Design and fabrication of an optimized Rogowski coil for plasma current sensing and the operation confidence of Alvand tokamak
To understand the fundamental parameters of Alvand tokamak, A Rogowski coil with an active integrator was designed and constructed. Considering the characteristics of the Alvand tokamak, the structural and electrical parameters affecting the sensor function, were designed. Calibration was performed...
Gespeichert in:
Veröffentlicht in: | Nuclear engineering and technology 2020, Vol.52 (11), p.2535-2542 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | kor |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 2542 |
---|---|
container_issue | 11 |
container_start_page | 2535 |
container_title | Nuclear engineering and technology |
container_volume | 52 |
creator | Eydan, Anna Shirani, Babak Sadeghi, Yahya Asgarian, Mohammad Ali Noori, Ehsanollah |
description | To understand the fundamental parameters of Alvand tokamak, A Rogowski coil with an active integrator was designed and constructed. Considering the characteristics of the Alvand tokamak, the structural and electrical parameters affecting the sensor function, were designed. Calibration was performed directly in the presence of plasma. The sensor has a high resistance against interference of external magnetic fields. Plasma current was measured in various experiments. Based on the plasma current profile and loop voltage signal, the time evolution of plasma discharge was investigated and plasma behavior was analyzed. Alvand tokamak discharge was divided into several regions that represents different physical phenomena in the plasma. During the plasma discharge time, plasma had significant changes and its characteristic was not uniform. To understand the plasma behavior in each of the phases, the Rogowski sensor should have sufficient time resolution. The Rogowski sensor with a frequency up to 15 kHz was appropriate for this purpose. |
format | Article |
fullrecord | <record><control><sourceid>kisti</sourceid><recordid>TN_cdi_kisti_ndsl_JAKO202031837625238</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>JAKO202031837625238</sourcerecordid><originalsourceid>FETCH-kisti_ndsl_JAKO2020318376252383</originalsourceid><addsrcrecordid>eNqNjMFKAzEURUOx0EH7D2_jcmCSdJxsi62ILgRx4a6kmWT6mExeyYsKLvx2a_UDXF24nHNmolJKr2rdmtcLUclOm7rttF6IJTPum5WSsmmNrMTXxjMOCWzqIdh9RmcLUgIKpwvoWHDCT9_DMw30wSOCI4wQKMMxWp4suLecfSrAPjGm4RwqB39Sff5NOUoBe5-c_6mu4_sZodFOdrwS82Aj--XfXorru-3L7X09IhfcpZ7j7mH9-KQa1WhpdHejWqWN_i_3DZeSUHw</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Design and fabrication of an optimized Rogowski coil for plasma current sensing and the operation confidence of Alvand tokamak</title><source>DOAJ Directory of Open Access Journals</source><source>Alma/SFX Local Collection</source><source>EZB Electronic Journals Library</source><creator>Eydan, Anna ; Shirani, Babak ; Sadeghi, Yahya ; Asgarian, Mohammad Ali ; Noori, Ehsanollah</creator><creatorcontrib>Eydan, Anna ; Shirani, Babak ; Sadeghi, Yahya ; Asgarian, Mohammad Ali ; Noori, Ehsanollah</creatorcontrib><description>To understand the fundamental parameters of Alvand tokamak, A Rogowski coil with an active integrator was designed and constructed. Considering the characteristics of the Alvand tokamak, the structural and electrical parameters affecting the sensor function, were designed. Calibration was performed directly in the presence of plasma. The sensor has a high resistance against interference of external magnetic fields. Plasma current was measured in various experiments. Based on the plasma current profile and loop voltage signal, the time evolution of plasma discharge was investigated and plasma behavior was analyzed. Alvand tokamak discharge was divided into several regions that represents different physical phenomena in the plasma. During the plasma discharge time, plasma had significant changes and its characteristic was not uniform. To understand the plasma behavior in each of the phases, the Rogowski sensor should have sufficient time resolution. The Rogowski sensor with a frequency up to 15 kHz was appropriate for this purpose.</description><identifier>ISSN: 1738-5733</identifier><identifier>EISSN: 2234-358X</identifier><language>kor</language><ispartof>Nuclear engineering and technology, 2020, Vol.52 (11), p.2535-2542</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,4024</link.rule.ids></links><search><creatorcontrib>Eydan, Anna</creatorcontrib><creatorcontrib>Shirani, Babak</creatorcontrib><creatorcontrib>Sadeghi, Yahya</creatorcontrib><creatorcontrib>Asgarian, Mohammad Ali</creatorcontrib><creatorcontrib>Noori, Ehsanollah</creatorcontrib><title>Design and fabrication of an optimized Rogowski coil for plasma current sensing and the operation confidence of Alvand tokamak</title><title>Nuclear engineering and technology</title><addtitle>Nuclear engineering and technology : an international journal of the Korean Nuclear Society</addtitle><description>To understand the fundamental parameters of Alvand tokamak, A Rogowski coil with an active integrator was designed and constructed. Considering the characteristics of the Alvand tokamak, the structural and electrical parameters affecting the sensor function, were designed. Calibration was performed directly in the presence of plasma. The sensor has a high resistance against interference of external magnetic fields. Plasma current was measured in various experiments. Based on the plasma current profile and loop voltage signal, the time evolution of plasma discharge was investigated and plasma behavior was analyzed. Alvand tokamak discharge was divided into several regions that represents different physical phenomena in the plasma. During the plasma discharge time, plasma had significant changes and its characteristic was not uniform. To understand the plasma behavior in each of the phases, the Rogowski sensor should have sufficient time resolution. The Rogowski sensor with a frequency up to 15 kHz was appropriate for this purpose.</description><issn>1738-5733</issn><issn>2234-358X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>JDI</sourceid><recordid>eNqNjMFKAzEURUOx0EH7D2_jcmCSdJxsi62ILgRx4a6kmWT6mExeyYsKLvx2a_UDXF24nHNmolJKr2rdmtcLUclOm7rttF6IJTPum5WSsmmNrMTXxjMOCWzqIdh9RmcLUgIKpwvoWHDCT9_DMw30wSOCI4wQKMMxWp4suLecfSrAPjGm4RwqB39Sff5NOUoBe5-c_6mu4_sZodFOdrwS82Aj--XfXorru-3L7X09IhfcpZ7j7mH9-KQa1WhpdHejWqWN_i_3DZeSUHw</recordid><startdate>2020</startdate><enddate>2020</enddate><creator>Eydan, Anna</creator><creator>Shirani, Babak</creator><creator>Sadeghi, Yahya</creator><creator>Asgarian, Mohammad Ali</creator><creator>Noori, Ehsanollah</creator><scope>JDI</scope></search><sort><creationdate>2020</creationdate><title>Design and fabrication of an optimized Rogowski coil for plasma current sensing and the operation confidence of Alvand tokamak</title><author>Eydan, Anna ; Shirani, Babak ; Sadeghi, Yahya ; Asgarian, Mohammad Ali ; Noori, Ehsanollah</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-kisti_ndsl_JAKO2020318376252383</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>kor</language><creationdate>2020</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Eydan, Anna</creatorcontrib><creatorcontrib>Shirani, Babak</creatorcontrib><creatorcontrib>Sadeghi, Yahya</creatorcontrib><creatorcontrib>Asgarian, Mohammad Ali</creatorcontrib><creatorcontrib>Noori, Ehsanollah</creatorcontrib><collection>KoreaScience</collection><jtitle>Nuclear engineering and technology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Eydan, Anna</au><au>Shirani, Babak</au><au>Sadeghi, Yahya</au><au>Asgarian, Mohammad Ali</au><au>Noori, Ehsanollah</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Design and fabrication of an optimized Rogowski coil for plasma current sensing and the operation confidence of Alvand tokamak</atitle><jtitle>Nuclear engineering and technology</jtitle><addtitle>Nuclear engineering and technology : an international journal of the Korean Nuclear Society</addtitle><date>2020</date><risdate>2020</risdate><volume>52</volume><issue>11</issue><spage>2535</spage><epage>2542</epage><pages>2535-2542</pages><issn>1738-5733</issn><eissn>2234-358X</eissn><abstract>To understand the fundamental parameters of Alvand tokamak, A Rogowski coil with an active integrator was designed and constructed. Considering the characteristics of the Alvand tokamak, the structural and electrical parameters affecting the sensor function, were designed. Calibration was performed directly in the presence of plasma. The sensor has a high resistance against interference of external magnetic fields. Plasma current was measured in various experiments. Based on the plasma current profile and loop voltage signal, the time evolution of plasma discharge was investigated and plasma behavior was analyzed. Alvand tokamak discharge was divided into several regions that represents different physical phenomena in the plasma. During the plasma discharge time, plasma had significant changes and its characteristic was not uniform. To understand the plasma behavior in each of the phases, the Rogowski sensor should have sufficient time resolution. The Rogowski sensor with a frequency up to 15 kHz was appropriate for this purpose.</abstract><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1738-5733 |
ispartof | Nuclear engineering and technology, 2020, Vol.52 (11), p.2535-2542 |
issn | 1738-5733 2234-358X |
language | kor |
recordid | cdi_kisti_ndsl_JAKO202031837625238 |
source | DOAJ Directory of Open Access Journals; Alma/SFX Local Collection; EZB Electronic Journals Library |
title | Design and fabrication of an optimized Rogowski coil for plasma current sensing and the operation confidence of Alvand tokamak |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-20T21%3A35%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-kisti&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Design%20and%20fabrication%20of%20an%20optimized%20Rogowski%20coil%20for%20plasma%20current%20sensing%20and%20the%20operation%20confidence%20of%20Alvand%20tokamak&rft.jtitle=Nuclear%20engineering%20and%20technology&rft.au=Eydan,%20Anna&rft.date=2020&rft.volume=52&rft.issue=11&rft.spage=2535&rft.epage=2542&rft.pages=2535-2542&rft.issn=1738-5733&rft.eissn=2234-358X&rft_id=info:doi/&rft_dat=%3Ckisti%3EJAKO202031837625238%3C/kisti%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |