Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography
Purpose: This study was performed to investigate the intra- and inter-observer variability in linear measurements with axial images obtained by PreXion (PreXion Inc., San Mateo, USA) and i-CAT(Imaging Sciences International, Xoran Technologies Inc., Hatfield, USA) CBCT scanners, with different voxel...
Gespeichert in:
Veröffentlicht in: | Imaging science in dentistry 2017, Vol.47 (1), p.11-15 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | kor |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 15 |
---|---|
container_issue | 1 |
container_start_page | 11 |
container_title | Imaging science in dentistry |
container_volume | 47 |
creator | Silva, Nathalia Cristine da Barriviera, Mauricio Junqueira, Jose Luiz Cintra Panzarella, Francine Kuhl Raitz, Ricardo |
description | Purpose: This study was performed to investigate the intra- and inter-observer variability in linear measurements with axial images obtained by PreXion (PreXion Inc., San Mateo, USA) and i-CAT(Imaging Sciences International, Xoran Technologies Inc., Hatfield, USA) CBCT scanners, with different voxel sizes. Materials and Methods: A cylindrical object made from nylon with radiopaque markers (phantom) was scanned by i-CAT and PreXion 3D devices. For each axial image, measurements were taken twice in the horizontal(distance A-B) and vertical (distance C-D) directions, randomly, with a one-week interval between measurements, by four oral radiologists with five years or more experience in the use of these measuring tools. Results: All of the obtained linear measurements had lower values than those of the phantom. The statistical analysis showed high intra- and inter-observer reliability (p=0.297). Compared to the real measurements, the measurements obtained using the i-CAT device and PreXion tomography, on average, revealed absolute errors ranging from 0.22 to 0.59 mm and from 0.23 to 0.63 mm, respectively. Conclusion: It can be concluded that both scanners are accurate, although the linear measurements are underestimations, with no significant differences between the evaluators. |
format | Article |
fullrecord | <record><control><sourceid>kisti</sourceid><recordid>TN_cdi_kisti_ndsl_JAKO201713647762564</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>JAKO201713647762564</sourcerecordid><originalsourceid>FETCH-kisti_ndsl_JAKO2017136477625643</originalsourceid><addsrcrecordid>eNqNirtOAzEQRS0EEhHJP0xDudKunazTIgTiUdDQR-PskEzwYzX2Irbj03GBUnObe-7jQi20NqaxW9Nenlnra7XK-dRWbfTW9t1C_TzHIphcJvkiAYwDcCwk50ZolDRMe3bsucx1Bc-RUCAQ5kkoUCwZUgT8ZvTAAQ9UsytYbwO4GfYpUuMIQ6UwTqW2JYV0EByP81JdfaDPtPrzG3X7-PB-_9R8ci68i0P2u5e71zfddrYz_draXm_6tfnv7xcbr1KL</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography</title><source>KoreaMed Synapse</source><source>KoreaMed Open Access</source><source>DOAJ Directory of Open Access Journals</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>PubMed Central</source><source>PubMed Central Open Access</source><creator>Silva, Nathalia Cristine da ; Barriviera, Mauricio ; Junqueira, Jose Luiz Cintra ; Panzarella, Francine Kuhl ; Raitz, Ricardo</creator><creatorcontrib>Silva, Nathalia Cristine da ; Barriviera, Mauricio ; Junqueira, Jose Luiz Cintra ; Panzarella, Francine Kuhl ; Raitz, Ricardo</creatorcontrib><description>Purpose: This study was performed to investigate the intra- and inter-observer variability in linear measurements with axial images obtained by PreXion (PreXion Inc., San Mateo, USA) and i-CAT(Imaging Sciences International, Xoran Technologies Inc., Hatfield, USA) CBCT scanners, with different voxel sizes. Materials and Methods: A cylindrical object made from nylon with radiopaque markers (phantom) was scanned by i-CAT and PreXion 3D devices. For each axial image, measurements were taken twice in the horizontal(distance A-B) and vertical (distance C-D) directions, randomly, with a one-week interval between measurements, by four oral radiologists with five years or more experience in the use of these measuring tools. Results: All of the obtained linear measurements had lower values than those of the phantom. The statistical analysis showed high intra- and inter-observer reliability (p=0.297). Compared to the real measurements, the measurements obtained using the i-CAT device and PreXion tomography, on average, revealed absolute errors ranging from 0.22 to 0.59 mm and from 0.23 to 0.63 mm, respectively. Conclusion: It can be concluded that both scanners are accurate, although the linear measurements are underestimations, with no significant differences between the evaluators.</description><identifier>ISSN: 2233-7822</identifier><identifier>EISSN: 2233-7830</identifier><language>kor</language><ispartof>Imaging science in dentistry, 2017, Vol.47 (1), p.11-15</ispartof><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,776,780,881,4010</link.rule.ids></links><search><creatorcontrib>Silva, Nathalia Cristine da</creatorcontrib><creatorcontrib>Barriviera, Mauricio</creatorcontrib><creatorcontrib>Junqueira, Jose Luiz Cintra</creatorcontrib><creatorcontrib>Panzarella, Francine Kuhl</creatorcontrib><creatorcontrib>Raitz, Ricardo</creatorcontrib><title>Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography</title><title>Imaging science in dentistry</title><addtitle>Imaging science in dentistry</addtitle><description>Purpose: This study was performed to investigate the intra- and inter-observer variability in linear measurements with axial images obtained by PreXion (PreXion Inc., San Mateo, USA) and i-CAT(Imaging Sciences International, Xoran Technologies Inc., Hatfield, USA) CBCT scanners, with different voxel sizes. Materials and Methods: A cylindrical object made from nylon with radiopaque markers (phantom) was scanned by i-CAT and PreXion 3D devices. For each axial image, measurements were taken twice in the horizontal(distance A-B) and vertical (distance C-D) directions, randomly, with a one-week interval between measurements, by four oral radiologists with five years or more experience in the use of these measuring tools. Results: All of the obtained linear measurements had lower values than those of the phantom. The statistical analysis showed high intra- and inter-observer reliability (p=0.297). Compared to the real measurements, the measurements obtained using the i-CAT device and PreXion tomography, on average, revealed absolute errors ranging from 0.22 to 0.59 mm and from 0.23 to 0.63 mm, respectively. Conclusion: It can be concluded that both scanners are accurate, although the linear measurements are underestimations, with no significant differences between the evaluators.</description><issn>2233-7822</issn><issn>2233-7830</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>JDI</sourceid><recordid>eNqNirtOAzEQRS0EEhHJP0xDudKunazTIgTiUdDQR-PskEzwYzX2Irbj03GBUnObe-7jQi20NqaxW9Nenlnra7XK-dRWbfTW9t1C_TzHIphcJvkiAYwDcCwk50ZolDRMe3bsucx1Bc-RUCAQ5kkoUCwZUgT8ZvTAAQ9UsytYbwO4GfYpUuMIQ6UwTqW2JYV0EByP81JdfaDPtPrzG3X7-PB-_9R8ci68i0P2u5e71zfddrYz_draXm_6tfnv7xcbr1KL</recordid><startdate>2017</startdate><enddate>2017</enddate><creator>Silva, Nathalia Cristine da</creator><creator>Barriviera, Mauricio</creator><creator>Junqueira, Jose Luiz Cintra</creator><creator>Panzarella, Francine Kuhl</creator><creator>Raitz, Ricardo</creator><scope>JDI</scope></search><sort><creationdate>2017</creationdate><title>Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography</title><author>Silva, Nathalia Cristine da ; Barriviera, Mauricio ; Junqueira, Jose Luiz Cintra ; Panzarella, Francine Kuhl ; Raitz, Ricardo</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-kisti_ndsl_JAKO2017136477625643</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>kor</language><creationdate>2017</creationdate><toplevel>online_resources</toplevel><creatorcontrib>Silva, Nathalia Cristine da</creatorcontrib><creatorcontrib>Barriviera, Mauricio</creatorcontrib><creatorcontrib>Junqueira, Jose Luiz Cintra</creatorcontrib><creatorcontrib>Panzarella, Francine Kuhl</creatorcontrib><creatorcontrib>Raitz, Ricardo</creatorcontrib><collection>KoreaScience</collection><jtitle>Imaging science in dentistry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Silva, Nathalia Cristine da</au><au>Barriviera, Mauricio</au><au>Junqueira, Jose Luiz Cintra</au><au>Panzarella, Francine Kuhl</au><au>Raitz, Ricardo</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography</atitle><jtitle>Imaging science in dentistry</jtitle><addtitle>Imaging science in dentistry</addtitle><date>2017</date><risdate>2017</risdate><volume>47</volume><issue>1</issue><spage>11</spage><epage>15</epage><pages>11-15</pages><issn>2233-7822</issn><eissn>2233-7830</eissn><abstract>Purpose: This study was performed to investigate the intra- and inter-observer variability in linear measurements with axial images obtained by PreXion (PreXion Inc., San Mateo, USA) and i-CAT(Imaging Sciences International, Xoran Technologies Inc., Hatfield, USA) CBCT scanners, with different voxel sizes. Materials and Methods: A cylindrical object made from nylon with radiopaque markers (phantom) was scanned by i-CAT and PreXion 3D devices. For each axial image, measurements were taken twice in the horizontal(distance A-B) and vertical (distance C-D) directions, randomly, with a one-week interval between measurements, by four oral radiologists with five years or more experience in the use of these measuring tools. Results: All of the obtained linear measurements had lower values than those of the phantom. The statistical analysis showed high intra- and inter-observer reliability (p=0.297). Compared to the real measurements, the measurements obtained using the i-CAT device and PreXion tomography, on average, revealed absolute errors ranging from 0.22 to 0.59 mm and from 0.23 to 0.63 mm, respectively. Conclusion: It can be concluded that both scanners are accurate, although the linear measurements are underestimations, with no significant differences between the evaluators.</abstract><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 2233-7822 |
ispartof | Imaging science in dentistry, 2017, Vol.47 (1), p.11-15 |
issn | 2233-7822 2233-7830 |
language | kor |
recordid | cdi_kisti_ndsl_JAKO201713647762564 |
source | KoreaMed Synapse; KoreaMed Open Access; DOAJ Directory of Open Access Journals; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; PubMed Central; PubMed Central Open Access |
title | Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-13T14%3A42%3A10IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-kisti&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Intraobserver%20and%20interobserver%20reproducibility%20in%20linear%20measurements%20on%20axial%20images%20obtained%20by%20cone-beam%20computed%20tomography&rft.jtitle=Imaging%20science%20in%20dentistry&rft.au=Silva,%20Nathalia%20Cristine%20da&rft.date=2017&rft.volume=47&rft.issue=1&rft.spage=11&rft.epage=15&rft.pages=11-15&rft.issn=2233-7822&rft.eissn=2233-7830&rft_id=info:doi/&rft_dat=%3Ckisti%3EJAKO201713647762564%3C/kisti%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |