Validation of an Automated Detection Platform for Use with the Roche Linear Array Human Papillomavirus Genotyping Test

An automated platform (BeeBlot) was evaluated in parallel with the recommended protocol for the hybridization and detection steps of the Roche Linear Array human papillomavirus (HPV) genotyping test using DNA from 143 cervical specimens. Genotyping profiles showed 100% concordance between the method...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Clinical Microbiology 2008-11, Vol.46 (11), p.3813-3816
Hauptverfasser: Stevens, Matthew P, Garland, Suzanne M, Tabrizi, Sepehr N
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An automated platform (BeeBlot) was evaluated in parallel with the recommended protocol for the hybridization and detection steps of the Roche Linear Array human papillomavirus (HPV) genotyping test using DNA from 143 cervical specimens. Genotyping profiles showed 100% concordance between the methods, suggesting that automation could complement the Roche Linear Array for enhanced speed and reproducibility.
ISSN:0095-1137
1098-660X
DOI:10.1128/JCM.01169-08