Effective nonapical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy

Current guidelines recommend avoiding apical left ventricular (LV) pacing for cardiac resynchronization therapy (CRT). We investigated the feasibility of nonapical pacing with the current quadripolar LV lead technology. We analyzed consecutive patients who received CRT with an LV quadripolar lead. T...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Kardiologia polska 2021-04, Vol.79 (4), p.442-448
Hauptverfasser: Forleo, Giovanni B, Schiavone, Marco, Della Rocca, Domenico, Solimene, Francesco, Schillaci, Vincenzo, Covino, Gregorio, Sassara, Massimo, Savarese, Gianluca, Donzelli, Stefano, Badolati, Sandra, Gerosa, Carmelo, Lavalle, Carlo, Gasperetti, Alessio, Mitacchione, Gianfranco, Lovecchio, Mariolina, Valsecchi, Sergio, Santini, Luca
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 448
container_issue 4
container_start_page 442
container_title Kardiologia polska
container_volume 79
creator Forleo, Giovanni B
Schiavone, Marco
Della Rocca, Domenico
Solimene, Francesco
Schillaci, Vincenzo
Covino, Gregorio
Sassara, Massimo
Savarese, Gianluca
Donzelli, Stefano
Badolati, Sandra
Gerosa, Carmelo
Lavalle, Carlo
Gasperetti, Alessio
Mitacchione, Gianfranco
Lovecchio, Mariolina
Valsecchi, Sergio
Santini, Luca
description Current guidelines recommend avoiding apical left ventricular (LV) pacing for cardiac resynchronization therapy (CRT). We investigated the feasibility of nonapical pacing with the current quadripolar LV lead technology. We analyzed consecutive patients who received CRT with an LV quadripolar lead. The post--implantation position of each electrode of the LV lead was designated as basal, mid, or apical. The pacing capture threshold (PCT) and phrenic nerve stimulation (PNS) threshold were assessed for each electrode. We enrolled 168 patients. A total of 8 CRT defibrillators were from Biotronik (with Sentus OTW QP leads), 98 were from Boston Scientific (with 21 Acuity X4 Spiral and 77 Acuity X4 Straight leads), and 62 from St. Jude Medical (with Quartet leads). The median (interquartile range) number of electrodes at nonapical segments per patient was 3 (1-4) with Biotronik Sentus leads, 4 (3-4) with spiral -design Boston Scientific leads, 4 (3-4) with straight Boston Scientific leads, and 3 (3-4) with St. Jude Medical Quartet leads (P = 0.045). Three patients (38%) with Biotronik Sentus leads, 21 (100%) with spiral -design Boston Scientific leads, 69 (90%) with straight -design Boston Scientific leads, and 49 (79%) with St. Jude Medical Quartet leads (P
doi_str_mv 10.33963/KP.15882
format Article
fullrecord <record><control><sourceid>pubmed_cross</sourceid><recordid>TN_cdi_crossref_primary_10_33963_KP_15882</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>33750083</sourcerecordid><originalsourceid>FETCH-LOGICAL-c320t-81112fde2819a948de29c000f34b92d8e3387dad46826e8360ea3559bc38c1353</originalsourceid><addsrcrecordid>eNo9kEtPAjEUhRujEUQX_gHTrYvBtnce7dIQUAOJLHQ9ufTh1AydsTNg8NeLoK7uuSdfzuIj5JqzMYDK4W6-HPNMSnFChlyqIklFoU7JkDEhEsVADMhF173vX5lzfk4GAEXGmIQhqabOWd37raWhCdh6jTWtrevp1oY-er2pMdIWtQ9v9NP3Ff3YoIm-bX762qLpqGsi1RiNR02j7XZBV7EJ_gt73wTaVzZiu7skZw7rzl793hF5nU1fJo_J4vnhaXK_SDQI1ieScy6csUJyhSqV-6Q0Y8xBulLCSAsgC4MmzaXIrYScWYQsUysNUnPIYERuj7s6Nl0XrSvb6NcYdyVn5cFWOV-WB1t79ubItpvV2pp_8k8PfAMGIGZ9</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Effective nonapical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy</title><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><creator>Forleo, Giovanni B ; Schiavone, Marco ; Della Rocca, Domenico ; Solimene, Francesco ; Schillaci, Vincenzo ; Covino, Gregorio ; Sassara, Massimo ; Savarese, Gianluca ; Donzelli, Stefano ; Badolati, Sandra ; Gerosa, Carmelo ; Lavalle, Carlo ; Gasperetti, Alessio ; Mitacchione, Gianfranco ; Lovecchio, Mariolina ; Valsecchi, Sergio ; Santini, Luca</creator><creatorcontrib>Forleo, Giovanni B ; Schiavone, Marco ; Della Rocca, Domenico ; Solimene, Francesco ; Schillaci, Vincenzo ; Covino, Gregorio ; Sassara, Massimo ; Savarese, Gianluca ; Donzelli, Stefano ; Badolati, Sandra ; Gerosa, Carmelo ; Lavalle, Carlo ; Gasperetti, Alessio ; Mitacchione, Gianfranco ; Lovecchio, Mariolina ; Valsecchi, Sergio ; Santini, Luca</creatorcontrib><description>Current guidelines recommend avoiding apical left ventricular (LV) pacing for cardiac resynchronization therapy (CRT). We investigated the feasibility of nonapical pacing with the current quadripolar LV lead technology. We analyzed consecutive patients who received CRT with an LV quadripolar lead. The post--implantation position of each electrode of the LV lead was designated as basal, mid, or apical. The pacing capture threshold (PCT) and phrenic nerve stimulation (PNS) threshold were assessed for each electrode. We enrolled 168 patients. A total of 8 CRT defibrillators were from Biotronik (with Sentus OTW QP leads), 98 were from Boston Scientific (with 21 Acuity X4 Spiral and 77 Acuity X4 Straight leads), and 62 from St. Jude Medical (with Quartet leads). The median (interquartile range) number of electrodes at nonapical segments per patient was 3 (1-4) with Biotronik Sentus leads, 4 (3-4) with spiral -design Boston Scientific leads, 4 (3-4) with straight Boston Scientific leads, and 3 (3-4) with St. Jude Medical Quartet leads (P = 0.045). Three patients (38%) with Biotronik Sentus leads, 21 (100%) with spiral -design Boston Scientific leads, 69 (90%) with straight -design Boston Scientific leads, and 49 (79%) with St. Jude Medical Quartet leads (P &lt;0.001) had at least 1 electrode located at nonapical segments linked with a PNS -PCT safety margin of more than 2 V. During the 6-month follow -up, PNS was detected in 4 patients and was eliminated with reprogramming. No significant changes in PCT were detected during follow -up. Quadripolar leads allowed nonapical pacing with acceptable electrical parameters in the majority of CRT recipients, although differences were found among the currently available devices.</description><identifier>ISSN: 0022-9032</identifier><identifier>EISSN: 1897-4279</identifier><identifier>DOI: 10.33963/KP.15882</identifier><identifier>PMID: 33750083</identifier><language>eng</language><publisher>Poland</publisher><ispartof>Kardiologia polska, 2021-04, Vol.79 (4), p.442-448</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c320t-81112fde2819a948de29c000f34b92d8e3387dad46826e8360ea3559bc38c1353</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33750083$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Forleo, Giovanni B</creatorcontrib><creatorcontrib>Schiavone, Marco</creatorcontrib><creatorcontrib>Della Rocca, Domenico</creatorcontrib><creatorcontrib>Solimene, Francesco</creatorcontrib><creatorcontrib>Schillaci, Vincenzo</creatorcontrib><creatorcontrib>Covino, Gregorio</creatorcontrib><creatorcontrib>Sassara, Massimo</creatorcontrib><creatorcontrib>Savarese, Gianluca</creatorcontrib><creatorcontrib>Donzelli, Stefano</creatorcontrib><creatorcontrib>Badolati, Sandra</creatorcontrib><creatorcontrib>Gerosa, Carmelo</creatorcontrib><creatorcontrib>Lavalle, Carlo</creatorcontrib><creatorcontrib>Gasperetti, Alessio</creatorcontrib><creatorcontrib>Mitacchione, Gianfranco</creatorcontrib><creatorcontrib>Lovecchio, Mariolina</creatorcontrib><creatorcontrib>Valsecchi, Sergio</creatorcontrib><creatorcontrib>Santini, Luca</creatorcontrib><title>Effective nonapical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy</title><title>Kardiologia polska</title><addtitle>Kardiol Pol</addtitle><description>Current guidelines recommend avoiding apical left ventricular (LV) pacing for cardiac resynchronization therapy (CRT). We investigated the feasibility of nonapical pacing with the current quadripolar LV lead technology. We analyzed consecutive patients who received CRT with an LV quadripolar lead. The post--implantation position of each electrode of the LV lead was designated as basal, mid, or apical. The pacing capture threshold (PCT) and phrenic nerve stimulation (PNS) threshold were assessed for each electrode. We enrolled 168 patients. A total of 8 CRT defibrillators were from Biotronik (with Sentus OTW QP leads), 98 were from Boston Scientific (with 21 Acuity X4 Spiral and 77 Acuity X4 Straight leads), and 62 from St. Jude Medical (with Quartet leads). The median (interquartile range) number of electrodes at nonapical segments per patient was 3 (1-4) with Biotronik Sentus leads, 4 (3-4) with spiral -design Boston Scientific leads, 4 (3-4) with straight Boston Scientific leads, and 3 (3-4) with St. Jude Medical Quartet leads (P = 0.045). Three patients (38%) with Biotronik Sentus leads, 21 (100%) with spiral -design Boston Scientific leads, 69 (90%) with straight -design Boston Scientific leads, and 49 (79%) with St. Jude Medical Quartet leads (P &lt;0.001) had at least 1 electrode located at nonapical segments linked with a PNS -PCT safety margin of more than 2 V. During the 6-month follow -up, PNS was detected in 4 patients and was eliminated with reprogramming. No significant changes in PCT were detected during follow -up. Quadripolar leads allowed nonapical pacing with acceptable electrical parameters in the majority of CRT recipients, although differences were found among the currently available devices.</description><issn>0022-9032</issn><issn>1897-4279</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNo9kEtPAjEUhRujEUQX_gHTrYvBtnce7dIQUAOJLHQ9ufTh1AydsTNg8NeLoK7uuSdfzuIj5JqzMYDK4W6-HPNMSnFChlyqIklFoU7JkDEhEsVADMhF173vX5lzfk4GAEXGmIQhqabOWd37raWhCdh6jTWtrevp1oY-er2pMdIWtQ9v9NP3Ff3YoIm-bX762qLpqGsi1RiNR02j7XZBV7EJ_gt73wTaVzZiu7skZw7rzl793hF5nU1fJo_J4vnhaXK_SDQI1ieScy6csUJyhSqV-6Q0Y8xBulLCSAsgC4MmzaXIrYScWYQsUysNUnPIYERuj7s6Nl0XrSvb6NcYdyVn5cFWOV-WB1t79ubItpvV2pp_8k8PfAMGIGZ9</recordid><startdate>20210423</startdate><enddate>20210423</enddate><creator>Forleo, Giovanni B</creator><creator>Schiavone, Marco</creator><creator>Della Rocca, Domenico</creator><creator>Solimene, Francesco</creator><creator>Schillaci, Vincenzo</creator><creator>Covino, Gregorio</creator><creator>Sassara, Massimo</creator><creator>Savarese, Gianluca</creator><creator>Donzelli, Stefano</creator><creator>Badolati, Sandra</creator><creator>Gerosa, Carmelo</creator><creator>Lavalle, Carlo</creator><creator>Gasperetti, Alessio</creator><creator>Mitacchione, Gianfranco</creator><creator>Lovecchio, Mariolina</creator><creator>Valsecchi, Sergio</creator><creator>Santini, Luca</creator><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>20210423</creationdate><title>Effective nonapical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy</title><author>Forleo, Giovanni B ; Schiavone, Marco ; Della Rocca, Domenico ; Solimene, Francesco ; Schillaci, Vincenzo ; Covino, Gregorio ; Sassara, Massimo ; Savarese, Gianluca ; Donzelli, Stefano ; Badolati, Sandra ; Gerosa, Carmelo ; Lavalle, Carlo ; Gasperetti, Alessio ; Mitacchione, Gianfranco ; Lovecchio, Mariolina ; Valsecchi, Sergio ; Santini, Luca</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c320t-81112fde2819a948de29c000f34b92d8e3387dad46826e8360ea3559bc38c1353</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Forleo, Giovanni B</creatorcontrib><creatorcontrib>Schiavone, Marco</creatorcontrib><creatorcontrib>Della Rocca, Domenico</creatorcontrib><creatorcontrib>Solimene, Francesco</creatorcontrib><creatorcontrib>Schillaci, Vincenzo</creatorcontrib><creatorcontrib>Covino, Gregorio</creatorcontrib><creatorcontrib>Sassara, Massimo</creatorcontrib><creatorcontrib>Savarese, Gianluca</creatorcontrib><creatorcontrib>Donzelli, Stefano</creatorcontrib><creatorcontrib>Badolati, Sandra</creatorcontrib><creatorcontrib>Gerosa, Carmelo</creatorcontrib><creatorcontrib>Lavalle, Carlo</creatorcontrib><creatorcontrib>Gasperetti, Alessio</creatorcontrib><creatorcontrib>Mitacchione, Gianfranco</creatorcontrib><creatorcontrib>Lovecchio, Mariolina</creatorcontrib><creatorcontrib>Valsecchi, Sergio</creatorcontrib><creatorcontrib>Santini, Luca</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><jtitle>Kardiologia polska</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Forleo, Giovanni B</au><au>Schiavone, Marco</au><au>Della Rocca, Domenico</au><au>Solimene, Francesco</au><au>Schillaci, Vincenzo</au><au>Covino, Gregorio</au><au>Sassara, Massimo</au><au>Savarese, Gianluca</au><au>Donzelli, Stefano</au><au>Badolati, Sandra</au><au>Gerosa, Carmelo</au><au>Lavalle, Carlo</au><au>Gasperetti, Alessio</au><au>Mitacchione, Gianfranco</au><au>Lovecchio, Mariolina</au><au>Valsecchi, Sergio</au><au>Santini, Luca</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Effective nonapical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy</atitle><jtitle>Kardiologia polska</jtitle><addtitle>Kardiol Pol</addtitle><date>2021-04-23</date><risdate>2021</risdate><volume>79</volume><issue>4</issue><spage>442</spage><epage>448</epage><pages>442-448</pages><issn>0022-9032</issn><eissn>1897-4279</eissn><abstract>Current guidelines recommend avoiding apical left ventricular (LV) pacing for cardiac resynchronization therapy (CRT). We investigated the feasibility of nonapical pacing with the current quadripolar LV lead technology. We analyzed consecutive patients who received CRT with an LV quadripolar lead. The post--implantation position of each electrode of the LV lead was designated as basal, mid, or apical. The pacing capture threshold (PCT) and phrenic nerve stimulation (PNS) threshold were assessed for each electrode. We enrolled 168 patients. A total of 8 CRT defibrillators were from Biotronik (with Sentus OTW QP leads), 98 were from Boston Scientific (with 21 Acuity X4 Spiral and 77 Acuity X4 Straight leads), and 62 from St. Jude Medical (with Quartet leads). The median (interquartile range) number of electrodes at nonapical segments per patient was 3 (1-4) with Biotronik Sentus leads, 4 (3-4) with spiral -design Boston Scientific leads, 4 (3-4) with straight Boston Scientific leads, and 3 (3-4) with St. Jude Medical Quartet leads (P = 0.045). Three patients (38%) with Biotronik Sentus leads, 21 (100%) with spiral -design Boston Scientific leads, 69 (90%) with straight -design Boston Scientific leads, and 49 (79%) with St. Jude Medical Quartet leads (P &lt;0.001) had at least 1 electrode located at nonapical segments linked with a PNS -PCT safety margin of more than 2 V. During the 6-month follow -up, PNS was detected in 4 patients and was eliminated with reprogramming. No significant changes in PCT were detected during follow -up. Quadripolar leads allowed nonapical pacing with acceptable electrical parameters in the majority of CRT recipients, although differences were found among the currently available devices.</abstract><cop>Poland</cop><pmid>33750083</pmid><doi>10.33963/KP.15882</doi><tpages>7</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0022-9032
ispartof Kardiologia polska, 2021-04, Vol.79 (4), p.442-448
issn 0022-9032
1897-4279
language eng
recordid cdi_crossref_primary_10_33963_KP_15882
source Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals
title Effective nonapical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-08T08%3A35%3A46IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-pubmed_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Effective%20nonapical%20left%20ventricular%20pacing%20with%20quadripolar%20leads%20for%20cardiac%20resynchronization%20therapy&rft.jtitle=Kardiologia%20polska&rft.au=Forleo,%20Giovanni%20B&rft.date=2021-04-23&rft.volume=79&rft.issue=4&rft.spage=442&rft.epage=448&rft.pages=442-448&rft.issn=0022-9032&rft.eissn=1897-4279&rft_id=info:doi/10.33963/KP.15882&rft_dat=%3Cpubmed_cross%3E33750083%3C/pubmed_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/33750083&rfr_iscdi=true