Coil and Magnet Design for Nuclear Magnetic Resonance in Inhomogeneous Field

High-resolution nuclear magnetic resonance (NMR) spectrometry is possible in nonhomogeneous magnetic fields-such as those in portable equipment-if the static and the radio-frequency (RF) magnetic fields are perpendicular and correlated in the measurement volume. From the easy-axis rotation theorem a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on magnetics 2006-12, Vol.42 (12), p.3861-3867
Hauptverfasser: Le Bec, G., Yonnet, J.-p., Raoof, K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:High-resolution nuclear magnetic resonance (NMR) spectrometry is possible in nonhomogeneous magnetic fields-such as those in portable equipment-if the static and the radio-frequency (RF) magnetic fields are perpendicular and correlated in the measurement volume. From the easy-axis rotation theorem and the Amperian currents model, it is possible to exactly match two magnetic fields in two-dimensional systems. We derive a basic probe element that fulfills these conditions. Then we present a portable NMR probe design. The static and RF magnetic fields of the probe are matched on a large volume
ISSN:0018-9464
1941-0069
DOI:10.1109/TMAG.2006.884385