A precision set-up and method for calibrating a DC voltage divider's ratios from 10 V to 1000 V
A precision set-up and method for calibrating the division ratios of a resistive divider for DC-voltage in the range 10 V to 1000 V is described. The method requires equipment commonly available in most standard laboratories. Uncertainties better than 3/spl times/10/sup -7/ (k = 2) in the ratio is a...
Gespeichert in:
Hauptverfasser: | , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A precision set-up and method for calibrating the division ratios of a resistive divider for DC-voltage in the range 10 V to 1000 V is described. The method requires equipment commonly available in most standard laboratories. Uncertainties better than 3/spl times/10/sup -7/ (k = 2) in the ratio is achieved. |
---|---|
DOI: | 10.1109/CPEM.2002.1034877 |