Abdominal CT radiation dose optimization at Siriraj Hospital
OBJECTIVE: To compare radiation dose and image quality between standard dose abdominal CT currently performed at our hospital and new low dose abdominal CT using various percentages (0%, 10%, 20%, and 30%) of Adaptive Statistical Iterative Reconstruction (ASiR). MATERIALS AND METHODS: We prospective...
Gespeichert in:
Veröffentlicht in: | The ASEAN Journal of Radiology 2020-08, Vol.21 (2), p.28-43 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 43 |
---|---|
container_issue | 2 |
container_start_page | 28 |
container_title | The ASEAN Journal of Radiology |
container_volume | 21 |
creator | Apisarnthanarak, Piyaporn Buranont, Chosita Boonma, Chulaluck Janpanich, Sureerat Suwatananonthakij, Tarntip Klinhom, Atchariya Muangsomboon, Kobkun Teerasamit, Wanwarang Suvannnarerg, Voraparee Saiviroonporn, Pairash |
description | OBJECTIVE: To compare radiation dose and image quality between standard dose abdominal CT currently performed at our hospital and new low dose abdominal CT using various percentages (0%, 10%, 20%, and 30%) of Adaptive Statistical Iterative Reconstruction (ASiR).
MATERIALS AND METHODS: We prospectively performed low dose abdominal CT (30% reduction of standard tube current) in 119 participants. The low dose CT images were post processed with four parameters (0%, 10%, 20% and 30%) of ASiR. The volume CT dose index (CTDIvol) of standard and low dose CT were compared. Four experienced abdominal radiologists independently assessed the quality of low dose CT with aforementioned ASiR parameters using a 5-point-scale satisfaction score (1 = unacceptable, 2 = poor, 3 = average, 4 = good, and 5 = excellent image quality) by using prior standard dose CT as a reference of excellent image quality (5). Each reader selected the preference ASiR parameter for each participant. The image noise of the liver and the aorta in all 5 (1 prior standard dose and 4 current low dose) image sets was measured.
RESULTS: The mean CTDIvol of low dose CT was significantly lower than of standard dose CT (7.17 ± 0.08 vs 12.02 ±1.61 mGy, p |
doi_str_mv | 10.46475/aseanjr.v21i2.80 |
format | Article |
fullrecord | <record><control><sourceid>crossref</sourceid><recordid>TN_cdi_crossref_primary_10_46475_aseanjr_v21i2_80</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>10_46475_aseanjr_v21i2_80</sourcerecordid><originalsourceid>FETCH-crossref_primary_10_46475_aseanjr_v21i2_803</originalsourceid><addsrcrecordid>eNqdztEKgjAYhuERBUl5AZ3tBrTNTZ3QSUjReR50Nn5T4Rd1sklQVx-YV9DRBy988BBy4CyUiUzjI7gahtaGr4hjFCq2Il6UpFGQiUysicdUnAVcyseW-M5hyWSScClS5ZHTuaxMjwN0NC-ohQphQjPQyriamnHCHj-_AhO9o0ULLb0ZN-IE3Z5sGuhc7S-7I_x6KfJb8LTGOVs3erTYg31rzvQs1YtUz1KtmPjn8wVjd0sf</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Abdominal CT radiation dose optimization at Siriraj Hospital</title><source>EZB-FREE-00999 freely available EZB journals</source><creator>Apisarnthanarak, Piyaporn ; Buranont, Chosita ; Boonma, Chulaluck ; Janpanich, Sureerat ; Suwatananonthakij, Tarntip ; Klinhom, Atchariya ; Muangsomboon, Kobkun ; Teerasamit, Wanwarang ; Suvannnarerg, Voraparee ; Saiviroonporn, Pairash</creator><creatorcontrib>Apisarnthanarak, Piyaporn ; Buranont, Chosita ; Boonma, Chulaluck ; Janpanich, Sureerat ; Suwatananonthakij, Tarntip ; Klinhom, Atchariya ; Muangsomboon, Kobkun ; Teerasamit, Wanwarang ; Suvannnarerg, Voraparee ; Saiviroonporn, Pairash</creatorcontrib><description>OBJECTIVE: To compare radiation dose and image quality between standard dose abdominal CT currently performed at our hospital and new low dose abdominal CT using various percentages (0%, 10%, 20%, and 30%) of Adaptive Statistical Iterative Reconstruction (ASiR).
MATERIALS AND METHODS: We prospectively performed low dose abdominal CT (30% reduction of standard tube current) in 119 participants. The low dose CT images were post processed with four parameters (0%, 10%, 20% and 30%) of ASiR. The volume CT dose index (CTDIvol) of standard and low dose CT were compared. Four experienced abdominal radiologists independently assessed the quality of low dose CT with aforementioned ASiR parameters using a 5-point-scale satisfaction score (1 = unacceptable, 2 = poor, 3 = average, 4 = good, and 5 = excellent image quality) by using prior standard dose CT as a reference of excellent image quality (5). Each reader selected the preference ASiR parameter for each participant. The image noise of the liver and the aorta in all 5 (1 prior standard dose and 4 current low dose) image sets was measured.
RESULTS: The mean CTDIvol of low dose CT was significantly lower than of standard dose CT (7.17 ± 0.08 vs 12.02 ±1.61 mGy, p<0.001). The mean satisfaction scores for low dose CT with 0%, 10%, 20% and 30% ASiR were 3.95, 3.99, 3.91 and 3.87, respectively with the ranges of 3 to 5 in all techniques. The preferred ASiR parameters of each participant randomly selected by each reader were varied, depending on the readers’ opinions. The mean image noise of the aorta on standard dose CT and low dose CT with 0%, 10%, 20%, and 30% ASiR was 29.07, 36.97, 33.92, 31.49, and 29.11, respectively, while the mean image noise of the liver was 24.60, 30.21, 28.33, 26.25, and 24.32, respectively.
CONCLUSION: Low dose CT with 30% reduction of standard mA had acceptable image quality with significantly reduced radiation dose. The increment of ASiR was helpful in reducing image noise. </description><identifier>ISSN: 0859-144X</identifier><identifier>EISSN: 2672-9393</identifier><identifier>DOI: 10.46475/aseanjr.v21i2.80</identifier><language>eng</language><ispartof>The ASEAN Journal of Radiology, 2020-08, Vol.21 (2), p.28-43</ispartof><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27922,27923</link.rule.ids></links><search><creatorcontrib>Apisarnthanarak, Piyaporn</creatorcontrib><creatorcontrib>Buranont, Chosita</creatorcontrib><creatorcontrib>Boonma, Chulaluck</creatorcontrib><creatorcontrib>Janpanich, Sureerat</creatorcontrib><creatorcontrib>Suwatananonthakij, Tarntip</creatorcontrib><creatorcontrib>Klinhom, Atchariya</creatorcontrib><creatorcontrib>Muangsomboon, Kobkun</creatorcontrib><creatorcontrib>Teerasamit, Wanwarang</creatorcontrib><creatorcontrib>Suvannnarerg, Voraparee</creatorcontrib><creatorcontrib>Saiviroonporn, Pairash</creatorcontrib><title>Abdominal CT radiation dose optimization at Siriraj Hospital</title><title>The ASEAN Journal of Radiology</title><description>OBJECTIVE: To compare radiation dose and image quality between standard dose abdominal CT currently performed at our hospital and new low dose abdominal CT using various percentages (0%, 10%, 20%, and 30%) of Adaptive Statistical Iterative Reconstruction (ASiR).
MATERIALS AND METHODS: We prospectively performed low dose abdominal CT (30% reduction of standard tube current) in 119 participants. The low dose CT images were post processed with four parameters (0%, 10%, 20% and 30%) of ASiR. The volume CT dose index (CTDIvol) of standard and low dose CT were compared. Four experienced abdominal radiologists independently assessed the quality of low dose CT with aforementioned ASiR parameters using a 5-point-scale satisfaction score (1 = unacceptable, 2 = poor, 3 = average, 4 = good, and 5 = excellent image quality) by using prior standard dose CT as a reference of excellent image quality (5). Each reader selected the preference ASiR parameter for each participant. The image noise of the liver and the aorta in all 5 (1 prior standard dose and 4 current low dose) image sets was measured.
RESULTS: The mean CTDIvol of low dose CT was significantly lower than of standard dose CT (7.17 ± 0.08 vs 12.02 ±1.61 mGy, p<0.001). The mean satisfaction scores for low dose CT with 0%, 10%, 20% and 30% ASiR were 3.95, 3.99, 3.91 and 3.87, respectively with the ranges of 3 to 5 in all techniques. The preferred ASiR parameters of each participant randomly selected by each reader were varied, depending on the readers’ opinions. The mean image noise of the aorta on standard dose CT and low dose CT with 0%, 10%, 20%, and 30% ASiR was 29.07, 36.97, 33.92, 31.49, and 29.11, respectively, while the mean image noise of the liver was 24.60, 30.21, 28.33, 26.25, and 24.32, respectively.
CONCLUSION: Low dose CT with 30% reduction of standard mA had acceptable image quality with significantly reduced radiation dose. The increment of ASiR was helpful in reducing image noise. </description><issn>0859-144X</issn><issn>2672-9393</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNqdztEKgjAYhuERBUl5AZ3tBrTNTZ3QSUjReR50Nn5T4Rd1sklQVx-YV9DRBy988BBy4CyUiUzjI7gahtaGr4hjFCq2Il6UpFGQiUysicdUnAVcyseW-M5hyWSScClS5ZHTuaxMjwN0NC-ohQphQjPQyriamnHCHj-_AhO9o0ULLb0ZN-IE3Z5sGuhc7S-7I_x6KfJb8LTGOVs3erTYg31rzvQs1YtUz1KtmPjn8wVjd0sf</recordid><startdate>20200830</startdate><enddate>20200830</enddate><creator>Apisarnthanarak, Piyaporn</creator><creator>Buranont, Chosita</creator><creator>Boonma, Chulaluck</creator><creator>Janpanich, Sureerat</creator><creator>Suwatananonthakij, Tarntip</creator><creator>Klinhom, Atchariya</creator><creator>Muangsomboon, Kobkun</creator><creator>Teerasamit, Wanwarang</creator><creator>Suvannnarerg, Voraparee</creator><creator>Saiviroonporn, Pairash</creator><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>20200830</creationdate><title>Abdominal CT radiation dose optimization at Siriraj Hospital</title><author>Apisarnthanarak, Piyaporn ; Buranont, Chosita ; Boonma, Chulaluck ; Janpanich, Sureerat ; Suwatananonthakij, Tarntip ; Klinhom, Atchariya ; Muangsomboon, Kobkun ; Teerasamit, Wanwarang ; Suvannnarerg, Voraparee ; Saiviroonporn, Pairash</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-crossref_primary_10_46475_aseanjr_v21i2_803</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><toplevel>online_resources</toplevel><creatorcontrib>Apisarnthanarak, Piyaporn</creatorcontrib><creatorcontrib>Buranont, Chosita</creatorcontrib><creatorcontrib>Boonma, Chulaluck</creatorcontrib><creatorcontrib>Janpanich, Sureerat</creatorcontrib><creatorcontrib>Suwatananonthakij, Tarntip</creatorcontrib><creatorcontrib>Klinhom, Atchariya</creatorcontrib><creatorcontrib>Muangsomboon, Kobkun</creatorcontrib><creatorcontrib>Teerasamit, Wanwarang</creatorcontrib><creatorcontrib>Suvannnarerg, Voraparee</creatorcontrib><creatorcontrib>Saiviroonporn, Pairash</creatorcontrib><collection>CrossRef</collection><jtitle>The ASEAN Journal of Radiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Apisarnthanarak, Piyaporn</au><au>Buranont, Chosita</au><au>Boonma, Chulaluck</au><au>Janpanich, Sureerat</au><au>Suwatananonthakij, Tarntip</au><au>Klinhom, Atchariya</au><au>Muangsomboon, Kobkun</au><au>Teerasamit, Wanwarang</au><au>Suvannnarerg, Voraparee</au><au>Saiviroonporn, Pairash</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Abdominal CT radiation dose optimization at Siriraj Hospital</atitle><jtitle>The ASEAN Journal of Radiology</jtitle><date>2020-08-30</date><risdate>2020</risdate><volume>21</volume><issue>2</issue><spage>28</spage><epage>43</epage><pages>28-43</pages><issn>0859-144X</issn><eissn>2672-9393</eissn><abstract>OBJECTIVE: To compare radiation dose and image quality between standard dose abdominal CT currently performed at our hospital and new low dose abdominal CT using various percentages (0%, 10%, 20%, and 30%) of Adaptive Statistical Iterative Reconstruction (ASiR).
MATERIALS AND METHODS: We prospectively performed low dose abdominal CT (30% reduction of standard tube current) in 119 participants. The low dose CT images were post processed with four parameters (0%, 10%, 20% and 30%) of ASiR. The volume CT dose index (CTDIvol) of standard and low dose CT were compared. Four experienced abdominal radiologists independently assessed the quality of low dose CT with aforementioned ASiR parameters using a 5-point-scale satisfaction score (1 = unacceptable, 2 = poor, 3 = average, 4 = good, and 5 = excellent image quality) by using prior standard dose CT as a reference of excellent image quality (5). Each reader selected the preference ASiR parameter for each participant. The image noise of the liver and the aorta in all 5 (1 prior standard dose and 4 current low dose) image sets was measured.
RESULTS: The mean CTDIvol of low dose CT was significantly lower than of standard dose CT (7.17 ± 0.08 vs 12.02 ±1.61 mGy, p<0.001). The mean satisfaction scores for low dose CT with 0%, 10%, 20% and 30% ASiR were 3.95, 3.99, 3.91 and 3.87, respectively with the ranges of 3 to 5 in all techniques. The preferred ASiR parameters of each participant randomly selected by each reader were varied, depending on the readers’ opinions. The mean image noise of the aorta on standard dose CT and low dose CT with 0%, 10%, 20%, and 30% ASiR was 29.07, 36.97, 33.92, 31.49, and 29.11, respectively, while the mean image noise of the liver was 24.60, 30.21, 28.33, 26.25, and 24.32, respectively.
CONCLUSION: Low dose CT with 30% reduction of standard mA had acceptable image quality with significantly reduced radiation dose. The increment of ASiR was helpful in reducing image noise. </abstract><doi>10.46475/aseanjr.v21i2.80</doi></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0859-144X |
ispartof | The ASEAN Journal of Radiology, 2020-08, Vol.21 (2), p.28-43 |
issn | 0859-144X 2672-9393 |
language | eng |
recordid | cdi_crossref_primary_10_46475_aseanjr_v21i2_80 |
source | EZB-FREE-00999 freely available EZB journals |
title | Abdominal CT radiation dose optimization at Siriraj Hospital |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-14T08%3A27%3A22IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-crossref&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Abdominal%20CT%20radiation%20dose%20optimization%20at%20Siriraj%20Hospital&rft.jtitle=The%20ASEAN%20Journal%20of%20Radiology&rft.au=Apisarnthanarak,%20Piyaporn&rft.date=2020-08-30&rft.volume=21&rft.issue=2&rft.spage=28&rft.epage=43&rft.pages=28-43&rft.issn=0859-144X&rft.eissn=2672-9393&rft_id=info:doi/10.46475/aseanjr.v21i2.80&rft_dat=%3Ccrossref%3E10_46475_aseanjr_v21i2_80%3C/crossref%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |