Inaccuracy of Thyroid to Background Uptake Ratio in Evaluating Technetium-99m-pertechnetate Thyroid Uptake and Establishing an Improved Algorithm
The aim of this study was to explore the accuracy of thyroid to background uptake ratio (UR) in the evaluation of Tc-pertechnetate thyroid uptake (TcTU) and establishment of an improved algorithm. This study was conducted on the thyroid images of 322 patients with thyroid diseases and 67 controls. F...
Gespeichert in:
Veröffentlicht in: | Asia Oceania journal of nuclear medicine & biology 2019-01, Vol.7 (2), p.160-171 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study was to explore the accuracy of thyroid to background uptake ratio (UR) in the evaluation of
Tc-pertechnetate thyroid uptake (TcTU) and establishment of an improved algorithm.
This study was conducted on the thyroid images of 322 patients with thyroid diseases and 67 controls. For the purpose of the study, URs of the images were calculated, and then corrected by standardized thyroid area size to establish a corrected uptake ratio (CUR). Subsequently, the accuracy between UR and CUR was compared.
The results of linear regression using weighted least squares (using TcTU as a dependent variable and CUR, UR, or thyroid area size as independent variables) showed that CUR (t=105.5, P=0.000), UR (t=31.9, P=0.000), and thyroid area size (t=15.9, P=0.000) are influential factors of TcTU. Furthermore, the standardized coefficient of CUR (β=0.983) was obviously higher than those of UR (β=0.851) and thyroid area size (β=0.629). The linear goodness-of-fit between CUR and TcTU (R=0.983) was better than that between UR and TcTU (R=0.851). In addition, the total concordance rate between CUR and TcTU (96.7%) was significantly higher than that between UR and TcTU (83.0%; χ
=42.9, P=0.000). Discordance rates of CUR in large thyroid area (1.4% vs. 13.4%, χ
=17.0, P=0.000) and small thyroid area (3.3% vs. 42.2 %, χ
=44.3, P=0.000), were significantly lower than that of UR. In the abnormal thyroid areas, the discordance rates of UR obviously increased as compared to those of CUR. The UR overestimated the thyroid uptake in small thyroid areas and underestimated it in large thyroid areas.
Based on the findings, CUR is more accurate than UR in measuring
TcO
- thyroid uptake; accordingly, it is more significant in the diagnosis of thyroid disease. |
---|---|
ISSN: | 2322-5718 2322-5726 |
DOI: | 10.22038/AOJNMB.2019.12734 |