Intrinsic Spatial Resolution and Parallax Correction Using Depth-Encoding PET Detector Modules Based on Position-Sensitive APD Readout

We are developing PET detectors with depth of interaction (DOI) capability based on lutetium oxyorthosilicate (LSO) scintillator arrays coupled at both ends to position-sensitive avalanche photodiodes (PSAPDs). The detector module consists of a 5x5 block of LSO crystals (1.5times1.5times20 mm 3 ) co...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science 2006-10, Vol.53 (5), p.2666-2670
Hauptverfasser: Dokhale, P.A., Silverman, R.W., Shah, K.S., Farrell, R., McClish, M.A., Entine, G., Cherry, S.R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We are developing PET detectors with depth of interaction (DOI) capability based on lutetium oxyorthosilicate (LSO) scintillator arrays coupled at both ends to position-sensitive avalanche photodiodes (PSAPDs). The detector module consists of a 5x5 block of LSO crystals (1.5times1.5times20 mm 3 ) coupled to 8times8 mm 2 PSAPDs at both ends. We present the intrinsic spatial resolution for two complete modules and the radial resolution component obtained with and without parallax correction using DOI information. DOI resolution was measured to be ~3 mm. Intrinsic spatial resolution (FWHM) averaged 1.15 mm for opposing crystal pairs. The detectors were then rotated and moved closer together consistent with a 16 cm diameter cylindrical geometry scanner. Resolution was measured at positions corresponding to a radial offset of 4.7 cm and 7 cm, with the annihilation photons incident on the detector surface at 30deg and 52deg respectively. At an angle of 52deg the intrinsic spatial resolution was degraded to 7.7 mm. By incorporating DOI information, the measured spatial resolution (FWHM) was improved to 1.8 mm. This demonstrates that approximately 90% of the resolution degradation due to the parallax error can be removed using these depth-encoding detectors for a field of view with a diameter that is 87.5% of the detector separation
ISSN:0018-9499
1558-1578
DOI:10.1109/TNS.2006.882807