Lung ultrasound features and relationships with respiratory mechanics of evolving BPD in preterm rabbits and human neonates
We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relations...
Gespeichert in:
Veröffentlicht in: | Journal of applied physiology (1985) 2021-09, Vol.131 (3), p.895-904 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 904 |
---|---|
container_issue | 3 |
container_start_page | 895 |
container_title | Journal of applied physiology (1985) |
container_volume | 131 |
creator | Loi, Barbara Casiraghi, Costanza Catozzi, Chiara Storti, Matteo Lucattelli, Monica Bartalesi, Barbara Yousef, Nadya Salomone, Fabrizio De Luca, Daniele |
description | We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.
Evolving bronchopulmonary dysplasia (BPD) is characterized by impaired alveolarization leading to lung aeration inhomogeneities. Hyperoxia-exposed preterm rabbits have been proposed to mimic evolving BPD; therefore, we aimed to verify if this model has the same lung ultrasound and mechanical features of evolving BPD in human neonates. Semiquantitative lung ultrasound and lung mechanics measurement was performed in 25 preterm rabbits (28 days of gestation) and 25 neonates (mean gestational age ≈ 26 wk) with evolving BPD. A modified rabbit lung ultrasound score (rLUS) and a validated neonatal lung ultrasound score (LUS) were used. Lung ultrasound images were recorded and evaluated by two independent observers blinded to each other’s evaluation. Lung ultrasound findings were equally heterogeneous both in rabbits as in human neonates and encompassed all the classical lung ultrasound semiology. Lung ultrasound and histology examination were also performed in 13 term rabbits kept under normoxia as further control and showed the absence of ultrasound and histology abnormalities compared with hyperoxia-exposed preterm rabbits. The interrater absolute agreement for the evaluation of lung ultrasound images in rabbits was very high [ICC: 0.989 (95%CI: 0.975–0.995); P < 0.0001], and there was no difference between the two observers. Lung mechanics parameters were similarly altered in both rabbits and human neonates. There were moderately significant correlations between airway resistances and lung ultrasound scores in rabbits ( ρ = 0.519; P = 0.008) and in neonates ( ρ = 0.409; P = 0.042). In conclusion, the preterm rabbit model fairly reproduces the lung ultrasound and mechanical characteristics of preterm neonates with evolving BPD.
NEW & NOTEWORTHY We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal mo |
doi_str_mv | 10.1152/japplphysiol.00300.2021 |
format | Article |
fullrecord | <record><control><sourceid>proquest_hal_p</sourceid><recordid>TN_cdi_hal_primary_oai_HAL_hal_04532551v1</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2570242384</sourcerecordid><originalsourceid>FETCH-LOGICAL-c401t-bcd6de86af2435f3f7ec6687c73cad487fb70937e96f2da91d3dfe4ba1a56ea33</originalsourceid><addsrcrecordid>eNpNkU1r3DAQhkVoIds0vyGCnHrwVp-W95ikHwkstIf2LMb2KNbilRxJ3rL0z9fbLSWnYd55eBl4CLnhbM25Fh93ME3jNByzj-OaMcnYWjDBL8hquYqK14y_IavGaFYZ3ZhL8i7nHWNcKc1X5Pd2Ds90HkuCHOfQU4dQ5oSZwrIkHKH4GPLgp0x_-TIsUZ58ghLTke6xGyD4LtPoKB7iePBL2f33T9QHOiUsmPY0Qdv6cu4b5j0EGjAGKJjfk7cOxozX_-YV-fnl84-Hx2r77evTw9226hTjpWq7vu6xqcEJJbWTzmBX143pjOygV41xrWEbaXBTO9HDhveyd6ha4KBrBCmvyIdz7wCjnZLfQzraCN4-3m3tKWNKS6E1P_CFvT2zU4ovM-Zid3FOYXnPCm2YUEI2aqHMmepSzDmh-1_LmT1Zsa-t2L9W7MmK_AOSnYdJ</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2570242384</pqid></control><display><type>article</type><title>Lung ultrasound features and relationships with respiratory mechanics of evolving BPD in preterm rabbits and human neonates</title><source>American Physiological Society</source><source>EZB-FREE-00999 freely available EZB journals</source><source>Alma/SFX Local Collection</source><creator>Loi, Barbara ; Casiraghi, Costanza ; Catozzi, Chiara ; Storti, Matteo ; Lucattelli, Monica ; Bartalesi, Barbara ; Yousef, Nadya ; Salomone, Fabrizio ; De Luca, Daniele</creator><creatorcontrib>Loi, Barbara ; Casiraghi, Costanza ; Catozzi, Chiara ; Storti, Matteo ; Lucattelli, Monica ; Bartalesi, Barbara ; Yousef, Nadya ; Salomone, Fabrizio ; De Luca, Daniele</creatorcontrib><description>We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.
Evolving bronchopulmonary dysplasia (BPD) is characterized by impaired alveolarization leading to lung aeration inhomogeneities. Hyperoxia-exposed preterm rabbits have been proposed to mimic evolving BPD; therefore, we aimed to verify if this model has the same lung ultrasound and mechanical features of evolving BPD in human neonates. Semiquantitative lung ultrasound and lung mechanics measurement was performed in 25 preterm rabbits (28 days of gestation) and 25 neonates (mean gestational age ≈ 26 wk) with evolving BPD. A modified rabbit lung ultrasound score (rLUS) and a validated neonatal lung ultrasound score (LUS) were used. Lung ultrasound images were recorded and evaluated by two independent observers blinded to each other’s evaluation. Lung ultrasound findings were equally heterogeneous both in rabbits as in human neonates and encompassed all the classical lung ultrasound semiology. Lung ultrasound and histology examination were also performed in 13 term rabbits kept under normoxia as further control and showed the absence of ultrasound and histology abnormalities compared with hyperoxia-exposed preterm rabbits. The interrater absolute agreement for the evaluation of lung ultrasound images in rabbits was very high [ICC: 0.989 (95%CI: 0.975–0.995); P < 0.0001], and there was no difference between the two observers. Lung mechanics parameters were similarly altered in both rabbits and human neonates. There were moderately significant correlations between airway resistances and lung ultrasound scores in rabbits ( ρ = 0.519; P = 0.008) and in neonates ( ρ = 0.409; P = 0.042). In conclusion, the preterm rabbit model fairly reproduces the lung ultrasound and mechanical characteristics of preterm neonates with evolving BPD.
NEW & NOTEWORTHY We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.</description><identifier>ISSN: 8750-7587</identifier><identifier>EISSN: 1522-1601</identifier><identifier>DOI: 10.1152/japplphysiol.00300.2021</identifier><language>eng</language><publisher>Bethesda: American Physiological Society</publisher><subject>Abnormalities ; Aeration ; Animal models ; Dysplasia ; Evaluation ; Evolution ; Exposure ; Gestation ; Gestational age ; Histology ; Hyperoxia ; Life Sciences ; Lungs ; Mechanical properties ; Mechanics ; Mechanics (physics) ; Neonates ; Observers ; Rabbits ; Respiratory tract ; Ultrasonic imaging ; Ultrasound</subject><ispartof>Journal of applied physiology (1985), 2021-09, Vol.131 (3), p.895-904</ispartof><rights>Copyright American Physiological Society Sep 2021</rights><rights>Distributed under a Creative Commons Attribution 4.0 International License</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c401t-bcd6de86af2435f3f7ec6687c73cad487fb70937e96f2da91d3dfe4ba1a56ea33</citedby><cites>FETCH-LOGICAL-c401t-bcd6de86af2435f3f7ec6687c73cad487fb70937e96f2da91d3dfe4ba1a56ea33</cites><orcidid>0000-0001-6756-7237 ; 0000-0002-3846-4834</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,3037,27923,27924</link.rule.ids><backlink>$$Uhttps://hal.science/hal-04532551$$DView record in HAL$$Hfree_for_read</backlink></links><search><creatorcontrib>Loi, Barbara</creatorcontrib><creatorcontrib>Casiraghi, Costanza</creatorcontrib><creatorcontrib>Catozzi, Chiara</creatorcontrib><creatorcontrib>Storti, Matteo</creatorcontrib><creatorcontrib>Lucattelli, Monica</creatorcontrib><creatorcontrib>Bartalesi, Barbara</creatorcontrib><creatorcontrib>Yousef, Nadya</creatorcontrib><creatorcontrib>Salomone, Fabrizio</creatorcontrib><creatorcontrib>De Luca, Daniele</creatorcontrib><title>Lung ultrasound features and relationships with respiratory mechanics of evolving BPD in preterm rabbits and human neonates</title><title>Journal of applied physiology (1985)</title><description>We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.
Evolving bronchopulmonary dysplasia (BPD) is characterized by impaired alveolarization leading to lung aeration inhomogeneities. Hyperoxia-exposed preterm rabbits have been proposed to mimic evolving BPD; therefore, we aimed to verify if this model has the same lung ultrasound and mechanical features of evolving BPD in human neonates. Semiquantitative lung ultrasound and lung mechanics measurement was performed in 25 preterm rabbits (28 days of gestation) and 25 neonates (mean gestational age ≈ 26 wk) with evolving BPD. A modified rabbit lung ultrasound score (rLUS) and a validated neonatal lung ultrasound score (LUS) were used. Lung ultrasound images were recorded and evaluated by two independent observers blinded to each other’s evaluation. Lung ultrasound findings were equally heterogeneous both in rabbits as in human neonates and encompassed all the classical lung ultrasound semiology. Lung ultrasound and histology examination were also performed in 13 term rabbits kept under normoxia as further control and showed the absence of ultrasound and histology abnormalities compared with hyperoxia-exposed preterm rabbits. The interrater absolute agreement for the evaluation of lung ultrasound images in rabbits was very high [ICC: 0.989 (95%CI: 0.975–0.995); P < 0.0001], and there was no difference between the two observers. Lung mechanics parameters were similarly altered in both rabbits and human neonates. There were moderately significant correlations between airway resistances and lung ultrasound scores in rabbits ( ρ = 0.519; P = 0.008) and in neonates ( ρ = 0.409; P = 0.042). In conclusion, the preterm rabbit model fairly reproduces the lung ultrasound and mechanical characteristics of preterm neonates with evolving BPD.
NEW & NOTEWORTHY We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.</description><subject>Abnormalities</subject><subject>Aeration</subject><subject>Animal models</subject><subject>Dysplasia</subject><subject>Evaluation</subject><subject>Evolution</subject><subject>Exposure</subject><subject>Gestation</subject><subject>Gestational age</subject><subject>Histology</subject><subject>Hyperoxia</subject><subject>Life Sciences</subject><subject>Lungs</subject><subject>Mechanical properties</subject><subject>Mechanics</subject><subject>Mechanics (physics)</subject><subject>Neonates</subject><subject>Observers</subject><subject>Rabbits</subject><subject>Respiratory tract</subject><subject>Ultrasonic imaging</subject><subject>Ultrasound</subject><issn>8750-7587</issn><issn>1522-1601</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNpNkU1r3DAQhkVoIds0vyGCnHrwVp-W95ikHwkstIf2LMb2KNbilRxJ3rL0z9fbLSWnYd55eBl4CLnhbM25Fh93ME3jNByzj-OaMcnYWjDBL8hquYqK14y_IavGaFYZ3ZhL8i7nHWNcKc1X5Pd2Ds90HkuCHOfQU4dQ5oSZwrIkHKH4GPLgp0x_-TIsUZ58ghLTke6xGyD4LtPoKB7iePBL2f33T9QHOiUsmPY0Qdv6cu4b5j0EGjAGKJjfk7cOxozX_-YV-fnl84-Hx2r77evTw9226hTjpWq7vu6xqcEJJbWTzmBX143pjOygV41xrWEbaXBTO9HDhveyd6ha4KBrBCmvyIdz7wCjnZLfQzraCN4-3m3tKWNKS6E1P_CFvT2zU4ovM-Zid3FOYXnPCm2YUEI2aqHMmepSzDmh-1_LmT1Zsa-t2L9W7MmK_AOSnYdJ</recordid><startdate>20210901</startdate><enddate>20210901</enddate><creator>Loi, Barbara</creator><creator>Casiraghi, Costanza</creator><creator>Catozzi, Chiara</creator><creator>Storti, Matteo</creator><creator>Lucattelli, Monica</creator><creator>Bartalesi, Barbara</creator><creator>Yousef, Nadya</creator><creator>Salomone, Fabrizio</creator><creator>De Luca, Daniele</creator><general>American Physiological Society</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7QP</scope><scope>7QR</scope><scope>7TK</scope><scope>7TS</scope><scope>7U7</scope><scope>8FD</scope><scope>C1K</scope><scope>FR3</scope><scope>P64</scope><scope>1XC</scope><orcidid>https://orcid.org/0000-0001-6756-7237</orcidid><orcidid>https://orcid.org/0000-0002-3846-4834</orcidid></search><sort><creationdate>20210901</creationdate><title>Lung ultrasound features and relationships with respiratory mechanics of evolving BPD in preterm rabbits and human neonates</title><author>Loi, Barbara ; Casiraghi, Costanza ; Catozzi, Chiara ; Storti, Matteo ; Lucattelli, Monica ; Bartalesi, Barbara ; Yousef, Nadya ; Salomone, Fabrizio ; De Luca, Daniele</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c401t-bcd6de86af2435f3f7ec6687c73cad487fb70937e96f2da91d3dfe4ba1a56ea33</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Abnormalities</topic><topic>Aeration</topic><topic>Animal models</topic><topic>Dysplasia</topic><topic>Evaluation</topic><topic>Evolution</topic><topic>Exposure</topic><topic>Gestation</topic><topic>Gestational age</topic><topic>Histology</topic><topic>Hyperoxia</topic><topic>Life Sciences</topic><topic>Lungs</topic><topic>Mechanical properties</topic><topic>Mechanics</topic><topic>Mechanics (physics)</topic><topic>Neonates</topic><topic>Observers</topic><topic>Rabbits</topic><topic>Respiratory tract</topic><topic>Ultrasonic imaging</topic><topic>Ultrasound</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Loi, Barbara</creatorcontrib><creatorcontrib>Casiraghi, Costanza</creatorcontrib><creatorcontrib>Catozzi, Chiara</creatorcontrib><creatorcontrib>Storti, Matteo</creatorcontrib><creatorcontrib>Lucattelli, Monica</creatorcontrib><creatorcontrib>Bartalesi, Barbara</creatorcontrib><creatorcontrib>Yousef, Nadya</creatorcontrib><creatorcontrib>Salomone, Fabrizio</creatorcontrib><creatorcontrib>De Luca, Daniele</creatorcontrib><collection>CrossRef</collection><collection>Calcium & Calcified Tissue Abstracts</collection><collection>Chemoreception Abstracts</collection><collection>Neurosciences Abstracts</collection><collection>Physical Education Index</collection><collection>Toxicology Abstracts</collection><collection>Technology Research Database</collection><collection>Environmental Sciences and Pollution Management</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>Hyper Article en Ligne (HAL)</collection><jtitle>Journal of applied physiology (1985)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Loi, Barbara</au><au>Casiraghi, Costanza</au><au>Catozzi, Chiara</au><au>Storti, Matteo</au><au>Lucattelli, Monica</au><au>Bartalesi, Barbara</au><au>Yousef, Nadya</au><au>Salomone, Fabrizio</au><au>De Luca, Daniele</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Lung ultrasound features and relationships with respiratory mechanics of evolving BPD in preterm rabbits and human neonates</atitle><jtitle>Journal of applied physiology (1985)</jtitle><date>2021-09-01</date><risdate>2021</risdate><volume>131</volume><issue>3</issue><spage>895</spage><epage>904</epage><pages>895-904</pages><issn>8750-7587</issn><eissn>1522-1601</eissn><abstract>We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.
Evolving bronchopulmonary dysplasia (BPD) is characterized by impaired alveolarization leading to lung aeration inhomogeneities. Hyperoxia-exposed preterm rabbits have been proposed to mimic evolving BPD; therefore, we aimed to verify if this model has the same lung ultrasound and mechanical features of evolving BPD in human neonates. Semiquantitative lung ultrasound and lung mechanics measurement was performed in 25 preterm rabbits (28 days of gestation) and 25 neonates (mean gestational age ≈ 26 wk) with evolving BPD. A modified rabbit lung ultrasound score (rLUS) and a validated neonatal lung ultrasound score (LUS) were used. Lung ultrasound images were recorded and evaluated by two independent observers blinded to each other’s evaluation. Lung ultrasound findings were equally heterogeneous both in rabbits as in human neonates and encompassed all the classical lung ultrasound semiology. Lung ultrasound and histology examination were also performed in 13 term rabbits kept under normoxia as further control and showed the absence of ultrasound and histology abnormalities compared with hyperoxia-exposed preterm rabbits. The interrater absolute agreement for the evaluation of lung ultrasound images in rabbits was very high [ICC: 0.989 (95%CI: 0.975–0.995); P < 0.0001], and there was no difference between the two observers. Lung mechanics parameters were similarly altered in both rabbits and human neonates. There were moderately significant correlations between airway resistances and lung ultrasound scores in rabbits ( ρ = 0.519; P = 0.008) and in neonates ( ρ = 0.409; P = 0.042). In conclusion, the preterm rabbit model fairly reproduces the lung ultrasound and mechanical characteristics of preterm neonates with evolving BPD.
NEW & NOTEWORTHY We have reported that hyperoxia-exposed preterm rabbits and human preterm neonates with evolving BPD have the same lung ultrasound appearance, and that lung ultrasound can be fruitfully applied on this model with a brief training. The animal model and human neonates also presented the same relationship between semiquantitative ultrasound-assessed lung aeration and airway resistances. In conclusion, this animal model fairly reproduce evolving BPD as it is seen in clinical practice.</abstract><cop>Bethesda</cop><pub>American Physiological Society</pub><doi>10.1152/japplphysiol.00300.2021</doi><tpages>10</tpages><orcidid>https://orcid.org/0000-0001-6756-7237</orcidid><orcidid>https://orcid.org/0000-0002-3846-4834</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 8750-7587 |
ispartof | Journal of applied physiology (1985), 2021-09, Vol.131 (3), p.895-904 |
issn | 8750-7587 1522-1601 |
language | eng |
recordid | cdi_hal_primary_oai_HAL_hal_04532551v1 |
source | American Physiological Society; EZB-FREE-00999 freely available EZB journals; Alma/SFX Local Collection |
subjects | Abnormalities Aeration Animal models Dysplasia Evaluation Evolution Exposure Gestation Gestational age Histology Hyperoxia Life Sciences Lungs Mechanical properties Mechanics Mechanics (physics) Neonates Observers Rabbits Respiratory tract Ultrasonic imaging Ultrasound |
title | Lung ultrasound features and relationships with respiratory mechanics of evolving BPD in preterm rabbits and human neonates |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-12T07%3A40%3A13IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_hal_p&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Lung%20ultrasound%20features%20and%20relationships%20with%20respiratory%20mechanics%20of%20evolving%20BPD%20in%20preterm%20rabbits%20and%20human%20neonates&rft.jtitle=Journal%20of%20applied%20physiology%20(1985)&rft.au=Loi,%20Barbara&rft.date=2021-09-01&rft.volume=131&rft.issue=3&rft.spage=895&rft.epage=904&rft.pages=895-904&rft.issn=8750-7587&rft.eissn=1522-1601&rft_id=info:doi/10.1152/japplphysiol.00300.2021&rft_dat=%3Cproquest_hal_p%3E2570242384%3C/proquest_hal_p%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2570242384&rft_id=info:pmid/&rfr_iscdi=true |