Storage stability of commonly used haematological parameters at 33 °C
Introduction: This study aimed to investigate the analytical bias in haematological parameters induced by storage at 33 ºC. Materials and methods: Blood from the diversion pouch of 20 blood donors were collected in K2EDTA vials and stored at 33 ºC. Readings from each vial were taken at 0, 4, 6, 12,...
Gespeichert in:
Veröffentlicht in: | Biochemia Medica 2018-06, Vol.28 (2), p.347 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | |
---|---|
container_issue | 2 |
container_start_page | 347 |
container_title | Biochemia Medica |
container_volume | 28 |
creator | Jain, Ashish Jain, Sanchit Singh, Neha Aswal, Priyanka Pal, Shweta Meinia, Sushant Kumar Chowdhury, Nilotpal |
description | Introduction: This study aimed to investigate the analytical bias in haematological parameters induced by storage at 33 ºC.
Materials and methods: Blood from the diversion pouch of 20 blood donors were collected in K2EDTA vials and stored at 33 ºC. Readings from
each vial were taken at 0, 4, 6, 12, 24, 48 and 72 hours after collection on the Sysmex XP-100 analyser (Sysmex Corporation, Kobe, Japan). The percent
difference from the baseline readings were calculated and subjected to a Wilcoxon signed rank test at a Holm corrected significance level of 0.05. A
median percent difference, which was statistically significant and greater than the maximum acceptable bias (taken from studies of biological variation),
was taken as evidence of unacceptable shift. If the median shift was lesser than the maximum acceptable bias, two one-sided Wilcoxon signed
rank tests for equivalence were used to determine whether the percent differences were significantly lesser than the maximum acceptable bias.
Results: Haemoglobin, red blood cell count, white blood cell count, mean corpuscular haemoglobin and lymphocyte count showed acceptable bias
after storage for at least 24 hours at 33 ºC. Haematocrit, mean corpuscular volume, mean corpuscular haemoglobin concentration, platelet count and
mean platelet volume showed unacceptable shift in less than 4 hours when stored at 33 ºC.
Conclusions: Since many haematological parameters show unacceptable bias within 4 hours of sample storage at 33 ºC, the recommended limit of
time from collection to processing should be revised for areas where high environmental temperatures are common. |
doi_str_mv | 10.11613/BM.2018.020901 |
format | Article |
fullrecord | <record><control><sourceid>crossref_hrcak</sourceid><recordid>TN_cdi_hrcak_primary_oai_hrcak_srce_hr_199000</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>10_11613_BM_2018_020901</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2351-d3c08ba3d0d540035cfc986274d7e23433227acbe53a423bc2c0f1437c7dce43</originalsourceid><addsrcrecordid>eNo9kEFOwzAURC0EEqWwZusLpP32t-NkSSsKSK1Y0L314zhtIMGVHRa9FWfgZBSCWM1oNDOLx9itgJkQucD5YjOTIIoZSChBnLGJKFSeGVXI85NHhAzKXF6yq5ReAbTWhZmw1csQIu08TwNVbdcORx4a7kLfh_fuyD-Sr_mefE9D6MKuddTxA0Xq_eBj4jRwRP71ubxmFw11yd_86ZRtV_fb5WO2fn54Wt6tMydRi6xGB0VFWEOtFQBq17iyyKVRtfESFaKUhlzlNZKSWDnpoBEKjTO18wqnLBtv99HRmz3Etqd4tIFaOyYpOn-yVpQlnP6nbD72XQwpRd_8TwTYX2p2sbE_1OxIDb8BP0RgCA</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Storage stability of commonly used haematological parameters at 33 °C</title><source>DOAJ Directory of Open Access Journals</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>PubMed Central Open Access</source><source>PubMed Central</source><creator>Jain, Ashish ; Jain, Sanchit ; Singh, Neha ; Aswal, Priyanka ; Pal, Shweta ; Meinia, Sushant Kumar ; Chowdhury, Nilotpal</creator><creatorcontrib>Jain, Ashish ; Jain, Sanchit ; Singh, Neha ; Aswal, Priyanka ; Pal, Shweta ; Meinia, Sushant Kumar ; Chowdhury, Nilotpal</creatorcontrib><description>Introduction: This study aimed to investigate the analytical bias in haematological parameters induced by storage at 33 ºC.
Materials and methods: Blood from the diversion pouch of 20 blood donors were collected in K2EDTA vials and stored at 33 ºC. Readings from
each vial were taken at 0, 4, 6, 12, 24, 48 and 72 hours after collection on the Sysmex XP-100 analyser (Sysmex Corporation, Kobe, Japan). The percent
difference from the baseline readings were calculated and subjected to a Wilcoxon signed rank test at a Holm corrected significance level of 0.05. A
median percent difference, which was statistically significant and greater than the maximum acceptable bias (taken from studies of biological variation),
was taken as evidence of unacceptable shift. If the median shift was lesser than the maximum acceptable bias, two one-sided Wilcoxon signed
rank tests for equivalence were used to determine whether the percent differences were significantly lesser than the maximum acceptable bias.
Results: Haemoglobin, red blood cell count, white blood cell count, mean corpuscular haemoglobin and lymphocyte count showed acceptable bias
after storage for at least 24 hours at 33 ºC. Haematocrit, mean corpuscular volume, mean corpuscular haemoglobin concentration, platelet count and
mean platelet volume showed unacceptable shift in less than 4 hours when stored at 33 ºC.
Conclusions: Since many haematological parameters show unacceptable bias within 4 hours of sample storage at 33 ºC, the recommended limit of
time from collection to processing should be revised for areas where high environmental temperatures are common.</description><identifier>ISSN: 1330-0962</identifier><identifier>EISSN: 1846-7482</identifier><identifier>DOI: 10.11613/BM.2018.020901</identifier><language>eng</language><publisher>Medicinska naklada</publisher><subject>erythrocyte indices ; laboratory errors ; platelet count ; pre-analytical error ; sample stability</subject><ispartof>Biochemia Medica, 2018-06, Vol.28 (2), p.347</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c2351-d3c08ba3d0d540035cfc986274d7e23433227acbe53a423bc2c0f1437c7dce43</citedby><cites>FETCH-LOGICAL-c2351-d3c08ba3d0d540035cfc986274d7e23433227acbe53a423bc2c0f1437c7dce43</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,864,885,27924,27925</link.rule.ids></links><search><creatorcontrib>Jain, Ashish</creatorcontrib><creatorcontrib>Jain, Sanchit</creatorcontrib><creatorcontrib>Singh, Neha</creatorcontrib><creatorcontrib>Aswal, Priyanka</creatorcontrib><creatorcontrib>Pal, Shweta</creatorcontrib><creatorcontrib>Meinia, Sushant Kumar</creatorcontrib><creatorcontrib>Chowdhury, Nilotpal</creatorcontrib><title>Storage stability of commonly used haematological parameters at 33 °C</title><title>Biochemia Medica</title><description>Introduction: This study aimed to investigate the analytical bias in haematological parameters induced by storage at 33 ºC.
Materials and methods: Blood from the diversion pouch of 20 blood donors were collected in K2EDTA vials and stored at 33 ºC. Readings from
each vial were taken at 0, 4, 6, 12, 24, 48 and 72 hours after collection on the Sysmex XP-100 analyser (Sysmex Corporation, Kobe, Japan). The percent
difference from the baseline readings were calculated and subjected to a Wilcoxon signed rank test at a Holm corrected significance level of 0.05. A
median percent difference, which was statistically significant and greater than the maximum acceptable bias (taken from studies of biological variation),
was taken as evidence of unacceptable shift. If the median shift was lesser than the maximum acceptable bias, two one-sided Wilcoxon signed
rank tests for equivalence were used to determine whether the percent differences were significantly lesser than the maximum acceptable bias.
Results: Haemoglobin, red blood cell count, white blood cell count, mean corpuscular haemoglobin and lymphocyte count showed acceptable bias
after storage for at least 24 hours at 33 ºC. Haematocrit, mean corpuscular volume, mean corpuscular haemoglobin concentration, platelet count and
mean platelet volume showed unacceptable shift in less than 4 hours when stored at 33 ºC.
Conclusions: Since many haematological parameters show unacceptable bias within 4 hours of sample storage at 33 ºC, the recommended limit of
time from collection to processing should be revised for areas where high environmental temperatures are common.</description><subject>erythrocyte indices</subject><subject>laboratory errors</subject><subject>platelet count</subject><subject>pre-analytical error</subject><subject>sample stability</subject><issn>1330-0962</issn><issn>1846-7482</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><recordid>eNo9kEFOwzAURC0EEqWwZusLpP32t-NkSSsKSK1Y0L314zhtIMGVHRa9FWfgZBSCWM1oNDOLx9itgJkQucD5YjOTIIoZSChBnLGJKFSeGVXI85NHhAzKXF6yq5ReAbTWhZmw1csQIu08TwNVbdcORx4a7kLfh_fuyD-Sr_mefE9D6MKuddTxA0Xq_eBj4jRwRP71ubxmFw11yd_86ZRtV_fb5WO2fn54Wt6tMydRi6xGB0VFWEOtFQBq17iyyKVRtfESFaKUhlzlNZKSWDnpoBEKjTO18wqnLBtv99HRmz3Etqd4tIFaOyYpOn-yVpQlnP6nbD72XQwpRd_8TwTYX2p2sbE_1OxIDb8BP0RgCA</recordid><startdate>20180615</startdate><enddate>20180615</enddate><creator>Jain, Ashish</creator><creator>Jain, Sanchit</creator><creator>Singh, Neha</creator><creator>Aswal, Priyanka</creator><creator>Pal, Shweta</creator><creator>Meinia, Sushant Kumar</creator><creator>Chowdhury, Nilotpal</creator><general>Medicinska naklada</general><scope>AAYXX</scope><scope>CITATION</scope><scope>VP8</scope></search><sort><creationdate>20180615</creationdate><title>Storage stability of commonly used haematological parameters at 33 °C</title><author>Jain, Ashish ; Jain, Sanchit ; Singh, Neha ; Aswal, Priyanka ; Pal, Shweta ; Meinia, Sushant Kumar ; Chowdhury, Nilotpal</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2351-d3c08ba3d0d540035cfc986274d7e23433227acbe53a423bc2c0f1437c7dce43</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>erythrocyte indices</topic><topic>laboratory errors</topic><topic>platelet count</topic><topic>pre-analytical error</topic><topic>sample stability</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Jain, Ashish</creatorcontrib><creatorcontrib>Jain, Sanchit</creatorcontrib><creatorcontrib>Singh, Neha</creatorcontrib><creatorcontrib>Aswal, Priyanka</creatorcontrib><creatorcontrib>Pal, Shweta</creatorcontrib><creatorcontrib>Meinia, Sushant Kumar</creatorcontrib><creatorcontrib>Chowdhury, Nilotpal</creatorcontrib><collection>CrossRef</collection><collection>Hrcak: Portal of scientific journals of Croatia</collection><jtitle>Biochemia Medica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Jain, Ashish</au><au>Jain, Sanchit</au><au>Singh, Neha</au><au>Aswal, Priyanka</au><au>Pal, Shweta</au><au>Meinia, Sushant Kumar</au><au>Chowdhury, Nilotpal</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Storage stability of commonly used haematological parameters at 33 °C</atitle><jtitle>Biochemia Medica</jtitle><date>2018-06-15</date><risdate>2018</risdate><volume>28</volume><issue>2</issue><spage>347</spage><pages>347-</pages><issn>1330-0962</issn><eissn>1846-7482</eissn><abstract>Introduction: This study aimed to investigate the analytical bias in haematological parameters induced by storage at 33 ºC.
Materials and methods: Blood from the diversion pouch of 20 blood donors were collected in K2EDTA vials and stored at 33 ºC. Readings from
each vial were taken at 0, 4, 6, 12, 24, 48 and 72 hours after collection on the Sysmex XP-100 analyser (Sysmex Corporation, Kobe, Japan). The percent
difference from the baseline readings were calculated and subjected to a Wilcoxon signed rank test at a Holm corrected significance level of 0.05. A
median percent difference, which was statistically significant and greater than the maximum acceptable bias (taken from studies of biological variation),
was taken as evidence of unacceptable shift. If the median shift was lesser than the maximum acceptable bias, two one-sided Wilcoxon signed
rank tests for equivalence were used to determine whether the percent differences were significantly lesser than the maximum acceptable bias.
Results: Haemoglobin, red blood cell count, white blood cell count, mean corpuscular haemoglobin and lymphocyte count showed acceptable bias
after storage for at least 24 hours at 33 ºC. Haematocrit, mean corpuscular volume, mean corpuscular haemoglobin concentration, platelet count and
mean platelet volume showed unacceptable shift in less than 4 hours when stored at 33 ºC.
Conclusions: Since many haematological parameters show unacceptable bias within 4 hours of sample storage at 33 ºC, the recommended limit of
time from collection to processing should be revised for areas where high environmental temperatures are common.</abstract><pub>Medicinska naklada</pub><doi>10.11613/BM.2018.020901</doi><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1330-0962 |
ispartof | Biochemia Medica, 2018-06, Vol.28 (2), p.347 |
issn | 1330-0962 1846-7482 |
language | eng |
recordid | cdi_hrcak_primary_oai_hrcak_srce_hr_199000 |
source | DOAJ Directory of Open Access Journals; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; PubMed Central Open Access; PubMed Central |
subjects | erythrocyte indices laboratory errors platelet count pre-analytical error sample stability |
title | Storage stability of commonly used haematological parameters at 33 °C |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T22%3A41%3A52IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-crossref_hrcak&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Storage%20stability%20of%20commonly%20used%20haematological%20parameters%20at%2033%20%C2%B0C&rft.jtitle=Biochemia%20Medica&rft.au=Jain,%20Ashish&rft.date=2018-06-15&rft.volume=28&rft.issue=2&rft.spage=347&rft.pages=347-&rft.issn=1330-0962&rft.eissn=1846-7482&rft_id=info:doi/10.11613/BM.2018.020901&rft_dat=%3Ccrossref_hrcak%3E10_11613_BM_2018_020901%3C/crossref_hrcak%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |