Near-field dosimetry of {sup 125}I sources for interstitial brachytherapy implants measured using thermoluminescent sheets

The dosimetric characteristics were measured for two types of {sup 125}I low-energy photon-emitting sources by using a wide and highly sensitive thermoluminescent (TL) sheet film, which was developed for two-dimensional dose distribution measurements. The TL film is made of Teflon homogeneously mixe...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2004-12, Vol.31 (12)
Hauptverfasser: Iwata, Kazuro, Yue, Ning J., Nath, Ravinder, Department of Therapeutic Radiology, Yale University School of Medicine, P.O. Box 208040, New Haven, Connecticut 06520-8040
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The dosimetric characteristics were measured for two types of {sup 125}I low-energy photon-emitting sources by using a wide and highly sensitive thermoluminescent (TL) sheet film, which was developed for two-dimensional dose distribution measurements. The TL film is made of Teflon homogeneously mixed with small powders of thermoluminescence (BaSO{sub 4}:Eu doped). Various dosimetric parameters (i.e., radial dose function, 2D and 1D anisotropy functions) of model 6711 and 6702 {sup 125}I sources were obtained at various distances from the source surfaces to 15 mm. These parameters obtained with TL sheet were compared with the data recommended in the updated AAPM TG-43 report. The radial dose functions measured with TL sheet are in agreement with those established data of model 6711 {sup 125}I seed and model 6702 {sup 125}I seed at most of the distances within 5% and 7%, respectively. All the measured anisotropy functions showed symmetry about the longitudinal source axis. The anisotropy of dose distributions was clearly present in the immediate vicinity of the source edges. The measured 2D anisotropy function values at 1 cm are in reasonably good agreement with the recommended values. The differences at two points in the 1D anisotropy functions measured with TL sheet and the established data at 1 cm from source center were 0.7% and 1.9% for model 6711 and 6702 {sup 125}I sources, respectively; the differences at 0.5 cm were 1.5% and 1.7% for model 6711 and 6702 {sup 125}I sources, respectively. The relative dosimetric characteristics in the vicinity of actual interstitial brachytherapy sources containing {sup 125}I have been experimentally determined by using the TL sheet as a 2D dosimeter.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.1820012