Detection of Low Energy Heavy Particles with Silicon Barrier Detectors
The performance characteristics of silicon surface barrier detectors have been investigated for detecting H+, He+, N+, Ne+ and Ar+ in the energy interval 6-100 keV. For a cryogenically cooled, 50 mm2, 1325 μ depletion depth detector the resolution for protons was 1.65 keV (FWHM). The resolution decr...
Gespeichert in:
Veröffentlicht in: | IEEE (Inst. Elec. Electron. Eng.), Trans. Nucl. Sci. NS-16: No. 1, 82-6(Feb 1969) Trans. Nucl. Sci. NS-16: No. 1, 82-6(Feb 1969), 1969-02, Vol.16 (1), p.82-86 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The performance characteristics of silicon surface barrier detectors have been investigated for detecting H+, He+, N+, Ne+ and Ar+ in the energy interval 6-100 keV. For a cryogenically cooled, 50 mm2, 1325 μ depletion depth detector the resolution for protons was 1.65 keV (FWHM). The resolution decreased as the particle mass increased being 6.5 keV for 40 keV Ar+. All particles exhibit pulse height linearity throughout the energy range. |
---|---|
ISSN: | 0018-9499 1558-1578 |
DOI: | 10.1109/TNS.1969.4325084 |