Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification
Purpose To assess errors associated with EPI-accelerated intracardiac 4D flow MRI (4DEPI) with EPI factor 5, compared with non-EPI gradient echo (4DGRE). Methods Three 3T MRI experiments were performed comparing 4DEPI to 4DGRE: steady flow through straight tubes, pulsatile flow in a left-ventricle p...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , , , |
---|---|
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | |
---|---|
container_issue | |
container_start_page | |
container_title | |
container_volume | |
creator | Westenberg, Jos J. M Assen, Hans C den Boogaard, Pieter J Goeman, Jelle J Saaid, Hicham Voorneveld, Jason Bosch, Johan Kenjeres, Sasa Claessens, Tom Garg, Pankaj Kouwenhoven, Marc Lamb, Hildo J |
description | Purpose To assess errors associated with EPI-accelerated intracardiac 4D flow MRI (4DEPI) with EPI factor 5, compared with non-EPI gradient echo (4DGRE). Methods Three 3T MRI experiments were performed comparing 4DEPI to 4DGRE: steady flow through straight tubes, pulsatile flow in a left-ventricle phantom, and intracardiac flow in 10 healthy volunteers. For each experiment, 4DEPI was repeated with readout and blip phase-encoding gradient in different orientations, parallel or perpendicular to the flow direction. In vitro flow rates were compared with timed volumetric collection. In the left-ventricle phantom and in vivo, voxel-based speed and spatio-temporal median speed were compared between sequences, as well as mitral and aortic transvalvular net forward volume. Results In steady-flow phantoms, the flow rate error was largest (12%) for high velocity (>2 m/s) with 4DEPI readout gradient parallel to the flow. Voxel-based speed and median speed in the left-ventricle phantom were |
format | Article |
fullrecord | <record><control><sourceid>ghent</sourceid><recordid>TN_cdi_ghent_librecat_oai_archive_ugent_be_8741101</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>oai_archive_ugent_be_8741101</sourcerecordid><originalsourceid>FETCH-ghent_librecat_oai_archive_ugent_be_87411013</originalsourceid><addsrcrecordid>eNqdi0sOgjAURTvQRPzs4W2ApEUIMlaMDpgY582jFHgGW22Lbl9MXIHJTc7gnDtjkciSJE6yIl2wpfc3znlR5GnEqlL1Fh4DGnRAd-zIdDGZZlS6Ae2cdR7ITAsOFbqGUEF6gHawb6guZ3iOaAK1pDCQNWs2b3HwevPjiiXH8ro_xV2vTZAD1U5PpbRIEp3q6aXl2H1VreUuT4XgYvvX6QNR7UmZ</addsrcrecordid><sourcetype>Institutional Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification</title><source>Wiley Online Library - AutoHoldings Journals</source><source>Ghent University Academic Bibliography</source><creator>Westenberg, Jos J. M ; Assen, Hans C ; den Boogaard, Pieter J ; Goeman, Jelle J ; Saaid, Hicham ; Voorneveld, Jason ; Bosch, Johan ; Kenjeres, Sasa ; Claessens, Tom ; Garg, Pankaj ; Kouwenhoven, Marc ; Lamb, Hildo J</creator><creatorcontrib>Westenberg, Jos J. M ; Assen, Hans C ; den Boogaard, Pieter J ; Goeman, Jelle J ; Saaid, Hicham ; Voorneveld, Jason ; Bosch, Johan ; Kenjeres, Sasa ; Claessens, Tom ; Garg, Pankaj ; Kouwenhoven, Marc ; Lamb, Hildo J</creatorcontrib><description>Purpose To assess errors associated with EPI-accelerated intracardiac 4D flow MRI (4DEPI) with EPI factor 5, compared with non-EPI gradient echo (4DGRE). Methods Three 3T MRI experiments were performed comparing 4DEPI to 4DGRE: steady flow through straight tubes, pulsatile flow in a left-ventricle phantom, and intracardiac flow in 10 healthy volunteers. For each experiment, 4DEPI was repeated with readout and blip phase-encoding gradient in different orientations, parallel or perpendicular to the flow direction. In vitro flow rates were compared with timed volumetric collection. In the left-ventricle phantom and in vivo, voxel-based speed and spatio-temporal median speed were compared between sequences, as well as mitral and aortic transvalvular net forward volume. Results In steady-flow phantoms, the flow rate error was largest (12%) for high velocity (>2 m/s) with 4DEPI readout gradient parallel to the flow. Voxel-based speed and median speed in the left-ventricle phantom were <= 5.5% different between sequences. In vivo, mean net forward volume inconsistency was largest (6.4 +/- 8.5%) for 4DEPI with nonblip phase-encoding gradient parallel to the main flow. The difference in median speed for 4DEPI versus 4DGRE was largest (9%) when the 4DEPI readout gradient was parallel to the flow. Conclusions Velocity and flow rate are inaccurate for 4DEPI with EPI factor 5 when flow is parallel to the readout or blip phase-encoding gradient. However, mean differences in flow rate, voxel-based speed, and spatio-temporal median speed were acceptable (<= 10%) when comparing 4DEPI to 4DGRE for intracardiac flow in healthy volunteers.</description><identifier>ISSN: 1522-2594</identifier><identifier>ISSN: 0740-3194</identifier><language>eng</language><subject>4D flow MRI ; ACCURACY ; echo planar imaging ; ECHOCARDIOGRAPHY ; HEART ; intracardiac blood flow ; Nuclear Medicine and imaging ; PATTERNS ; phase-contrast MRI ; Radiology ; REGURGITATION ; SENSITIVITY ; Technology and Engineering ; VALIDATION ; VALVE ; VESSELS ; VISUALIZATION</subject><creationdate>2022</creationdate><rights>Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International Public License (CC BY-NC-ND 4.0) info:eu-repo/semantics/openAccess</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>315,316,782,786,4026,27867</link.rule.ids></links><search><creatorcontrib>Westenberg, Jos J. M</creatorcontrib><creatorcontrib>Assen, Hans C</creatorcontrib><creatorcontrib>den Boogaard, Pieter J</creatorcontrib><creatorcontrib>Goeman, Jelle J</creatorcontrib><creatorcontrib>Saaid, Hicham</creatorcontrib><creatorcontrib>Voorneveld, Jason</creatorcontrib><creatorcontrib>Bosch, Johan</creatorcontrib><creatorcontrib>Kenjeres, Sasa</creatorcontrib><creatorcontrib>Claessens, Tom</creatorcontrib><creatorcontrib>Garg, Pankaj</creatorcontrib><creatorcontrib>Kouwenhoven, Marc</creatorcontrib><creatorcontrib>Lamb, Hildo J</creatorcontrib><title>Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification</title><description>Purpose To assess errors associated with EPI-accelerated intracardiac 4D flow MRI (4DEPI) with EPI factor 5, compared with non-EPI gradient echo (4DGRE). Methods Three 3T MRI experiments were performed comparing 4DEPI to 4DGRE: steady flow through straight tubes, pulsatile flow in a left-ventricle phantom, and intracardiac flow in 10 healthy volunteers. For each experiment, 4DEPI was repeated with readout and blip phase-encoding gradient in different orientations, parallel or perpendicular to the flow direction. In vitro flow rates were compared with timed volumetric collection. In the left-ventricle phantom and in vivo, voxel-based speed and spatio-temporal median speed were compared between sequences, as well as mitral and aortic transvalvular net forward volume. Results In steady-flow phantoms, the flow rate error was largest (12%) for high velocity (>2 m/s) with 4DEPI readout gradient parallel to the flow. Voxel-based speed and median speed in the left-ventricle phantom were <= 5.5% different between sequences. In vivo, mean net forward volume inconsistency was largest (6.4 +/- 8.5%) for 4DEPI with nonblip phase-encoding gradient parallel to the main flow. The difference in median speed for 4DEPI versus 4DGRE was largest (9%) when the 4DEPI readout gradient was parallel to the flow. Conclusions Velocity and flow rate are inaccurate for 4DEPI with EPI factor 5 when flow is parallel to the readout or blip phase-encoding gradient. However, mean differences in flow rate, voxel-based speed, and spatio-temporal median speed were acceptable (<= 10%) when comparing 4DEPI to 4DGRE for intracardiac flow in healthy volunteers.</description><subject>4D flow MRI</subject><subject>ACCURACY</subject><subject>echo planar imaging</subject><subject>ECHOCARDIOGRAPHY</subject><subject>HEART</subject><subject>intracardiac blood flow</subject><subject>Nuclear Medicine and imaging</subject><subject>PATTERNS</subject><subject>phase-contrast MRI</subject><subject>Radiology</subject><subject>REGURGITATION</subject><subject>SENSITIVITY</subject><subject>Technology and Engineering</subject><subject>VALIDATION</subject><subject>VALVE</subject><subject>VESSELS</subject><subject>VISUALIZATION</subject><issn>1522-2594</issn><issn>0740-3194</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><sourceid>ADGLB</sourceid><recordid>eNqdi0sOgjAURTvQRPzs4W2ApEUIMlaMDpgY582jFHgGW22Lbl9MXIHJTc7gnDtjkciSJE6yIl2wpfc3znlR5GnEqlL1Fh4DGnRAd-zIdDGZZlS6Ae2cdR7ITAsOFbqGUEF6gHawb6guZ3iOaAK1pDCQNWs2b3HwevPjiiXH8ro_xV2vTZAD1U5PpbRIEp3q6aXl2H1VreUuT4XgYvvX6QNR7UmZ</recordid><startdate>2022</startdate><enddate>2022</enddate><creator>Westenberg, Jos J. M</creator><creator>Assen, Hans C</creator><creator>den Boogaard, Pieter J</creator><creator>Goeman, Jelle J</creator><creator>Saaid, Hicham</creator><creator>Voorneveld, Jason</creator><creator>Bosch, Johan</creator><creator>Kenjeres, Sasa</creator><creator>Claessens, Tom</creator><creator>Garg, Pankaj</creator><creator>Kouwenhoven, Marc</creator><creator>Lamb, Hildo J</creator><scope>ADGLB</scope></search><sort><creationdate>2022</creationdate><title>Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification</title><author>Westenberg, Jos J. M ; Assen, Hans C ; den Boogaard, Pieter J ; Goeman, Jelle J ; Saaid, Hicham ; Voorneveld, Jason ; Bosch, Johan ; Kenjeres, Sasa ; Claessens, Tom ; Garg, Pankaj ; Kouwenhoven, Marc ; Lamb, Hildo J</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-ghent_librecat_oai_archive_ugent_be_87411013</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><topic>4D flow MRI</topic><topic>ACCURACY</topic><topic>echo planar imaging</topic><topic>ECHOCARDIOGRAPHY</topic><topic>HEART</topic><topic>intracardiac blood flow</topic><topic>Nuclear Medicine and imaging</topic><topic>PATTERNS</topic><topic>phase-contrast MRI</topic><topic>Radiology</topic><topic>REGURGITATION</topic><topic>SENSITIVITY</topic><topic>Technology and Engineering</topic><topic>VALIDATION</topic><topic>VALVE</topic><topic>VESSELS</topic><topic>VISUALIZATION</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Westenberg, Jos J. M</creatorcontrib><creatorcontrib>Assen, Hans C</creatorcontrib><creatorcontrib>den Boogaard, Pieter J</creatorcontrib><creatorcontrib>Goeman, Jelle J</creatorcontrib><creatorcontrib>Saaid, Hicham</creatorcontrib><creatorcontrib>Voorneveld, Jason</creatorcontrib><creatorcontrib>Bosch, Johan</creatorcontrib><creatorcontrib>Kenjeres, Sasa</creatorcontrib><creatorcontrib>Claessens, Tom</creatorcontrib><creatorcontrib>Garg, Pankaj</creatorcontrib><creatorcontrib>Kouwenhoven, Marc</creatorcontrib><creatorcontrib>Lamb, Hildo J</creatorcontrib><collection>Ghent University Academic Bibliography</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Westenberg, Jos J. M</au><au>Assen, Hans C</au><au>den Boogaard, Pieter J</au><au>Goeman, Jelle J</au><au>Saaid, Hicham</au><au>Voorneveld, Jason</au><au>Bosch, Johan</au><au>Kenjeres, Sasa</au><au>Claessens, Tom</au><au>Garg, Pankaj</au><au>Kouwenhoven, Marc</au><au>Lamb, Hildo J</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification</atitle><date>2022</date><risdate>2022</risdate><issn>1522-2594</issn><issn>0740-3194</issn><abstract>Purpose To assess errors associated with EPI-accelerated intracardiac 4D flow MRI (4DEPI) with EPI factor 5, compared with non-EPI gradient echo (4DGRE). Methods Three 3T MRI experiments were performed comparing 4DEPI to 4DGRE: steady flow through straight tubes, pulsatile flow in a left-ventricle phantom, and intracardiac flow in 10 healthy volunteers. For each experiment, 4DEPI was repeated with readout and blip phase-encoding gradient in different orientations, parallel or perpendicular to the flow direction. In vitro flow rates were compared with timed volumetric collection. In the left-ventricle phantom and in vivo, voxel-based speed and spatio-temporal median speed were compared between sequences, as well as mitral and aortic transvalvular net forward volume. Results In steady-flow phantoms, the flow rate error was largest (12%) for high velocity (>2 m/s) with 4DEPI readout gradient parallel to the flow. Voxel-based speed and median speed in the left-ventricle phantom were <= 5.5% different between sequences. In vivo, mean net forward volume inconsistency was largest (6.4 +/- 8.5%) for 4DEPI with nonblip phase-encoding gradient parallel to the main flow. The difference in median speed for 4DEPI versus 4DGRE was largest (9%) when the 4DEPI readout gradient was parallel to the flow. Conclusions Velocity and flow rate are inaccurate for 4DEPI with EPI factor 5 when flow is parallel to the readout or blip phase-encoding gradient. However, mean differences in flow rate, voxel-based speed, and spatio-temporal median speed were acceptable (<= 10%) when comparing 4DEPI to 4DGRE for intracardiac flow in healthy volunteers.</abstract><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1522-2594 |
ispartof | |
issn | 1522-2594 0740-3194 |
language | eng |
recordid | cdi_ghent_librecat_oai_archive_ugent_be_8741101 |
source | Wiley Online Library - AutoHoldings Journals; Ghent University Academic Bibliography |
subjects | 4D flow MRI ACCURACY echo planar imaging ECHOCARDIOGRAPHY HEART intracardiac blood flow Nuclear Medicine and imaging PATTERNS phase-contrast MRI Radiology REGURGITATION SENSITIVITY Technology and Engineering VALIDATION VALVE VESSELS VISUALIZATION |
title | Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-04T01%3A37%3A42IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-ghent&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Echo%20planar%20imaging-induced%20errors%20in%20intracardiac%204D%20flow%20MRI%20quantification&rft.au=Westenberg,%20Jos%20J.%20M&rft.date=2022&rft.issn=1522-2594&rft_id=info:doi/&rft_dat=%3Cghent%3Eoai_archive_ugent_be_8741101%3C/ghent%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |