
Hauptverfasser: Fiorini, Carlo, Perotti, Francesco
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 044303-8
container_issue 4
container_start_page 044303
container_title Review of scientific instruments
container_volume 76
creator Fiorini, Carlo
Perotti, Francesco
description In this work we present a small prototype of Anger camera based on a monolithic array of silicon drift detectors (SDDs) used as photodetectors for a single CsI(Tl) scintillator crystal. This prototype, having a total sensitive area of about 1 cm 2 , has been realized in order to evaluate the performances attainable in γ -ray imaging by using a SDD array instead of the more conventionally employed photomultiplier tubes. An intrinsic resolution better than 200 μ m has been measured with this camera. This result was achieved thanks to the extremely low electronics noise presented by the SDD units, fully exploited by the integration of the front-end junction field effect transistor directly on the detector chip. The main features of the detector as well as the results of its extended experimental characterization with 60 and 122 keV gamma rays are presented and discussed in the article. The present device has many potential applications in the field of medical imaging, in which outstanding position resolutions are essential.
doi_str_mv 10.1063/1.1891485
format Article
fullrecord <record><control><sourceid>scitation_osti_</sourceid><recordid>TN_cdi_osti_scitechconnect_20646480</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>rsi</sourcerecordid><originalsourceid>FETCH-LOGICAL-c312t-d05020737c7d76219c6b66e4a54f0ee5f8d7d1f9f163ed0d79e078d02d2c75e23</originalsourceid><addsrcrecordid>eNp1kMFKxDAQhoMouK4efIOCJw9dJ02bpBdhWVwVFjyo59BNJm6kbUqSRXx7W3fx5lxmYD5-Zj5CriksKHB2RxdU1rSU1QmZUZB1LnjBTskMgJU5F6U8JxcxfsJYFaUzsn7tmrbNhuCTT98DZt5my_4DQ6abDkOTfbm0y-J-27m2dR2mcTP46JLzfRYw-nY_jZfkzDZtxKtjn5P39cPb6infvDw-r5abXDNapNxABQUIJrQw42G01nzLOZZNVVpArKw0wlBbW8oZGjCiRhDSQGEKLSos2JzcHHJ9TE5F7RLqnfZ9jzqpAnjJSwkjdXugdPAxBrRqCK5rwreioCZNiqqjppG9P7BTWDP98j_860r9uVLeqiX7AaG0b3U</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Small prototype of Anger camera with submillimeter position resolution</title><source>AIP Journals Complete</source><source>AIP Digital Archive</source><creator>Fiorini, Carlo ; Perotti, Francesco</creator><creatorcontrib>Fiorini, Carlo ; Perotti, Francesco</creatorcontrib><description>In this work we present a small prototype of Anger camera based on a monolithic array of silicon drift detectors (SDDs) used as photodetectors for a single CsI(Tl) scintillator crystal. This prototype, having a total sensitive area of about 1 cm 2 , has been realized in order to evaluate the performances attainable in γ -ray imaging by using a SDD array instead of the more conventionally employed photomultiplier tubes. An intrinsic resolution better than 200 μ m has been measured with this camera. This result was achieved thanks to the extremely low electronics noise presented by the SDD units, fully exploited by the integration of the front-end junction field effect transistor directly on the detector chip. The main features of the detector as well as the results of its extended experimental characterization with 60 and 122 keV gamma rays are presented and discussed in the article. The present device has many potential applications in the field of medical imaging, in which outstanding position resolutions are essential.</description><identifier>ISSN: 0034-6748</identifier><identifier>EISSN: 1089-7623</identifier><identifier>DOI: 10.1063/1.1891485</identifier><identifier>CODEN: RSINAK</identifier><language>eng</language><publisher>United States: American Institute of Physics</publisher><subject>CAMERAS ; CRYSTALS ; FIELD EFFECT TRANSISTORS ; GAMMA DETECTION ; INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY ; NOISE ; PHOTODETECTORS ; PHOTOMULTIPLIERS ; SCINTILLATION COUNTERS ; SI SEMICONDUCTOR DETECTORS ; SILICON ; SPATIAL RESOLUTION</subject><ispartof>Review of scientific instruments, 2005-04, Vol.76 (4), p.044303-044303-8</ispartof><rights>2005 American Institute of Physics</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c312t-d05020737c7d76219c6b66e4a54f0ee5f8d7d1f9f163ed0d79e078d02d2c75e23</citedby><cites>FETCH-LOGICAL-c312t-d05020737c7d76219c6b66e4a54f0ee5f8d7d1f9f163ed0d79e078d02d2c75e23</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://pubs.aip.org/rsi/article-lookup/doi/10.1063/1.1891485$$EHTML$$P50$$Gscitation$$H</linktohtml><link.rule.ids>230,314,776,780,790,881,1553,4498,27901,27902,76126,76132</link.rule.ids><backlink>$$Uhttps://www.osti.gov/biblio/20646480$$D View this record in Osti.gov$$Hfree_for_read</backlink></links><search><creatorcontrib>Fiorini, Carlo</creatorcontrib><creatorcontrib>Perotti, Francesco</creatorcontrib><title>Small prototype of Anger camera with submillimeter position resolution</title><title>Review of scientific instruments</title><description>In this work we present a small prototype of Anger camera based on a monolithic array of silicon drift detectors (SDDs) used as photodetectors for a single CsI(Tl) scintillator crystal. This prototype, having a total sensitive area of about 1 cm 2 , has been realized in order to evaluate the performances attainable in γ -ray imaging by using a SDD array instead of the more conventionally employed photomultiplier tubes. An intrinsic resolution better than 200 μ m has been measured with this camera. This result was achieved thanks to the extremely low electronics noise presented by the SDD units, fully exploited by the integration of the front-end junction field effect transistor directly on the detector chip. The main features of the detector as well as the results of its extended experimental characterization with 60 and 122 keV gamma rays are presented and discussed in the article. The present device has many potential applications in the field of medical imaging, in which outstanding position resolutions are essential.</description><subject>CAMERAS</subject><subject>CRYSTALS</subject><subject>FIELD EFFECT TRANSISTORS</subject><subject>GAMMA DETECTION</subject><subject>INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY</subject><subject>NOISE</subject><subject>PHOTODETECTORS</subject><subject>PHOTOMULTIPLIERS</subject><subject>SCINTILLATION COUNTERS</subject><subject>SI SEMICONDUCTOR DETECTORS</subject><subject>SILICON</subject><subject>SPATIAL RESOLUTION</subject><issn>0034-6748</issn><issn>1089-7623</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><recordid>eNp1kMFKxDAQhoMouK4efIOCJw9dJ02bpBdhWVwVFjyo59BNJm6kbUqSRXx7W3fx5lxmYD5-Zj5CriksKHB2RxdU1rSU1QmZUZB1LnjBTskMgJU5F6U8JxcxfsJYFaUzsn7tmrbNhuCTT98DZt5my_4DQ6abDkOTfbm0y-J-27m2dR2mcTP46JLzfRYw-nY_jZfkzDZtxKtjn5P39cPb6infvDw-r5abXDNapNxABQUIJrQw42G01nzLOZZNVVpArKw0wlBbW8oZGjCiRhDSQGEKLSos2JzcHHJ9TE5F7RLqnfZ9jzqpAnjJSwkjdXugdPAxBrRqCK5rwreioCZNiqqjppG9P7BTWDP98j_860r9uVLeqiX7AaG0b3U</recordid><startdate>20050401</startdate><enddate>20050401</enddate><creator>Fiorini, Carlo</creator><creator>Perotti, Francesco</creator><general>American Institute of Physics</general><scope>AAYXX</scope><scope>CITATION</scope><scope>OTOTI</scope></search><sort><creationdate>20050401</creationdate><title>Small prototype of Anger camera with submillimeter position resolution</title><author>Fiorini, Carlo ; Perotti, Francesco</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c312t-d05020737c7d76219c6b66e4a54f0ee5f8d7d1f9f163ed0d79e078d02d2c75e23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>CAMERAS</topic><topic>CRYSTALS</topic><topic>FIELD EFFECT TRANSISTORS</topic><topic>GAMMA DETECTION</topic><topic>INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY</topic><topic>NOISE</topic><topic>PHOTODETECTORS</topic><topic>PHOTOMULTIPLIERS</topic><topic>SCINTILLATION COUNTERS</topic><topic>SI SEMICONDUCTOR DETECTORS</topic><topic>SILICON</topic><topic>SPATIAL RESOLUTION</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fiorini, Carlo</creatorcontrib><creatorcontrib>Perotti, Francesco</creatorcontrib><collection>CrossRef</collection><collection>OSTI.GOV</collection><jtitle>Review of scientific instruments</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fiorini, Carlo</au><au>Perotti, Francesco</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Small prototype of Anger camera with submillimeter position resolution</atitle><jtitle>Review of scientific instruments</jtitle><date>2005-04-01</date><risdate>2005</risdate><volume>76</volume><issue>4</issue><spage>044303</spage><epage>044303-8</epage><pages>044303-044303-8</pages><issn>0034-6748</issn><eissn>1089-7623</eissn><coden>RSINAK</coden><abstract>In this work we present a small prototype of Anger camera based on a monolithic array of silicon drift detectors (SDDs) used as photodetectors for a single CsI(Tl) scintillator crystal. This prototype, having a total sensitive area of about 1 cm 2 , has been realized in order to evaluate the performances attainable in γ -ray imaging by using a SDD array instead of the more conventionally employed photomultiplier tubes. An intrinsic resolution better than 200 μ m has been measured with this camera. This result was achieved thanks to the extremely low electronics noise presented by the SDD units, fully exploited by the integration of the front-end junction field effect transistor directly on the detector chip. The main features of the detector as well as the results of its extended experimental characterization with 60 and 122 keV gamma rays are presented and discussed in the article. The present device has many potential applications in the field of medical imaging, in which outstanding position resolutions are essential.</abstract><cop>United States</cop><pub>American Institute of Physics</pub><doi>10.1063/1.1891485</doi></addata></record>
fulltext fulltext
identifier ISSN: 0034-6748
ispartof Review of scientific instruments, 2005-04, Vol.76 (4), p.044303-044303-8
issn 0034-6748
1089-7623
language eng
recordid cdi_osti_scitechconnect_20646480
source AIP Journals Complete; AIP Digital Archive
subjects CAMERAS
CRYSTALS
FIELD EFFECT TRANSISTORS
GAMMA DETECTION
INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
NOISE
PHOTODETECTORS
PHOTOMULTIPLIERS
SCINTILLATION COUNTERS
SI SEMICONDUCTOR DETECTORS
SILICON
SPATIAL RESOLUTION
title Small prototype of Anger camera with submillimeter position resolution
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-12T09%3A51%3A05IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-scitation_osti_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Small%20prototype%20of%20Anger%20camera%20with%20submillimeter%20position%20resolution&rft.jtitle=Review%20of%20scientific%20instruments&rft.au=Fiorini,%20Carlo&rft.date=2005-04-01&rft.volume=76&rft.issue=4&rft.spage=044303&rft.epage=044303-8&rft.pages=044303-044303-8&rft.issn=0034-6748&rft.eissn=1089-7623&rft.coden=RSINAK&rft_id=info:doi/10.1063/1.1891485&rft_dat=%3Cscitation_osti_%3Ersi%3C/scitation_osti_%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true