First measurements of a 512 PSAPD prototype of a sub-mm resolution clinical PET camera
We present measurements from a first prototype of a 1 mm3 resolution clinical PET camera dedicated to breast imaging. This first prototype consists of two cartridges of 8 detector layers. Each detector layer comprises 16 dual LYSO-PSAPD modules arranged side by side. The LYSO scintillator is an 8×8...
Gespeichert in:
Hauptverfasser: | , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present measurements from a first prototype of a 1 mm3 resolution clinical PET camera dedicated to breast imaging. This first prototype consists of two cartridges of 8 detector layers. Each detector layer comprises 16 dual LYSO-PSAPD modules arranged side by side. The LYSO scintillator is an 8×8 array of 0.9×0.9×1 mm3 crystal pixels. Each array is coupled to a PSAPD. The modules are oriented edge-on with respect to the incoming 511 keV photons. The two cartridges thus contain 256 PSAPDs and 16,284 crystals each. The entire camera will have 2 imaging heads of 9 cartridges each. Charge created by the PSAPDs was routed towards a signal conditioning board that was connected to another circuit board that houses the RENA-3 ASIC. The RENA3 chip contains 36 channels each with preamplifier, leading edge discriminator and sample and hold circuitry. The RENA3-boards were connected to a DAQ board that linked to a host PC using 8 USB2 connections. We measured an energy resolution of 10.62 + 0.04% FWHM at 511 keV, together with a coincidence time resolution of 15.7 + 0.2 ns FWHM. We determined a crystal misidentification probability of |
---|---|
ISSN: | 1082-3654 2577-0829 |
DOI: | 10.1109/NSSMIC.2013.6829406 |