Accuracy of Zero-Heat-Flux Cutaneous Temperature in Intensive Care Adults
OBJECTIVES:To compare accuracy of a continuous noninvasive cutaneous temperature using zero-heat-flux method to esophageal temperature and arterial temperature. DESIGN:Prospective study. SETTING:ICU and NeuroICU, University Hospital. PATIENTS:Fifty-two ICU patients over a 4-month period who required...
Gespeichert in:
Veröffentlicht in: | Critical care medicine 2017-07, Vol.45 (7), p.e715-e717 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVES:To compare accuracy of a continuous noninvasive cutaneous temperature using zero-heat-flux method to esophageal temperature and arterial temperature.
DESIGN:Prospective study.
SETTING:ICU and NeuroICU, University Hospital.
PATIENTS:Fifty-two ICU patients over a 4-month period who required continuous temperature monitoring were included in the study, after informed consent.
INTERVENTIONS:All patients had esophageal temperature probe and a noninvasive cutaneous device to monitor their core temperature continuously. In seven patients who required cardiac output monitoring, continuous iliac arterial temperature was collected. Simultaneous core temperatures were recorded from 1 to 5 days. Comparison to the esophageal temperature, considered as the reference in this study, used the Bland and Altman method with adjustment for multiple measurements per patient.
MEASUREMENTS AND MAIN RESULTS:The esophageal temperature ranged from 33°C to 39.7°C, 61,298 pairs of temperature using zero-heat-flux and esophageal temperature were collected and 1,850 triple of temperature using zero-heat-flux, esophageal temperature, and arterial temperature. Bias and limits of agreement for temperature using zero-heat-flux were 0.19°C ± 0.53°C compared with esophageal temperature with an absolute difference of temperature pairs equal to or lower than 0.5°C of 92.6% (95% CI, 91.9–93.4%) of cases and equal to or lower than 1°C for 99.9% (95% CI, 99.7–100.0%) of cases. Compared with arterial temperature, bias and limits of agreement were –0.00°C ± 0.36°C with an absolute difference of temperature pairs equal to or lower than 0.5°C of 99.8% (95% CI, 95.3–100%) of cases. All absolute difference of temperature pairs between temperature using zero-heat-flux and arterial temperature and between arterial temperature and esophageal temperature were equal to or lower than 1°C. No local or systemic serious complication was observed.
CONCLUSIONS:These results suggest a comparable reliability of the cutaneous sensor using the zero-heat-flux method compared with esophageal or iliac arterial temperatures measurements. |
---|---|