Field correction factors for a PTW‐31016 Pinpoint ionization chamber for both flattened and unflattened beams. Study of the main sources of uncertainties

Purpose The primary aim of this study was to determine correction factors, kQclin,Qmsrfclin,fmsr for a PTW‐31016 ionization chamber on field sizes from 0.5 cm × 0.5 cm to 2 cm × 2 cm for both flattened (FF) and flattened filter‐free (FFF) beams produced in a TrueBeam clinical accelerator. The second...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2017-05, Vol.44 (5), p.1930-1938
Hauptverfasser: Puxeu‐Vaqué, Josep, Duch, Maria A., Nailon, William H., Cruz Lizuain, M., Ginjaume, Mercè
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1938
container_issue 5
container_start_page 1930
container_title Medical physics (Lancaster)
container_volume 44
creator Puxeu‐Vaqué, Josep
Duch, Maria A.
Nailon, William H.
Cruz Lizuain, M.
Ginjaume, Mercè
description Purpose The primary aim of this study was to determine correction factors, kQclin,Qmsrfclin,fmsr for a PTW‐31016 ionization chamber on field sizes from 0.5 cm × 0.5 cm to 2 cm × 2 cm for both flattened (FF) and flattened filter‐free (FFF) beams produced in a TrueBeam clinical accelerator. The secondary objective was the determination of field output factors, ΩQclin,Qmsrfclin,fmsr over this range of field sizes using both Monte Carlo (MC) simulation and measurements. Methods kQclin,Qmsrfclin,fmsr for the PTW‐31016 chamber were calculated by MC simulation for field sizes of 0.5 cm × 0.5 cm, 1 cm × 1 cm, and 2 cm × 2 cm. MC simulations were performed with the PENELOPE code system for the 10 MV FFF Particle Space File from a TrueBeam linear accelerator (LINAC) provided by the manufacturer (Varian Medical Systems, Inc. Palo Alto, CA, USA). Simulations were repeated taking into account chamber manufacturing tolerances and accelerator jaw positioning in order to assess the uncertainty of the calculated correction factors. Output ratios were measured on square fields ranging from 0.5 cm × 0.5 cm to 10 cm × 10 cm for 6 MV and 10 MV FF and FFF beams produced by a TrueBeam using a PTW‐31016 ionization chamber; a Sun Nuclear Edge detector (SunNuclear Corp., Melbourne, FL, USA) and TLD–700R (Harshaw, Thermo Scientific, Waltham, MA, USA). The validity of the proposed correction factors was verified using the calculated correction factors for the determination of ΩQclin,Qmsrfclin,fmsr using a PTW‐31016 at the four TrueBeam energies and comparing the results with both TLD‐700R measurements and MC simulations. Finally, the proposed correction factors were used to assess the correction factors of the SunNuclear Edge detector. Results The present work provides a set of MC calculated correction factors for a PTW‐31016 chamber used on a TrueBeam FF and FFF mode. For the 0.5 cm × 0.5 cm square field size, kQclin,Qmsrfclin,fmsr is equal to 1.17 with a combined uncertainty of 2% (k = 1). A detailed analysis of the most influential parameters is presented in this work. PTW‐31016 corrected measurements were used for the determination of ΩQclin,Qmsrfclin,fmsr for 6 MV and 10 MV FF and FFF and the results were in agreement with values obtained using a TLD‐700R detector (differences < 3% for a 0.5 cm square field) for the four energies studied. Uncertainty in field collimation was found to be the main source of influence of ΩQclin,Qmsrfclin,fmsr and caused differences of up to 15% bet
doi_str_mv 10.1002/mp.12189
format Article
fullrecord <record><control><sourceid>proquest_csuc_</sourceid><recordid>TN_cdi_csuc_recercat_oai_recercat_cat_2072_288543</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1874784977</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3639-dcc4dbbbdf4986969c77f894b5d27b2317075791d78d564a06dc0b285983074d3</originalsourceid><addsrcrecordid>eNp1kc2KFDEUhYM4OO0o-ASSpZtq81eVZCmDo8KIDTPiMuSv6EhVUiYphnblI7j37XwSq7vH6dUsLpd7-M5Z3APAK4zWGCHydpzWmGAhn4AVYZw2jCD5FKwQkqwhDLXn4Hkp3xFCHW3RM3BOBOmwwHwF_lwFPzhoU87e1pAi7LWtKRfYpww13Nx--_vrN8UId3AT4pRCrHDBwk99oO1Wj8bnA21S3cJ-0LX66B3U0cE5nm7j9VjW8KbObgdTD-vWw1GHCEuas_Vlr83R-lwXsQZfXoCzXg_Fv7zfF-Dr1fvby4_N9ZcPny7fXTeWdlQ2zlrmjDGuZ1J0spOW815IZlpHuCEUc8RbLrHjwrUd06hzFhkiWiko4szRC4CPubbMVi1_8NnqqpIOp2M_BHGiiBAto4vnzdEz5fRj9qWqMRTrh0FHn-aisOCMCyY5P6E2p1Ky79WUw6jzTmGk9vWpcVKH-hb09X3qbEbvHsD_fS1AcwTuwuB3jwapz5tj4D9yHKTi</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1874784977</pqid></control><display><type>article</type><title>Field correction factors for a PTW‐31016 Pinpoint ionization chamber for both flattened and unflattened beams. Study of the main sources of uncertainties</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Recercat</source><source>Alma/SFX Local Collection</source><creator>Puxeu‐Vaqué, Josep ; Duch, Maria A. ; Nailon, William H. ; Cruz Lizuain, M. ; Ginjaume, Mercè</creator><creatorcontrib>Puxeu‐Vaqué, Josep ; Duch, Maria A. ; Nailon, William H. ; Cruz Lizuain, M. ; Ginjaume, Mercè</creatorcontrib><description>Purpose The primary aim of this study was to determine correction factors, kQclin,Qmsrfclin,fmsr for a PTW‐31016 ionization chamber on field sizes from 0.5 cm × 0.5 cm to 2 cm × 2 cm for both flattened (FF) and flattened filter‐free (FFF) beams produced in a TrueBeam clinical accelerator. The secondary objective was the determination of field output factors, ΩQclin,Qmsrfclin,fmsr over this range of field sizes using both Monte Carlo (MC) simulation and measurements. Methods kQclin,Qmsrfclin,fmsr for the PTW‐31016 chamber were calculated by MC simulation for field sizes of 0.5 cm × 0.5 cm, 1 cm × 1 cm, and 2 cm × 2 cm. MC simulations were performed with the PENELOPE code system for the 10 MV FFF Particle Space File from a TrueBeam linear accelerator (LINAC) provided by the manufacturer (Varian Medical Systems, Inc. Palo Alto, CA, USA). Simulations were repeated taking into account chamber manufacturing tolerances and accelerator jaw positioning in order to assess the uncertainty of the calculated correction factors. Output ratios were measured on square fields ranging from 0.5 cm × 0.5 cm to 10 cm × 10 cm for 6 MV and 10 MV FF and FFF beams produced by a TrueBeam using a PTW‐31016 ionization chamber; a Sun Nuclear Edge detector (SunNuclear Corp., Melbourne, FL, USA) and TLD–700R (Harshaw, Thermo Scientific, Waltham, MA, USA). The validity of the proposed correction factors was verified using the calculated correction factors for the determination of ΩQclin,Qmsrfclin,fmsr using a PTW‐31016 at the four TrueBeam energies and comparing the results with both TLD‐700R measurements and MC simulations. Finally, the proposed correction factors were used to assess the correction factors of the SunNuclear Edge detector. Results The present work provides a set of MC calculated correction factors for a PTW‐31016 chamber used on a TrueBeam FF and FFF mode. For the 0.5 cm × 0.5 cm square field size, kQclin,Qmsrfclin,fmsr is equal to 1.17 with a combined uncertainty of 2% (k = 1). A detailed analysis of the most influential parameters is presented in this work. PTW‐31016 corrected measurements were used for the determination of ΩQclin,Qmsrfclin,fmsr for 6 MV and 10 MV FF and FFF and the results were in agreement with values obtained using a TLD‐700R detector (differences &lt; 3% for a 0.5 cm square field) for the four energies studied. Uncertainty in field collimation was found to be the main source of influence of ΩQclin,Qmsrfclin,fmsr and caused differences of up to 15% between calculations and measurements for the 0.5 cm × 0.5 cm field. This was also confirmed by repeating the same measurements at two different institutions. Conclusions This study confirms the need to introduce correction factors when using a PTW‐31016 chamber and the hypothesis of their low energy dependence. MC simulation has been shown to be a useful methodology to determine detector correction factors for small fields and to analyze the main sources of uncertainty. However, due to the influence of the LINAC jaw setup for field sizes below or equal to 1 cm, MC methods are not recommended in this range for field output factor calculations.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1002/mp.12189</identifier><identifier>PMID: 28261817</identifier><language>eng</language><publisher>United States</publisher><subject>Ciències de la salut ; Correction factors ; Flattening filter-free ; Física ; Física mèdica ; Humans ; Medical physics ; Medicina ; Monte Carlo ; Monte Carlo Method ; Montecarlo, Mètode de ; Output factors ; Particle Accelerators ; Photons ; Radiometry ; Setup inaccuracies ; Small fields ; Uncertainty ; Àrees temàtiques de la UPC</subject><ispartof>Medical physics (Lancaster), 2017-05, Vol.44 (5), p.1930-1938</ispartof><rights>2017 American Association of Physicists in Medicine</rights><rights>2017 American Association of Physicists in Medicine.</rights><rights>Attribution-NonCommercial-NoDerivs 3.0 Spain info:eu-repo/semantics/openAccess &lt;a href="http://creativecommons.org/licenses/by-nc-nd/3.0/es/"&gt;http://creativecommons.org/licenses/by-nc-nd/3.0/es/&lt;/a&gt;</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3639-dcc4dbbbdf4986969c77f894b5d27b2317075791d78d564a06dc0b285983074d3</citedby><cites>FETCH-LOGICAL-c3639-dcc4dbbbdf4986969c77f894b5d27b2317075791d78d564a06dc0b285983074d3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1002%2Fmp.12189$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1002%2Fmp.12189$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>230,314,776,780,881,1411,26953,27903,27904,45552,45553</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28261817$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Puxeu‐Vaqué, Josep</creatorcontrib><creatorcontrib>Duch, Maria A.</creatorcontrib><creatorcontrib>Nailon, William H.</creatorcontrib><creatorcontrib>Cruz Lizuain, M.</creatorcontrib><creatorcontrib>Ginjaume, Mercè</creatorcontrib><title>Field correction factors for a PTW‐31016 Pinpoint ionization chamber for both flattened and unflattened beams. Study of the main sources of uncertainties</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose The primary aim of this study was to determine correction factors, kQclin,Qmsrfclin,fmsr for a PTW‐31016 ionization chamber on field sizes from 0.5 cm × 0.5 cm to 2 cm × 2 cm for both flattened (FF) and flattened filter‐free (FFF) beams produced in a TrueBeam clinical accelerator. The secondary objective was the determination of field output factors, ΩQclin,Qmsrfclin,fmsr over this range of field sizes using both Monte Carlo (MC) simulation and measurements. Methods kQclin,Qmsrfclin,fmsr for the PTW‐31016 chamber were calculated by MC simulation for field sizes of 0.5 cm × 0.5 cm, 1 cm × 1 cm, and 2 cm × 2 cm. MC simulations were performed with the PENELOPE code system for the 10 MV FFF Particle Space File from a TrueBeam linear accelerator (LINAC) provided by the manufacturer (Varian Medical Systems, Inc. Palo Alto, CA, USA). Simulations were repeated taking into account chamber manufacturing tolerances and accelerator jaw positioning in order to assess the uncertainty of the calculated correction factors. Output ratios were measured on square fields ranging from 0.5 cm × 0.5 cm to 10 cm × 10 cm for 6 MV and 10 MV FF and FFF beams produced by a TrueBeam using a PTW‐31016 ionization chamber; a Sun Nuclear Edge detector (SunNuclear Corp., Melbourne, FL, USA) and TLD–700R (Harshaw, Thermo Scientific, Waltham, MA, USA). The validity of the proposed correction factors was verified using the calculated correction factors for the determination of ΩQclin,Qmsrfclin,fmsr using a PTW‐31016 at the four TrueBeam energies and comparing the results with both TLD‐700R measurements and MC simulations. Finally, the proposed correction factors were used to assess the correction factors of the SunNuclear Edge detector. Results The present work provides a set of MC calculated correction factors for a PTW‐31016 chamber used on a TrueBeam FF and FFF mode. For the 0.5 cm × 0.5 cm square field size, kQclin,Qmsrfclin,fmsr is equal to 1.17 with a combined uncertainty of 2% (k = 1). A detailed analysis of the most influential parameters is presented in this work. PTW‐31016 corrected measurements were used for the determination of ΩQclin,Qmsrfclin,fmsr for 6 MV and 10 MV FF and FFF and the results were in agreement with values obtained using a TLD‐700R detector (differences &lt; 3% for a 0.5 cm square field) for the four energies studied. Uncertainty in field collimation was found to be the main source of influence of ΩQclin,Qmsrfclin,fmsr and caused differences of up to 15% between calculations and measurements for the 0.5 cm × 0.5 cm field. This was also confirmed by repeating the same measurements at two different institutions. Conclusions This study confirms the need to introduce correction factors when using a PTW‐31016 chamber and the hypothesis of their low energy dependence. MC simulation has been shown to be a useful methodology to determine detector correction factors for small fields and to analyze the main sources of uncertainty. However, due to the influence of the LINAC jaw setup for field sizes below or equal to 1 cm, MC methods are not recommended in this range for field output factor calculations.</description><subject>Ciències de la salut</subject><subject>Correction factors</subject><subject>Flattening filter-free</subject><subject>Física</subject><subject>Física mèdica</subject><subject>Humans</subject><subject>Medical physics</subject><subject>Medicina</subject><subject>Monte Carlo</subject><subject>Monte Carlo Method</subject><subject>Montecarlo, Mètode de</subject><subject>Output factors</subject><subject>Particle Accelerators</subject><subject>Photons</subject><subject>Radiometry</subject><subject>Setup inaccuracies</subject><subject>Small fields</subject><subject>Uncertainty</subject><subject>Àrees temàtiques de la UPC</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><sourceid>XX2</sourceid><recordid>eNp1kc2KFDEUhYM4OO0o-ASSpZtq81eVZCmDo8KIDTPiMuSv6EhVUiYphnblI7j37XwSq7vH6dUsLpd7-M5Z3APAK4zWGCHydpzWmGAhn4AVYZw2jCD5FKwQkqwhDLXn4Hkp3xFCHW3RM3BOBOmwwHwF_lwFPzhoU87e1pAi7LWtKRfYpww13Nx--_vrN8UId3AT4pRCrHDBwk99oO1Wj8bnA21S3cJ-0LX66B3U0cE5nm7j9VjW8KbObgdTD-vWw1GHCEuas_Vlr83R-lwXsQZfXoCzXg_Fv7zfF-Dr1fvby4_N9ZcPny7fXTeWdlQ2zlrmjDGuZ1J0spOW815IZlpHuCEUc8RbLrHjwrUd06hzFhkiWiko4szRC4CPubbMVi1_8NnqqpIOp2M_BHGiiBAto4vnzdEz5fRj9qWqMRTrh0FHn-aisOCMCyY5P6E2p1Ky79WUw6jzTmGk9vWpcVKH-hb09X3qbEbvHsD_fS1AcwTuwuB3jwapz5tj4D9yHKTi</recordid><startdate>201705</startdate><enddate>201705</enddate><creator>Puxeu‐Vaqué, Josep</creator><creator>Duch, Maria A.</creator><creator>Nailon, William H.</creator><creator>Cruz Lizuain, M.</creator><creator>Ginjaume, Mercè</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>XX2</scope></search><sort><creationdate>201705</creationdate><title>Field correction factors for a PTW‐31016 Pinpoint ionization chamber for both flattened and unflattened beams. Study of the main sources of uncertainties</title><author>Puxeu‐Vaqué, Josep ; Duch, Maria A. ; Nailon, William H. ; Cruz Lizuain, M. ; Ginjaume, Mercè</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3639-dcc4dbbbdf4986969c77f894b5d27b2317075791d78d564a06dc0b285983074d3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Ciències de la salut</topic><topic>Correction factors</topic><topic>Flattening filter-free</topic><topic>Física</topic><topic>Física mèdica</topic><topic>Humans</topic><topic>Medical physics</topic><topic>Medicina</topic><topic>Monte Carlo</topic><topic>Monte Carlo Method</topic><topic>Montecarlo, Mètode de</topic><topic>Output factors</topic><topic>Particle Accelerators</topic><topic>Photons</topic><topic>Radiometry</topic><topic>Setup inaccuracies</topic><topic>Small fields</topic><topic>Uncertainty</topic><topic>Àrees temàtiques de la UPC</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Puxeu‐Vaqué, Josep</creatorcontrib><creatorcontrib>Duch, Maria A.</creatorcontrib><creatorcontrib>Nailon, William H.</creatorcontrib><creatorcontrib>Cruz Lizuain, M.</creatorcontrib><creatorcontrib>Ginjaume, Mercè</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Recercat</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Puxeu‐Vaqué, Josep</au><au>Duch, Maria A.</au><au>Nailon, William H.</au><au>Cruz Lizuain, M.</au><au>Ginjaume, Mercè</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Field correction factors for a PTW‐31016 Pinpoint ionization chamber for both flattened and unflattened beams. Study of the main sources of uncertainties</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2017-05</date><risdate>2017</risdate><volume>44</volume><issue>5</issue><spage>1930</spage><epage>1938</epage><pages>1930-1938</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><abstract>Purpose The primary aim of this study was to determine correction factors, kQclin,Qmsrfclin,fmsr for a PTW‐31016 ionization chamber on field sizes from 0.5 cm × 0.5 cm to 2 cm × 2 cm for both flattened (FF) and flattened filter‐free (FFF) beams produced in a TrueBeam clinical accelerator. The secondary objective was the determination of field output factors, ΩQclin,Qmsrfclin,fmsr over this range of field sizes using both Monte Carlo (MC) simulation and measurements. Methods kQclin,Qmsrfclin,fmsr for the PTW‐31016 chamber were calculated by MC simulation for field sizes of 0.5 cm × 0.5 cm, 1 cm × 1 cm, and 2 cm × 2 cm. MC simulations were performed with the PENELOPE code system for the 10 MV FFF Particle Space File from a TrueBeam linear accelerator (LINAC) provided by the manufacturer (Varian Medical Systems, Inc. Palo Alto, CA, USA). Simulations were repeated taking into account chamber manufacturing tolerances and accelerator jaw positioning in order to assess the uncertainty of the calculated correction factors. Output ratios were measured on square fields ranging from 0.5 cm × 0.5 cm to 10 cm × 10 cm for 6 MV and 10 MV FF and FFF beams produced by a TrueBeam using a PTW‐31016 ionization chamber; a Sun Nuclear Edge detector (SunNuclear Corp., Melbourne, FL, USA) and TLD–700R (Harshaw, Thermo Scientific, Waltham, MA, USA). The validity of the proposed correction factors was verified using the calculated correction factors for the determination of ΩQclin,Qmsrfclin,fmsr using a PTW‐31016 at the four TrueBeam energies and comparing the results with both TLD‐700R measurements and MC simulations. Finally, the proposed correction factors were used to assess the correction factors of the SunNuclear Edge detector. Results The present work provides a set of MC calculated correction factors for a PTW‐31016 chamber used on a TrueBeam FF and FFF mode. For the 0.5 cm × 0.5 cm square field size, kQclin,Qmsrfclin,fmsr is equal to 1.17 with a combined uncertainty of 2% (k = 1). A detailed analysis of the most influential parameters is presented in this work. PTW‐31016 corrected measurements were used for the determination of ΩQclin,Qmsrfclin,fmsr for 6 MV and 10 MV FF and FFF and the results were in agreement with values obtained using a TLD‐700R detector (differences &lt; 3% for a 0.5 cm square field) for the four energies studied. Uncertainty in field collimation was found to be the main source of influence of ΩQclin,Qmsrfclin,fmsr and caused differences of up to 15% between calculations and measurements for the 0.5 cm × 0.5 cm field. This was also confirmed by repeating the same measurements at two different institutions. Conclusions This study confirms the need to introduce correction factors when using a PTW‐31016 chamber and the hypothesis of their low energy dependence. MC simulation has been shown to be a useful methodology to determine detector correction factors for small fields and to analyze the main sources of uncertainty. However, due to the influence of the LINAC jaw setup for field sizes below or equal to 1 cm, MC methods are not recommended in this range for field output factor calculations.</abstract><cop>United States</cop><pmid>28261817</pmid><doi>10.1002/mp.12189</doi><tpages>9</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2017-05, Vol.44 (5), p.1930-1938
issn 0094-2405
2473-4209
language eng
recordid cdi_csuc_recercat_oai_recercat_cat_2072_288543
source MEDLINE; Wiley Online Library Journals Frontfile Complete; Recercat; Alma/SFX Local Collection
subjects Ciències de la salut
Correction factors
Flattening filter-free
Física
Física mèdica
Humans
Medical physics
Medicina
Monte Carlo
Monte Carlo Method
Montecarlo, Mètode de
Output factors
Particle Accelerators
Photons
Radiometry
Setup inaccuracies
Small fields
Uncertainty
Àrees temàtiques de la UPC
title Field correction factors for a PTW‐31016 Pinpoint ionization chamber for both flattened and unflattened beams. Study of the main sources of uncertainties
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-28T09%3A42%3A31IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_csuc_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Field%20correction%20factors%20for%20a%20PTW%E2%80%9031016%20Pinpoint%20ionization%20chamber%20for%20both%20flattened%20and%20unflattened%20beams.%20Study%20of%20the%20main%20sources%20of%20uncertainties&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=Puxeu%E2%80%90Vaqu%C3%A9,%20Josep&rft.date=2017-05&rft.volume=44&rft.issue=5&rft.spage=1930&rft.epage=1938&rft.pages=1930-1938&rft.issn=0094-2405&rft.eissn=2473-4209&rft_id=info:doi/10.1002/mp.12189&rft_dat=%3Cproquest_csuc_%3E1874784977%3C/proquest_csuc_%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1874784977&rft_id=info:pmid/28261817&rfr_iscdi=true