Application of a mass spectrometer as a capnograph

The feasibility of using a mass spectrometer for monitoring the carbon dioxide and inhalational anesthetic concentrations in the breathing circuit of an apparatus for inhalational anesthesia are demonstrated. Mass-spectrometric data for the CO 2 and inhalational anesthetic concentrations are compare...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Technical physics 2010-12, Vol.55 (12), p.1814-1816
Hauptverfasser: Elokhin, V. A., Ershov, T. D., Levshankov, A. I., Nikolaev, V. I., Elizarov, A. Yu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The feasibility of using a mass spectrometer for monitoring the carbon dioxide and inhalational anesthetic concentrations in the breathing circuit of an apparatus for inhalational anesthesia are demonstrated. Mass-spectrometric data for the CO 2 and inhalational anesthetic concentrations are compared with related optical data. The advantages of the mass spectrometer as a capnograph over the optical spectrometer are indicated. The variation of the inhalational anesthetic content in expired air is shown to depend on the muscle relaxation efficiency.
ISSN:1063-7842
1090-6525
DOI:10.1134/S1063784210120194