Rohrer's constant, [K.sub.2], as a factor of determining inspiratory resistance of common adult endotracheal tubes
The aim of the study was to calculate the in vitro inspiratory resistance ([R.sub.ETT]) of adult endotracheal tubes (ETT), via the end-inspiratory occlusion method, and to apply this method in vivo in order to estimate [R.sub.ETT] value in real time. By plotting [R.sub.ETT] over inspiratory flow (V)...
Gespeichert in:
Veröffentlicht in: | Anaesthesia and intensive care 2011-05, Vol.39 (3), p.410 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of the study was to calculate the in vitro inspiratory resistance ([R.sub.ETT]) of adult endotracheal tubes (ETT), via the end-inspiratory occlusion method, and to apply this method in vivo in order to estimate [R.sub.ETT] value in real time. By plotting [R.sub.ETT] over inspiratory flow (V) and calculating Rohrer's coefficients of linear and nonlinear resistance, [K.sub.1] and [K.sub.2] respectively, we determined the resistive behaviour of each ETT. Peak and plateau pressures were recorded at both proximal and distal sites of the ETT after applying a three-second occlusion under constant flow. Distal pressure was obtained via an intraluminal catheter. [R.sub.ETT] was calculated as ([P.sub.peak]-[P.sub.plateau])/([??]), at both sites. [R.sub.ETT] value resulted from the difference [R.sub.proximal]-[R.sub.distal]. Graph [R.sub.ETT] over ([??]) was plotted and Rohrer's constants were calculated by the method of least squares. For ETTs with inner diameter 9.0, 8.5, 8.0, 7.5, 7.0 and 6.5 mm, K2 was 2.42, 3.05, 4.65, 6.01, 9.17 and 12.80 cm[H.sub.2]O/l/s, respectively. The intraluminal catheter increased [R.sub.ETT] No.7.0 by an average of 49%. Finally, ten patients with partially obstructed ETTs were tested and K2 in vivo constants found to be higher than their corresponding in vitro values (P value 0.00012). Therefore, knowing the performing size of an ETT may help the clinicians identify ETT obstruction and deal with weaning problems. Key Words: endotracheal tube, inspiratory resistance, Rohrer's constant |
---|---|
ISSN: | 0310-057X |