Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure
Speckle Plethysmography (SPG) and Photoplethysmography (PPG) are different biophotonics technologies that allow for measurement of haemodynamics. As the difference between SPG and PPG under low perfusion conditions is not fully understood, a Cold Pressor Test (CPT-60 s full hand immersion in ice wat...
Gespeichert in:
Veröffentlicht in: | SENSORS 2023-05, Vol.23 (11) |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | |
---|---|
container_issue | 11 |
container_start_page | |
container_title | SENSORS |
container_volume | 23 |
creator | Olazabal, Jorge Herranz Lorato, Ilde Kling, Jesse Verhoeven, Marc Wieringa, Fokko Van Hoof, Chris Verkruijsse, Willem Hermeling, Evelien |
description | Speckle Plethysmography (SPG) and Photoplethysmography (PPG) are different biophotonics technologies that allow for measurement of haemodynamics. As the difference between SPG and PPG under low perfusion conditions is not fully understood, a Cold Pressor Test (CPT-60 s full hand immersion in ice water), was used to modulate blood pressure and peripheral circulation. A custom-built setup simultaneously derived SPG and PPG from the same video streams at two wavelengths (639 nm and 850 nm). SPG and PPG were measured at the right index finger location before and during the CPT using finger Arterial Pressure (fiAP) as a reference. The effect of the CPT on the Alternating Component amplitude (AC) and Signal-to-Noise Ratio (SNR) of dual-wavelength SPG and PPG signals was analysed across participants. Furthermore, waveform differences between SPG, PPG, and fiAP based on frequency harmonic ratios were analysed for each subject (n = 10). Both PPG and SPG at 850 nm show a significant reduction during the CPT in both AC and SNR. However, SPG showed significantly higher and more stable SNR than PPG in both study phases. Harmonic ratios were found substantially higher in SPG than PPG. Therefore, in low perfusion conditions, SPG seems to offer a more robust pulse wave monitoring with higher harmonic ratios than PPG. |
format | Article |
fullrecord | <record><control><sourceid>kuleuven</sourceid><recordid>TN_cdi_kuleuven_dspace_20_500_12942_726015</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>20_500_12942_726015</sourcerecordid><originalsourceid>FETCH-kuleuven_dspace_20_500_12942_7260153</originalsourceid><addsrcrecordid>eNqVysFOAjEUheGGSAKK73DXJphOOygsyUTikgj7ps4cmJHSNrcdlbUvLokujCtdnT8530CMi1KV07lS8uJHj8RlSi9SKq31fCw-qnCMlrsUPD0jvwGeNhH1wYHWDrk9pWPYs43tiaxvaN2GHOKvo-m583uqgjsDRkqBaYuU6Qk7MHyNhnKg1RmBackZ3Fn3RXvGRAx31iVcf--VuFk9bKvH6aF36F_hTZOirWGUNDMpTaEWpTL36k4WM_1PfPtnbPJ71p_oTGWD</addsrcrecordid><sourcetype>Institutional Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure</title><source>Lirias (KU Leuven Association)</source><source>Directory of Open Access Journals</source><source>MDPI - Multidisciplinary Digital Publishing Institute</source><source>PubMed Central</source><source>Free Full-Text Journals in Chemistry</source><source>EZB Electronic Journals Library</source><creator>Olazabal, Jorge Herranz ; Lorato, Ilde ; Kling, Jesse ; Verhoeven, Marc ; Wieringa, Fokko ; Van Hoof, Chris ; Verkruijsse, Willem ; Hermeling, Evelien</creator><creatorcontrib>Olazabal, Jorge Herranz ; Lorato, Ilde ; Kling, Jesse ; Verhoeven, Marc ; Wieringa, Fokko ; Van Hoof, Chris ; Verkruijsse, Willem ; Hermeling, Evelien</creatorcontrib><description>Speckle Plethysmography (SPG) and Photoplethysmography (PPG) are different biophotonics technologies that allow for measurement of haemodynamics. As the difference between SPG and PPG under low perfusion conditions is not fully understood, a Cold Pressor Test (CPT-60 s full hand immersion in ice water), was used to modulate blood pressure and peripheral circulation. A custom-built setup simultaneously derived SPG and PPG from the same video streams at two wavelengths (639 nm and 850 nm). SPG and PPG were measured at the right index finger location before and during the CPT using finger Arterial Pressure (fiAP) as a reference. The effect of the CPT on the Alternating Component amplitude (AC) and Signal-to-Noise Ratio (SNR) of dual-wavelength SPG and PPG signals was analysed across participants. Furthermore, waveform differences between SPG, PPG, and fiAP based on frequency harmonic ratios were analysed for each subject (n = 10). Both PPG and SPG at 850 nm show a significant reduction during the CPT in both AC and SNR. However, SPG showed significantly higher and more stable SNR than PPG in both study phases. Harmonic ratios were found substantially higher in SPG than PPG. Therefore, in low perfusion conditions, SPG seems to offer a more robust pulse wave monitoring with higher harmonic ratios than PPG.</description><identifier>ISSN: 1424-8220</identifier><identifier>EISSN: 1424-8220</identifier><language>eng</language><publisher>MDPI</publisher><ispartof>SENSORS, 2023-05, Vol.23 (11)</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>315,316,781,785,27862</link.rule.ids></links><search><creatorcontrib>Olazabal, Jorge Herranz</creatorcontrib><creatorcontrib>Lorato, Ilde</creatorcontrib><creatorcontrib>Kling, Jesse</creatorcontrib><creatorcontrib>Verhoeven, Marc</creatorcontrib><creatorcontrib>Wieringa, Fokko</creatorcontrib><creatorcontrib>Van Hoof, Chris</creatorcontrib><creatorcontrib>Verkruijsse, Willem</creatorcontrib><creatorcontrib>Hermeling, Evelien</creatorcontrib><title>Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure</title><title>SENSORS</title><description>Speckle Plethysmography (SPG) and Photoplethysmography (PPG) are different biophotonics technologies that allow for measurement of haemodynamics. As the difference between SPG and PPG under low perfusion conditions is not fully understood, a Cold Pressor Test (CPT-60 s full hand immersion in ice water), was used to modulate blood pressure and peripheral circulation. A custom-built setup simultaneously derived SPG and PPG from the same video streams at two wavelengths (639 nm and 850 nm). SPG and PPG were measured at the right index finger location before and during the CPT using finger Arterial Pressure (fiAP) as a reference. The effect of the CPT on the Alternating Component amplitude (AC) and Signal-to-Noise Ratio (SNR) of dual-wavelength SPG and PPG signals was analysed across participants. Furthermore, waveform differences between SPG, PPG, and fiAP based on frequency harmonic ratios were analysed for each subject (n = 10). Both PPG and SPG at 850 nm show a significant reduction during the CPT in both AC and SNR. However, SPG showed significantly higher and more stable SNR than PPG in both study phases. Harmonic ratios were found substantially higher in SPG than PPG. Therefore, in low perfusion conditions, SPG seems to offer a more robust pulse wave monitoring with higher harmonic ratios than PPG.</description><issn>1424-8220</issn><issn>1424-8220</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><sourceid>FZOIL</sourceid><recordid>eNqVysFOAjEUheGGSAKK73DXJphOOygsyUTikgj7ps4cmJHSNrcdlbUvLokujCtdnT8530CMi1KV07lS8uJHj8RlSi9SKq31fCw-qnCMlrsUPD0jvwGeNhH1wYHWDrk9pWPYs43tiaxvaN2GHOKvo-m583uqgjsDRkqBaYuU6Qk7MHyNhnKg1RmBackZ3Fn3RXvGRAx31iVcf--VuFk9bKvH6aF36F_hTZOirWGUNDMpTaEWpTL36k4WM_1PfPtnbPJ71p_oTGWD</recordid><startdate>20230524</startdate><enddate>20230524</enddate><creator>Olazabal, Jorge Herranz</creator><creator>Lorato, Ilde</creator><creator>Kling, Jesse</creator><creator>Verhoeven, Marc</creator><creator>Wieringa, Fokko</creator><creator>Van Hoof, Chris</creator><creator>Verkruijsse, Willem</creator><creator>Hermeling, Evelien</creator><general>MDPI</general><scope>FZOIL</scope></search><sort><creationdate>20230524</creationdate><title>Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure</title><author>Olazabal, Jorge Herranz ; Lorato, Ilde ; Kling, Jesse ; Verhoeven, Marc ; Wieringa, Fokko ; Van Hoof, Chris ; Verkruijsse, Willem ; Hermeling, Evelien</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-kuleuven_dspace_20_500_12942_7260153</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Olazabal, Jorge Herranz</creatorcontrib><creatorcontrib>Lorato, Ilde</creatorcontrib><creatorcontrib>Kling, Jesse</creatorcontrib><creatorcontrib>Verhoeven, Marc</creatorcontrib><creatorcontrib>Wieringa, Fokko</creatorcontrib><creatorcontrib>Van Hoof, Chris</creatorcontrib><creatorcontrib>Verkruijsse, Willem</creatorcontrib><creatorcontrib>Hermeling, Evelien</creatorcontrib><collection>Lirias (KU Leuven Association)</collection><jtitle>SENSORS</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Olazabal, Jorge Herranz</au><au>Lorato, Ilde</au><au>Kling, Jesse</au><au>Verhoeven, Marc</au><au>Wieringa, Fokko</au><au>Van Hoof, Chris</au><au>Verkruijsse, Willem</au><au>Hermeling, Evelien</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure</atitle><jtitle>SENSORS</jtitle><date>2023-05-24</date><risdate>2023</risdate><volume>23</volume><issue>11</issue><issn>1424-8220</issn><eissn>1424-8220</eissn><abstract>Speckle Plethysmography (SPG) and Photoplethysmography (PPG) are different biophotonics technologies that allow for measurement of haemodynamics. As the difference between SPG and PPG under low perfusion conditions is not fully understood, a Cold Pressor Test (CPT-60 s full hand immersion in ice water), was used to modulate blood pressure and peripheral circulation. A custom-built setup simultaneously derived SPG and PPG from the same video streams at two wavelengths (639 nm and 850 nm). SPG and PPG were measured at the right index finger location before and during the CPT using finger Arterial Pressure (fiAP) as a reference. The effect of the CPT on the Alternating Component amplitude (AC) and Signal-to-Noise Ratio (SNR) of dual-wavelength SPG and PPG signals was analysed across participants. Furthermore, waveform differences between SPG, PPG, and fiAP based on frequency harmonic ratios were analysed for each subject (n = 10). Both PPG and SPG at 850 nm show a significant reduction during the CPT in both AC and SNR. However, SPG showed significantly higher and more stable SNR than PPG in both study phases. Harmonic ratios were found substantially higher in SPG than PPG. Therefore, in low perfusion conditions, SPG seems to offer a more robust pulse wave monitoring with higher harmonic ratios than PPG.</abstract><pub>MDPI</pub><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1424-8220 |
ispartof | SENSORS, 2023-05, Vol.23 (11) |
issn | 1424-8220 1424-8220 |
language | eng |
recordid | cdi_kuleuven_dspace_20_500_12942_726015 |
source | Lirias (KU Leuven Association); Directory of Open Access Journals; MDPI - Multidisciplinary Digital Publishing Institute; PubMed Central; Free Full-Text Journals in Chemistry; EZB Electronic Journals Library |
title | Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-17T16%3A48%3A44IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-kuleuven&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Comparison%20between%20Speckle%20Plethysmography%20and%20Photoplethysmography%20during%20Cold%20Pressor%20Test%20Referenced%20to%20Finger%20Arterial%20Pressure&rft.jtitle=SENSORS&rft.au=Olazabal,%20Jorge%20Herranz&rft.date=2023-05-24&rft.volume=23&rft.issue=11&rft.issn=1424-8220&rft.eissn=1424-8220&rft_id=info:doi/&rft_dat=%3Ckuleuven%3E20_500_12942_726015%3C/kuleuven%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |