Towards a Miniaturized 3D Receiver WPT System for Capsule Endoscopy
The optimization, manufacturing, and performance characterization of a miniaturized 3D receiver (RX)-based wireless power transfer (WPT) system fed by a multi-transmitter (multi-TX) array is presented in this study for applications in capsule endoscopy (CE). The 200 mm outer diameter, 35 μm thick pr...
Gespeichert in:
Veröffentlicht in: | Micromachines (Basel) 2019-08, Vol.10 (8), p.545 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The optimization, manufacturing, and performance characterization of a miniaturized 3D receiver (RX)-based wireless power transfer (WPT) system fed by a multi-transmitter (multi-TX) array is presented in this study for applications in capsule endoscopy (CE). The 200 mm outer diameter, 35 μm thick printed spiral TX coils of 2.8 g weight, is manufactured on a flexible substrate to enable bendability and portability of the transmitters by the patients. The 8.9 mm diameter-4.8 mm long, miniaturized 3D RX-includes a 4 mm diameter ferrite road to increase power transfer efficiency (PTE) and is dimensionally compatible for insertion into current endoscopic capsules. The multi-TX is activated using a custom-made high-efficiency dual class-E power amplifier operated in subnominal condition. A resulting link and system PTE of 1% and 0.7%, respectively, inside a phantom tissue is demonstrated for the proposed 3D WPT system. The specific absorption rate (SAR) is simulated using the HFSS
software (15.0) at 0.66 W/kg at 1 MHz operation frequency, which is below the IEEE guidelines for tissue safety. The maximum variation in temperature was also measured as 1.9 °C for the typical duration of the capsule's travel in the gastrointestinal tract to demonstrate the patients' tissues safety. |
---|---|
ISSN: | 2072-666X 2072-666X |
DOI: | 10.3390/mi10080545 |