Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study

In the nuclear medicine imaging, quality control (QC) process using quadrant bar phantom is fundamental aspect of evaluating the spatial resolution. In addition, QC process of gamma camera is performed by daily or weekly. Recently, Monte Carlo simulation using the Geant4 application for tomographic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear engineering and technology 2021, Vol.53 (6), p.1947-1954
Hauptverfasser: Park, Chan Rok, Kang, Seong-Hyeon, Lee, Youngjin
Format: Artikel
Sprache:kor
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1954
container_issue 6
container_start_page 1947
container_title Nuclear engineering and technology
container_volume 53
creator Park, Chan Rok
Kang, Seong-Hyeon
Lee, Youngjin
description In the nuclear medicine imaging, quality control (QC) process using quadrant bar phantom is fundamental aspect of evaluating the spatial resolution. In addition, QC process of gamma camera is performed by daily or weekly. Recently, Monte Carlo simulation using the Geant4 application for tomographic emission (GATE) is widely applied in the pre-clinical nuclear medicine field for modeling gamma cameras with pixelated cadmium telluride (CdTe) semiconductor detector. In this study, we modeled a pixelated CdTe semiconductor detector and quadrant bar phantom (0.5, 1.0, 1.5, and 2.0 mm bar thicknesses) using the GATE tool. Similarity analysis based on correlation coefficients and peak signal-to-noise ratios was performed to compare image qualities for various source to collimator distances (0, 2, 4, 6, and 8 cm) and collimator lengths (0.2, 0.4, 0.6, 0.8, and 1.0 cm). To this end, we selected reference images based on collimator length and source to collimator distance settings. The results demonstrate that as the collimator length increases and the source to collimator distance decreases, the similarity to reference images improves. Therefore, our simulation results represent valuable information for the modeling of CdTe-based semiconductor gamma imaging systems and QC phantoms in the field of nuclear medicine.
format Article
fullrecord <record><control><sourceid>kisti</sourceid><recordid>TN_cdi_kisti_ndsl_JAKO202124452738530</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>JAKO202124452738530</sourcerecordid><originalsourceid>FETCH-kisti_ndsl_JAKO2021244527385303</originalsourceid><addsrcrecordid>eNqNj81KxEAQhAdRcNF9h754DCQzCRu8SVBEEQ_uwdvSm-nExvlZp2fAvIjPaw4-gKcqiq8o6kxttDZtZbr-_Vxtmp3pq25nzKXaivCxbnXT1F3fbNTPG3t2mDgvgAHdIiwQJzjxNznMZGGwewIhz2MMtow5JpjRe4QRPSUE9jgTFOEwA8JXQZswZDhigtPH6qKHae2EMjpaM0-WRw50Cy8xZIIBk4sg7Ms6xzGA5GKXa3UxoRPa_umVunm43w-P1SdL5kOw4g5Pd8-vutaNbttOrw87U5v_cr_Zzlm_</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study</title><source>DOAJ Directory of Open Access Journals</source><source>EZB-FREE-00999 freely available EZB journals</source><source>Alma/SFX Local Collection</source><creator>Park, Chan Rok ; Kang, Seong-Hyeon ; Lee, Youngjin</creator><creatorcontrib>Park, Chan Rok ; Kang, Seong-Hyeon ; Lee, Youngjin</creatorcontrib><description>In the nuclear medicine imaging, quality control (QC) process using quadrant bar phantom is fundamental aspect of evaluating the spatial resolution. In addition, QC process of gamma camera is performed by daily or weekly. Recently, Monte Carlo simulation using the Geant4 application for tomographic emission (GATE) is widely applied in the pre-clinical nuclear medicine field for modeling gamma cameras with pixelated cadmium telluride (CdTe) semiconductor detector. In this study, we modeled a pixelated CdTe semiconductor detector and quadrant bar phantom (0.5, 1.0, 1.5, and 2.0 mm bar thicknesses) using the GATE tool. Similarity analysis based on correlation coefficients and peak signal-to-noise ratios was performed to compare image qualities for various source to collimator distances (0, 2, 4, 6, and 8 cm) and collimator lengths (0.2, 0.4, 0.6, 0.8, and 1.0 cm). To this end, we selected reference images based on collimator length and source to collimator distance settings. The results demonstrate that as the collimator length increases and the source to collimator distance decreases, the similarity to reference images improves. Therefore, our simulation results represent valuable information for the modeling of CdTe-based semiconductor gamma imaging systems and QC phantoms in the field of nuclear medicine.</description><identifier>ISSN: 1738-5733</identifier><identifier>EISSN: 2234-358X</identifier><language>kor</language><ispartof>Nuclear engineering and technology, 2021, Vol.53 (6), p.1947-1954</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,4024</link.rule.ids></links><search><creatorcontrib>Park, Chan Rok</creatorcontrib><creatorcontrib>Kang, Seong-Hyeon</creatorcontrib><creatorcontrib>Lee, Youngjin</creatorcontrib><title>Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study</title><title>Nuclear engineering and technology</title><addtitle>Nuclear engineering and technology : an international journal of the Korean Nuclear Society</addtitle><description>In the nuclear medicine imaging, quality control (QC) process using quadrant bar phantom is fundamental aspect of evaluating the spatial resolution. In addition, QC process of gamma camera is performed by daily or weekly. Recently, Monte Carlo simulation using the Geant4 application for tomographic emission (GATE) is widely applied in the pre-clinical nuclear medicine field for modeling gamma cameras with pixelated cadmium telluride (CdTe) semiconductor detector. In this study, we modeled a pixelated CdTe semiconductor detector and quadrant bar phantom (0.5, 1.0, 1.5, and 2.0 mm bar thicknesses) using the GATE tool. Similarity analysis based on correlation coefficients and peak signal-to-noise ratios was performed to compare image qualities for various source to collimator distances (0, 2, 4, 6, and 8 cm) and collimator lengths (0.2, 0.4, 0.6, 0.8, and 1.0 cm). To this end, we selected reference images based on collimator length and source to collimator distance settings. The results demonstrate that as the collimator length increases and the source to collimator distance decreases, the similarity to reference images improves. Therefore, our simulation results represent valuable information for the modeling of CdTe-based semiconductor gamma imaging systems and QC phantoms in the field of nuclear medicine.</description><issn>1738-5733</issn><issn>2234-358X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>JDI</sourceid><recordid>eNqNj81KxEAQhAdRcNF9h754DCQzCRu8SVBEEQ_uwdvSm-nExvlZp2fAvIjPaw4-gKcqiq8o6kxttDZtZbr-_Vxtmp3pq25nzKXaivCxbnXT1F3fbNTPG3t2mDgvgAHdIiwQJzjxNznMZGGwewIhz2MMtow5JpjRe4QRPSUE9jgTFOEwA8JXQZswZDhigtPH6qKHae2EMjpaM0-WRw50Cy8xZIIBk4sg7Ms6xzGA5GKXa3UxoRPa_umVunm43w-P1SdL5kOw4g5Pd8-vutaNbttOrw87U5v_cr_Zzlm_</recordid><startdate>2021</startdate><enddate>2021</enddate><creator>Park, Chan Rok</creator><creator>Kang, Seong-Hyeon</creator><creator>Lee, Youngjin</creator><scope>JDI</scope></search><sort><creationdate>2021</creationdate><title>Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study</title><author>Park, Chan Rok ; Kang, Seong-Hyeon ; Lee, Youngjin</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-kisti_ndsl_JAKO2021244527385303</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>kor</language><creationdate>2021</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Park, Chan Rok</creatorcontrib><creatorcontrib>Kang, Seong-Hyeon</creatorcontrib><creatorcontrib>Lee, Youngjin</creatorcontrib><collection>KoreaScience</collection><jtitle>Nuclear engineering and technology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Park, Chan Rok</au><au>Kang, Seong-Hyeon</au><au>Lee, Youngjin</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study</atitle><jtitle>Nuclear engineering and technology</jtitle><addtitle>Nuclear engineering and technology : an international journal of the Korean Nuclear Society</addtitle><date>2021</date><risdate>2021</risdate><volume>53</volume><issue>6</issue><spage>1947</spage><epage>1954</epage><pages>1947-1954</pages><issn>1738-5733</issn><eissn>2234-358X</eissn><abstract>In the nuclear medicine imaging, quality control (QC) process using quadrant bar phantom is fundamental aspect of evaluating the spatial resolution. In addition, QC process of gamma camera is performed by daily or weekly. Recently, Monte Carlo simulation using the Geant4 application for tomographic emission (GATE) is widely applied in the pre-clinical nuclear medicine field for modeling gamma cameras with pixelated cadmium telluride (CdTe) semiconductor detector. In this study, we modeled a pixelated CdTe semiconductor detector and quadrant bar phantom (0.5, 1.0, 1.5, and 2.0 mm bar thicknesses) using the GATE tool. Similarity analysis based on correlation coefficients and peak signal-to-noise ratios was performed to compare image qualities for various source to collimator distances (0, 2, 4, 6, and 8 cm) and collimator lengths (0.2, 0.4, 0.6, 0.8, and 1.0 cm). To this end, we selected reference images based on collimator length and source to collimator distance settings. The results demonstrate that as the collimator length increases and the source to collimator distance decreases, the similarity to reference images improves. Therefore, our simulation results represent valuable information for the modeling of CdTe-based semiconductor gamma imaging systems and QC phantoms in the field of nuclear medicine.</abstract><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1738-5733
ispartof Nuclear engineering and technology, 2021, Vol.53 (6), p.1947-1954
issn 1738-5733
2234-358X
language kor
recordid cdi_kisti_ndsl_JAKO202124452738530
source DOAJ Directory of Open Access Journals; EZB-FREE-00999 freely available EZB journals; Alma/SFX Local Collection
title Similarity analysis of pixelated CdTe semiconductor gamma camera image using a quadrant bar phantom for nuclear medicine: Monte Carlo simulation study
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-24T03%3A21%3A54IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-kisti&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Similarity%20analysis%20of%20pixelated%20CdTe%20semiconductor%20gamma%20camera%20image%20using%20a%20quadrant%20bar%20phantom%20for%20nuclear%20medicine:%20Monte%20Carlo%20simulation%20study&rft.jtitle=Nuclear%20engineering%20and%20technology&rft.au=Park,%20Chan%20Rok&rft.date=2021&rft.volume=53&rft.issue=6&rft.spage=1947&rft.epage=1954&rft.pages=1947-1954&rft.issn=1738-5733&rft.eissn=2234-358X&rft_id=info:doi/&rft_dat=%3Ckisti%3EJAKO202124452738530%3C/kisti%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true