Comparison of the Transpalpebral Tonometer TGDc-01 with Goldmann Applanation Tonometry
Purpose This study aimed to compare intraocular pressure (IOP) measurements obtained by the transpalpebral tonometer TGDc-01 and by the Goldmann applanation tonometer (GAT). Methods IOP was measured by the conventional GAT and the TGDc-01 tonometer. Central corneal thickness (CCI) was measured for a...
Gespeichert in:
Veröffentlicht in: | European journal of ophthalmology 2010-09, Vol.20 (5), p.902-906 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
This study aimed to compare intraocular pressure (IOP) measurements obtained by the transpalpebral tonometer TGDc-01 and by the Goldmann applanation tonometer (GAT).
Methods
IOP was measured by the conventional GAT and the TGDc-01 tonometer. Central corneal thickness (CCI) was measured for all eyes.
Results
Sixty-seven consecutive patients (123 eyes) participated. The mean difference between the 2 techniques (TGDc-01-GAT) was −0.93 mmHg, SD=2.74 (p=0.0002, paired t-test), and 70% of the measurements were within ±2 mmHg. The correlation coefficient was 0.67 (p |
---|---|