Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors

The present study aimed to dosimetrically evaluate the small-fields of a 6 MV flattening filter-free (FFF) photon beam using different detectors.The 6 MV FFF photon beam was used for measurement of output factor, depth dose, and beam profile of small-fields of sizes 0.6 cm × 0.6 cm to 6.0 cm × 6.0 c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical physics & engineering express 2021-07, Vol.7 (4), p.45004
Hauptverfasser: Dwivedi, Shekhar, Kansal, Sandeep, Dangwal, Vinod Kumar, Bharati, Avinav, Shukla, Jooli
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue 4
container_start_page 45004
container_title Biomedical physics & engineering express
container_volume 7
creator Dwivedi, Shekhar
Kansal, Sandeep
Dangwal, Vinod Kumar
Bharati, Avinav
Shukla, Jooli
description The present study aimed to dosimetrically evaluate the small-fields of a 6 MV flattening filter-free (FFF) photon beam using different detectors.The 6 MV FFF photon beam was used for measurement of output factor, depth dose, and beam profile of small-fields of sizes 0.6 cm × 0.6 cm to 6.0 cm × 6.0 cm. The five detectors used were SNC125c, PinPoint, EDGE, EBT3, and TLD-100. All measurements were performed as per the International Atomic Energy Agency TRS 483 protocol. Output factors measured using different detectors as direct reading ratios showed significant variation for the smallest fields, whereas after correcting them according to TRS 483, all sets of output factors were nearly compatible with each other when measurement uncertainty was also considered. The beam profile measured using SNC125c showed the largest penumbra for all field sizes, whereas the smallest was recorded with EDGE. Compared with that of EBT3, the surface dose was found to be much higher for all the other detectors. PinPoint, EBT3, TLD-100, and EDGE were found to be the detector of choice for small-field output factor measurements; however, PinPoint needs special attention when used for the smallest field size (0.6 cm × 0.6 cm). EDGE and EBT3 are optimal for measuring beam profiles. EBT3, PinPoint, and EDGE can be selected for depth dose measurements, and EBT3 is suitable for surface dose estimation.
doi_str_mv 10.1088/2057-1976/abfd80
format Article
fullrecord <record><control><sourceid>proquest_iop_j</sourceid><recordid>TN_cdi_iop_journals_10_1088_2057_1976_abfd80</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2520870180</sourcerecordid><originalsourceid>FETCH-LOGICAL-c368t-456c33125db8bdbb4887170728881ad351ce518cef32f0c72aaff06ba8c7dd8f3</originalsourceid><addsrcrecordid>eNp9kE1P3DAQhq2qqCDYe0-Vb-2BwDiOY--xouVDAnEBrpY_xm1QEqe2g-Dfk9VuUQ8VpxmNnnk18xDymcEJA6VOaxCyYmvZnhobvIIP5OBt9PGffp-scn4EANbWbbsWn8g-52sOSooDcvcj5m7Akl5oDNTQlt480NCbUnDsxl80dH3BVIWESPNg-p5Ov2OJI7VoBjrnDfNkUhfnTD0WdCWmfET2gukzrnb1kNyf_7w7u6yuby-uzr5fV463qlSNaB3nrBbeKuutbZSSTIKslVLMeC6YQ8GUw8DrAE7WxoQArTXKSe9V4Ifk2zZ3SvHPjLnoocsO-96MuByka1EvXwJTsKCwRV2KOScMekrdYNKLZqA3OvXGl9740ludy8qXXfpsB_RvC3_lLcDxFujipB_jnMbl2ffyvv4HtxM-a6kbDY0AaPTkA38FAeyMJg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2520870180</pqid></control><display><type>article</type><title>Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors</title><source>MEDLINE</source><source>Institute of Physics Journals</source><creator>Dwivedi, Shekhar ; Kansal, Sandeep ; Dangwal, Vinod Kumar ; Bharati, Avinav ; Shukla, Jooli</creator><creatorcontrib>Dwivedi, Shekhar ; Kansal, Sandeep ; Dangwal, Vinod Kumar ; Bharati, Avinav ; Shukla, Jooli</creatorcontrib><description>The present study aimed to dosimetrically evaluate the small-fields of a 6 MV flattening filter-free (FFF) photon beam using different detectors.The 6 MV FFF photon beam was used for measurement of output factor, depth dose, and beam profile of small-fields of sizes 0.6 cm × 0.6 cm to 6.0 cm × 6.0 cm. The five detectors used were SNC125c, PinPoint, EDGE, EBT3, and TLD-100. All measurements were performed as per the International Atomic Energy Agency TRS 483 protocol. Output factors measured using different detectors as direct reading ratios showed significant variation for the smallest fields, whereas after correcting them according to TRS 483, all sets of output factors were nearly compatible with each other when measurement uncertainty was also considered. The beam profile measured using SNC125c showed the largest penumbra for all field sizes, whereas the smallest was recorded with EDGE. Compared with that of EBT3, the surface dose was found to be much higher for all the other detectors. PinPoint, EBT3, TLD-100, and EDGE were found to be the detector of choice for small-field output factor measurements; however, PinPoint needs special attention when used for the smallest field size (0.6 cm × 0.6 cm). EDGE and EBT3 are optimal for measuring beam profiles. EBT3, PinPoint, and EDGE can be selected for depth dose measurements, and EBT3 is suitable for surface dose estimation.</description><identifier>ISSN: 2057-1976</identifier><identifier>EISSN: 2057-1976</identifier><identifier>DOI: 10.1088/2057-1976/abfd80</identifier><identifier>PMID: 33930875</identifier><identifier>CODEN: NJOPFM</identifier><language>eng</language><publisher>England: IOP Publishing</publisher><subject>detector ; FFF photon beam ; International Agencies ; Nuclear Energy ; output factor ; Photons ; Radiometry ; small-field dosimetry ; Uncertainty</subject><ispartof>Biomedical physics &amp; engineering express, 2021-07, Vol.7 (4), p.45004</ispartof><rights>2021 IOP Publishing Ltd</rights><rights>2021 IOP Publishing Ltd.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c368t-456c33125db8bdbb4887170728881ad351ce518cef32f0c72aaff06ba8c7dd8f3</citedby><cites>FETCH-LOGICAL-c368t-456c33125db8bdbb4887170728881ad351ce518cef32f0c72aaff06ba8c7dd8f3</cites><orcidid>0000-0001-8047-8036</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://iopscience.iop.org/article/10.1088/2057-1976/abfd80/pdf$$EPDF$$P50$$Giop$$H</linktopdf><link.rule.ids>314,780,784,27922,27923,53844,53891</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33930875$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Dwivedi, Shekhar</creatorcontrib><creatorcontrib>Kansal, Sandeep</creatorcontrib><creatorcontrib>Dangwal, Vinod Kumar</creatorcontrib><creatorcontrib>Bharati, Avinav</creatorcontrib><creatorcontrib>Shukla, Jooli</creatorcontrib><title>Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors</title><title>Biomedical physics &amp; engineering express</title><addtitle>BPEX</addtitle><addtitle>Biomed. Phys. Eng. Express</addtitle><description>The present study aimed to dosimetrically evaluate the small-fields of a 6 MV flattening filter-free (FFF) photon beam using different detectors.The 6 MV FFF photon beam was used for measurement of output factor, depth dose, and beam profile of small-fields of sizes 0.6 cm × 0.6 cm to 6.0 cm × 6.0 cm. The five detectors used were SNC125c, PinPoint, EDGE, EBT3, and TLD-100. All measurements were performed as per the International Atomic Energy Agency TRS 483 protocol. Output factors measured using different detectors as direct reading ratios showed significant variation for the smallest fields, whereas after correcting them according to TRS 483, all sets of output factors were nearly compatible with each other when measurement uncertainty was also considered. The beam profile measured using SNC125c showed the largest penumbra for all field sizes, whereas the smallest was recorded with EDGE. Compared with that of EBT3, the surface dose was found to be much higher for all the other detectors. PinPoint, EBT3, TLD-100, and EDGE were found to be the detector of choice for small-field output factor measurements; however, PinPoint needs special attention when used for the smallest field size (0.6 cm × 0.6 cm). EDGE and EBT3 are optimal for measuring beam profiles. EBT3, PinPoint, and EDGE can be selected for depth dose measurements, and EBT3 is suitable for surface dose estimation.</description><subject>detector</subject><subject>FFF photon beam</subject><subject>International Agencies</subject><subject>Nuclear Energy</subject><subject>output factor</subject><subject>Photons</subject><subject>Radiometry</subject><subject>small-field dosimetry</subject><subject>Uncertainty</subject><issn>2057-1976</issn><issn>2057-1976</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kE1P3DAQhq2qqCDYe0-Vb-2BwDiOY--xouVDAnEBrpY_xm1QEqe2g-Dfk9VuUQ8VpxmNnnk18xDymcEJA6VOaxCyYmvZnhobvIIP5OBt9PGffp-scn4EANbWbbsWn8g-52sOSooDcvcj5m7Akl5oDNTQlt480NCbUnDsxl80dH3BVIWESPNg-p5Ov2OJI7VoBjrnDfNkUhfnTD0WdCWmfET2gukzrnb1kNyf_7w7u6yuby-uzr5fV463qlSNaB3nrBbeKuutbZSSTIKslVLMeC6YQ8GUw8DrAE7WxoQArTXKSe9V4Ifk2zZ3SvHPjLnoocsO-96MuByka1EvXwJTsKCwRV2KOScMekrdYNKLZqA3OvXGl9740ludy8qXXfpsB_RvC3_lLcDxFujipB_jnMbl2ffyvv4HtxM-a6kbDY0AaPTkA38FAeyMJg</recordid><startdate>20210701</startdate><enddate>20210701</enddate><creator>Dwivedi, Shekhar</creator><creator>Kansal, Sandeep</creator><creator>Dangwal, Vinod Kumar</creator><creator>Bharati, Avinav</creator><creator>Shukla, Jooli</creator><general>IOP Publishing</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0001-8047-8036</orcidid></search><sort><creationdate>20210701</creationdate><title>Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors</title><author>Dwivedi, Shekhar ; Kansal, Sandeep ; Dangwal, Vinod Kumar ; Bharati, Avinav ; Shukla, Jooli</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c368t-456c33125db8bdbb4887170728881ad351ce518cef32f0c72aaff06ba8c7dd8f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>detector</topic><topic>FFF photon beam</topic><topic>International Agencies</topic><topic>Nuclear Energy</topic><topic>output factor</topic><topic>Photons</topic><topic>Radiometry</topic><topic>small-field dosimetry</topic><topic>Uncertainty</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Dwivedi, Shekhar</creatorcontrib><creatorcontrib>Kansal, Sandeep</creatorcontrib><creatorcontrib>Dangwal, Vinod Kumar</creatorcontrib><creatorcontrib>Bharati, Avinav</creatorcontrib><creatorcontrib>Shukla, Jooli</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Biomedical physics &amp; engineering express</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Dwivedi, Shekhar</au><au>Kansal, Sandeep</au><au>Dangwal, Vinod Kumar</au><au>Bharati, Avinav</au><au>Shukla, Jooli</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors</atitle><jtitle>Biomedical physics &amp; engineering express</jtitle><stitle>BPEX</stitle><addtitle>Biomed. Phys. Eng. Express</addtitle><date>2021-07-01</date><risdate>2021</risdate><volume>7</volume><issue>4</issue><spage>45004</spage><pages>45004-</pages><issn>2057-1976</issn><eissn>2057-1976</eissn><coden>NJOPFM</coden><abstract>The present study aimed to dosimetrically evaluate the small-fields of a 6 MV flattening filter-free (FFF) photon beam using different detectors.The 6 MV FFF photon beam was used for measurement of output factor, depth dose, and beam profile of small-fields of sizes 0.6 cm × 0.6 cm to 6.0 cm × 6.0 cm. The five detectors used were SNC125c, PinPoint, EDGE, EBT3, and TLD-100. All measurements were performed as per the International Atomic Energy Agency TRS 483 protocol. Output factors measured using different detectors as direct reading ratios showed significant variation for the smallest fields, whereas after correcting them according to TRS 483, all sets of output factors were nearly compatible with each other when measurement uncertainty was also considered. The beam profile measured using SNC125c showed the largest penumbra for all field sizes, whereas the smallest was recorded with EDGE. Compared with that of EBT3, the surface dose was found to be much higher for all the other detectors. PinPoint, EBT3, TLD-100, and EDGE were found to be the detector of choice for small-field output factor measurements; however, PinPoint needs special attention when used for the smallest field size (0.6 cm × 0.6 cm). EDGE and EBT3 are optimal for measuring beam profiles. EBT3, PinPoint, and EDGE can be selected for depth dose measurements, and EBT3 is suitable for surface dose estimation.</abstract><cop>England</cop><pub>IOP Publishing</pub><pmid>33930875</pmid><doi>10.1088/2057-1976/abfd80</doi><tpages>14</tpages><orcidid>https://orcid.org/0000-0001-8047-8036</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 2057-1976
ispartof Biomedical physics & engineering express, 2021-07, Vol.7 (4), p.45004
issn 2057-1976
2057-1976
language eng
recordid cdi_iop_journals_10_1088_2057_1976_abfd80
source MEDLINE; Institute of Physics Journals
subjects detector
FFF photon beam
International Agencies
Nuclear Energy
output factor
Photons
Radiometry
small-field dosimetry
Uncertainty
title Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-10T08%3A07%3A49IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_iop_j&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Dosimetry%20of%20a%206%20MV%20flattening%20filter-free%20small%20photon%20beam%20using%20various%20detectors&rft.jtitle=Biomedical%20physics%20&%20engineering%20express&rft.au=Dwivedi,%20Shekhar&rft.date=2021-07-01&rft.volume=7&rft.issue=4&rft.spage=45004&rft.pages=45004-&rft.issn=2057-1976&rft.eissn=2057-1976&rft.coden=NJOPFM&rft_id=info:doi/10.1088/2057-1976/abfd80&rft_dat=%3Cproquest_iop_j%3E2520870180%3C/proquest_iop_j%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2520870180&rft_id=info:pmid/33930875&rfr_iscdi=true