A SPECT camera for simultaneous SPECT/MRI
We describe a single photon emission computed tomograph (SPECT) which can be operated inside state-of-the-art magnetic resonance imaging (MRI) systems. The combined SPECT/MRI system allows one to acquire simultaneously the data from both modalities and co-register the images in space and time. The n...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe a single photon emission computed tomograph (SPECT) which can be operated inside state-of-the-art magnetic resonance imaging (MRI) systems. The combined SPECT/MRI system allows one to acquire simultaneously the data from both modalities and co-register the images in space and time. The new SPECT is based on the semiconductor cadmium zinc telluride (CZT) and application specific integrated circuits (ASICs) - two technologies which are MR-compatible and almost insensitive to magnetic fields. The SPECT camera has an energy resolution of 5.4 keV full-width-half-maximum (FWHM) at the 140-keV photo peak from 99m Tc for either inside or outside of the MRI. We acquired the first SPECT and MR images from resolution phantoms and mice in-vivo. The experiments show that the MRI is minimally affected by the SPECT camera inside the bore. Preliminary results for the SPECT camera show high sensitivity of about 1.6% and a spatial resolution of 4 mm FWHM. |
---|---|
ISSN: | 1082-3654 2577-0829 |
DOI: | 10.1109/NSSMIC.2009.5402269 |