Hip Arthroplasty Using the Bencox® Hip System: An Evaluation of a Consecutive Series of One Thousand Cases
This is a report on the outcomes associated with a consecutive series of 1,000 cementless hip arthroplasties utilizing the Bencox® hip stem-the first Korean-developed hip prosthesis. A consecutive series of 1,000 hip arthroplasties using the Bencox® hip stem were analyzed, starting from its initial...
Gespeichert in:
Veröffentlicht in: | Hip & pelvis 2018, 30(4), , pp.210-218 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This is a report on the outcomes associated with a consecutive series of 1,000 cementless hip arthroplasties utilizing the Bencox® hip stem-the first Korean-developed hip prosthesis.
A consecutive series of 1,000 hip arthroplasties using the Bencox® hip stem were analyzed, starting from its initial release (September 2006) until June 2014. Patients in this consecutive series underwent surgery for fractures (n=552), arthritis (n=155), avascular necrosis (n=209), and revisions (n=84). Of these 1,000 cases, patients with a minimum follow-up of at least 1 year (n=616) were retrospectively analyzed for radiographic and clinical outcomes (i.e., Harris hip score). The stability of the prosthesis was evaluated by examining subsidence.
During the follow-up period (mean follow-up period of 54.8 months), there were 2 cases requiring revision of the femoral stem-both were caused by periprosthetic fractures and neither involved stem loosening. The mean Harris hip score during follow-up was 95.5. Bone ongrowth occurred in 95% of patients; no cases of subsidence or aseptic loosening of the stem were detected, and no cases of postoperative complications such as ceramic breakage were observed.
Clinical and radiographic evaluations of hip arthroplasty using the Bencox® hip stem revealed excellent outcomes with an average of 54.8 month follow-up in a consecutive series of 1,000 cases. |
---|---|
ISSN: | 2287-3260 2287-3279 |
DOI: | 10.5371/hp.2018.30.4.210 |